Replacement Parts - Fluke PM6685 Service Manual

Programmable frequency counter
Hide thumbs Also See for PM6685:
Table of Contents

Advertisement

Select NULL by pressing s or t.
Press ENTER twice.
Set the measuring time to 10 s.
Connect the 10 MHz reference to the input A of the counter.
Adjustment procedure
Remove the seal from the front panel.
Adjust the potentiometer beyond the seal until the display
-3
reads 0.5
Hz or less.
Check that the value is stable over time, (more than 30 min -
utes).
Cover the "CALIBRATION ADJUSTMENT" hole on the front
panel with a relevant seal if necessary.

Replacement Parts

Pos
Description
Cover, (incl. front part).
Fan
Text plate kit
Rear plate
A1
Power supply
D1
LED, HLMP-1300, red
L1
Toroid
L2
Toroid
L3
Toroid
P1-P3
Cable kit
R1
Potentiometer, 1 kW
R2
Resistor 3.83 kW, 1% 0.5 W MRS25
NOTE: The rubidium time base (unit A2) must be sent to a
Fluke service center for repair. Follow the exchange
procedure.
9-14 Replacement Parts
Part Number
P
5322 447 92194
P
2822 031 01327
R
4031 100 62440
R
4031 100 53930
P
5322 214 91268
P
5322 130 81921
5322 526 10545
5322 526 10545
5322 526 10545
4031 100 61530
P
5322 101 11298
4822 050 23832

Hide quick links: