User Manual
Controls, Indicators, and Connections
1
READY
2
STANDBY
3
INCREASE/
DECREASE POWER setting to the desired level (from 1 to 5). The level chosen will be shown on the
LEVEL
4
POWER
5
VOLUME
6
MIN
7
MAX
8
ALARM INDICATOR This red indicator appears only if a system alarm occurs in response to a component
4
Fig. 2-1 Front Panel
When this indicator is green, the system is ready for activation.
Push this button to toggle between Standby and Ready modes. In Standby mode,
this button, and the STANDBY icon, light up and all power is removed from the
hand piece. Both the foot switch and hand switch are disabled. Upon power-up, the
system defaults to Standby mode enabled.
Push this button to increase or decrease the minimum (MIN) power
graphic display. The power level may be adjusted when the generator is in Ready or
Standby mode.
This switch controls the main electrical power to the generator.
Turn this knob to adjust the volume of the activation tones. A tone will sound
indicating the volume level selected.
Indicates the user-settable minimum power level setting. When this power level is
activated (by foot switch or hand switch), the "MIN" indicator will flash. The
system defaults to "MIN" power level 3. Refer to the instruments' package inserts
for the recommended minimum power level.
Indicates the maximum power level setting. This setting is always "5". When this
power level is activated (by foot switch or hand switch), the "MAX" indicator
will flash.
or generator problem.