1. Fill Unit
Move drain valve to closed position.
n
At the beginning of each day you will need to "Degas" your unit. In order to perform this
n
important function please fill you ultrasonic cleaner to the max fill line and run the unit for
15 minutes. This will minimize the microscopic bubbles in the tank and improve cleaning
effectiveness.
If heat is desired please adjust to appropriate temperature (110
n
"Temp" then press the "Heating" button to initiate the feature. (Note: Actual temperature will
display once heat is started and may take up to 30 minutes to reach desired temperature).
After completing the Degas operation, please add the correct amount of enzymatic cleaning
n
solution. (Tri-Clean triple enzymatic cleaner = 1 oz. per gal.)
Carefully place items to be cleaned in basket and lower into solution.
n
Place lid on the top of the unit.
n
2. Start Cleaning Cycle
Caution, while using heat does improve your cleaning results, it also results in the solution
n
becoming hot. Please use extreme caution when operating the unit and do not place hands in
the solution or spill solution on you.
Set desired time of operation by using the arrows labeled "Time".
n
Start the cleaning cycle by pressing the button labeled "Ultrasonic".
n
3. Unit Operation
Wait for cleaning cycle to finish or terminate the cycle early by pressing the ultrasonic button.
n
(Note: Do not stop the cycle by using the large green illuminated switch labeled "Power").
Turn off power to ultrasonic cleaner by using the green illuminated switch labeled "Power".
n
(Note: Power switch for recessed models is not illuminated and located on side of control panel).
Carefully remove basket from solution and allow excess liquid to drain off before moving away
n
from tank.
Drain contents of tank by opening the valve located on the side of unit. (Note: Recessed models
n
must be powered on to activate drain valve by pressing button labeled "Drain" on control panel).
Flush tank with warm water and clean with a non-acetone or non-phenol based cleaner if film is
n
present on tank walls.
www.DuralineSystems.com
Please visit
www.bit.ly/cleaninstrument
Note – please contact the instrument manufacturer to confirm cleaning guidelines.
Instructions for Use
F) using the arrows labeled
o
for basic tips on effective instrument cleaning.
BrandMax
®

Advertisement

loading
Need help?

Need help?

Do you have a question about the BrandMax TRI-CLEAN and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Duraline Systems BrandMax TRI-CLEAN

  • Page 1 Instructions for Use TRI-CLEAN ® 1. Fill Unit Move drain valve to closed position. At the beginning of each day you will need to “Degas” your unit. In order to perform this important function please fill you ultrasonic cleaner to the max fill line and run the unit for 15 minutes.
  • Page 2 Maintenance Schedule DO THIS... DAILY WEEKLY MONTHLY TRI-CLEAN ® Change ultrasonic solution when visibly soiled or once a day, which ever is more often. Rinse the tank thoroughly and wipe down with soft cloth at the end of each work day. Daily, flush drain tube with hot water to ensure smooth draining cycles.

This manual is also suitable for:

U-13lhrecU-19lhrecU-10lhrecU-20lhrec