Powerex FirstCall AAP7 Operating & Maintenance Manual

Powerex FirstCall AAP7 Operating & Maintenance Manual

Medical gas alarm

Advertisement

Quick Links

FirstCall™ Medical Gas Alarm
Operating & Maintenance Manual
Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product
described. Protect yourself and others by observing all safety information. Failure to comply with instructions could result in personal
injury and/or property damage! Retain instructions for future reference.
Powerex • 150 Production Drive • Harrison, OH 45030 • USA
IN599102AV • 15 September 2020
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
Page 1 of 36

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FirstCall AAP7 and is the answer not in the manual?

Questions and answers

Summary of Contents for Powerex FirstCall AAP7

  • Page 1 Protect yourself and others by observing all safety information. Failure to comply with instructions could result in personal injury and/or property damage! Retain instructions for future reference. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 •...
  • Page 3: Table Of Contents

    Setup and Configuration Login to Settings Keyboard Basics Zone Served Block Gas Badges Settings Menu Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 3 of 36...
  • Page 4: Responsibilities

    The alarm systems are ETL listed to alarm signals – also customizable UL 1069 and comply with the latest edition of NFPA 99. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 5: Configuration Of Models

    Example: CAP10-OAV2CW-55 Combo Alarm 10” screen Oxygen, Air, Vacuum, N2O, CO2, WAGD, 55 input signals Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 5 of 36...
  • Page 6: Installation

    Risers are type L copper tubing with lead free brass fitting. Risers require gas specific DISS fitting Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 7: Wiring For Power

    Hot wire – Strip the insulation and connect the hot lead to the appropriate terminal block (120V AC). Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 8: Low Voltage Wiring

    Attach front cover by placing threaded studs on lower flange of cover through the provided hinge holes. Place provided hex nuts over studs and tighten. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 9 Plug wired connectors from terminal blocks into appropriate input blocks in PLC. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 9 of 36...
  • Page 10: Wiring Schematics

    Note:Transducers and transducer wiring are field installed TRANS. TRANS. TRANS. TRANS. All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or critical branch of the essential electrical system as required in NFPA 99 RES RES RES RES GF GAS...
  • Page 11: Gas Area Alarm Panels

    Operating & Maintenance Manual Area Alarm Panel Wiring Schematic 5-8 Gas Area Alarm Panels – AP000A008AJ Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 11 of 36...
  • Page 12: Digital Input Master Alarm Panel

    Operating & Maintenance Manual Master Alarm Panel Wiring Schematic 23 Digital Input Master Alarm Panel – AP000M230AJ Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 13: Digital Input Master Alarm Panel

    Operating & Maintenance Manual Master Alarm Panel Wiring Schematic 39 Digital Input Master Alarm Panel – AP000M390AJ Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 14: Digital Input Master Alarm Panel

    Operating & Maintenance Manual Master Alarm Panel Wiring Schematic 55 Digital Input Master Alarm Panel – AP000M550AJ Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 15: Digital Inputs And 1-4 Gas Combo Alarm Panels

    Note:Transducers and transducer wiring is field installed TRANS. TRANS. TRANS. TRANS. All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or critical branch of the essential electrical system as required in NFPA 99 BLUE BLUE-WHITE...
  • Page 16: Digital Inputs And 5-6 Gas Combo Alarm Panels

    Note:Transducers and transducer wiring is field installed TRANS. TRANS. TRANS. TRANS. All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or critical branch of the essential electrical system as required in NFPA 99 BLUE-WHITE BLUE...
  • Page 17: Digital Inputs And 1-4 Gas Combo Alarm Panels

    BLUE COM2 BLUE-WHITE GREEN All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or critical branch of the essential electrical system as required in NFPA 99 Note:Transducers and transducer wiring is field installed TRANS.
  • Page 18: Digital Inputs And 5-6 Gas Combo Alarm Panels

    BLUE COM2 BLUE-WHITE GREEN All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or critical branch of the essential electrical system as required in NFPA 99 Note:Transducers and transducer wiring is field installed TRANS.
  • Page 19: Digital Inputs And 1-4 Gas Combo Alarm Panels

    INPUT BAR 2 (GRAY) BLUE COM2 BLUE-WHITE All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or GREEN critical branch of the essential electrical system as required in NFPA 99 Note:Transducers and transducer wiring is field installed TRANS.
  • Page 20: Digital Inputs And 5-6 Gas Combo Alarm Panels

    BLUE (14X) PLC/HMI PLC/HMI INPUT BAR 2 (GRAY) BLUE All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or COM2 critical branch of the essential electrical system as required in NFPA 99 BLUE-WHITE GREEN Note:Transducers and transducer wiring is field installed TRANS.
  • Page 21: Setup And Configuration

    Text can be added to the black-colored bar underneath the gas to further specify the location where each gas is being used. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 22: Settings Menu

    Press the “SAVE” button at the bottom to save and input signals available in the Powerex FirstCall™ changes. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 23: Alarm Summary

    Instrument Air This screen is only applicable to Area alarms and Combo alarms. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 23 of 36...
  • Page 24: Master Alarm Settings

    There is also an option to manually add the alarm signal description by Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 25: Dry Contact Settings

    The lower right bar on the home screen that is typically Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 26: Transducer Settings

    20 percent from the normal Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 27: Operation

    3. Low – gas pressure less than 20% below nominal. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 28: Silencing The Alarm

    To see the specific alarm, press the gas badge that’s in alarm to view the list of alarm conditions. The specific alarm will be highlighted in red. Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 29: Maintenance

    On the main screen, the Green “Normal” bar changes to Yellow “Maintenance.” Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 30: Parts Maintenance List

    Digital Input Only Expansion Module PE0004104AV Digital Input and Output Expansion Module PE0004105AV Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 30 of 36...
  • Page 31: Troubleshooting Guide

    Replace sensor if necessary (see replacement part list for part number Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 32: Specifications

    Front Panel – Large: 14” X 15.4” Transducers: 1.18” Diameter X 2.03” tall (Does not include DISS fitting) Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com...
  • Page 33: Appendix A. Bacnet Object Names

    General Fault status. A displayed value of G6HiAlarm Gen Fault 0=Normal, 1=Alarm. G7HiAlarm G8HiAlarm Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 33 of 36...
  • Page 34 FirstCall™ Medical Gas Alarm Operating & Maintenance Manual Notes Powerex • 150 Production Drive • Harrison, OH 45030 • USA IN599102AV • 15 September 2020 P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com Page 34 of 36...
  • Page 35: Warranty

    Defect, and Powerex shall replace any Product which it determines to have had a Defect; provided, however, that Powerex may elect, upon return of the Product, to refund to buyer any part of the purchase price of such Products paid to Powerex.
  • Page 36 (b) Powerex will not assume any expense or liability for repairs to products made outside of its plant by third parties. In the event Powerex elects to replace a defective product, costs of installation, labor, service, and all other costs to replace the product shall be the responsibility of Buyer.