Nemesys M
Hardware Manual
ORIGINAL INSTRUCTIONS 1.00 – APRIL 2021

Advertisement

Table of Contents
loading

Summary of Contents for cetoni Nemesys M

  • Page 1 Nemesys M Hardware Manual ORIGINAL INSTRUCTIONS 1.00 – APRIL 2021...
  • Page 2 CETONI GmbH Wiesenring 6 07554 Korbussen Germany +49 (0) 36602 338-0 +49 (0) 36602 338-11 info@cetoni.de www.cetoni.de CETONI Nemesys M Hardware Manual...
  • Page 3 If you control the products with a software from CETONI GmbH, you agree to the applicable license agreement, which can be read in the corresponding software manual. This and all other current product manuals can be found at https://www.cetoni.de/en/downloads/manuals.
  • Page 4 CETONI Nemesys M Hardware Manual...
  • Page 5: Table Of Contents

    Reasonably Foreseeable Faulty Application 2.4.4 Safety measures 2.4.5 Measures for Safe Operation Warranty and Liability Scope of Delivery Technical Data Environment Mechanical Data Electrical Data Interfaces Dimensional Drawing Dosing Performance Materials (Syringes) Operating the Hardware CETONI Nemesys M Hardware Manual...
  • Page 6 OEM RS232 Cable Set Overload shut-off Syringe Maintenance 5.8.1 Piston Removal/Installation 5.8.2 Piston (Dis)Assembly 5.8.3 Front Cap Removal/Installation 5.8.4 Mounting the seal wash set 5.8.5 Maintenance and Care Transport and Storage Maintenance and Care Disposal CETONI Nemesys M Hardware Manual...
  • Page 7: Revision History

    Please let us know your device type and email address and we will send you the appropriate manual as a pdf file. The type of your product can be found on the label behind "Type:", according to the marked number in the following example: CETONI Nemesys M Hardware Manual...
  • Page 8: Introduction

    Introduction Preface Thank you for purchasing a product from CETONI. With this user manual we would like to support you as well as possible when handling the device. If you have any questions or suggestions, please do not hesitate to contact us.
  • Page 9: Norms And Directives

    European directives. Application Purpose 2.4.1 General Description of the Device The CETONI Nemesys M is a syringe pump. It allows emptying and filling syringes by the relative linear movement of a syringe- and a piston holder. 2.4.2 Intended Use The CETONI Nemesys M is used for highly precise and low-pulsation dosing of fluids in the range of nanoliters to milliliters per second.
  • Page 10: Safety Measures

    2.4.4 Safety measures The safety of the user and a failure-free operation of the CETONI Nemesys M are assured only if original parts are used. Only original accessories may be used. Warranty claims will not be accepted for damage due to the use of alien accessories or expendables.
  • Page 11: Measures For Safe Operation

    2.4.5 Measures for Safe Operation 2.4.5.1 ELECTROMAGNETIC EMISSIONS The CETONI Nemesys M is intended for use in any type of facility, connected directly to the public power supply network that supplies buildings used for domestic purposes. 2.4.5.2 ELECTROSTATIC DISCHARGE Floors should be made of wood, concrete, or ceramic tiles. If the flooring is made of a synthetic material;...
  • Page 12: Warranty And Liability

    CETONI GmbH considers itself responsible for the devices with regard to safety, reliability and function only if assembly, new settings, changes, extensions and repairs are carried out by CETONI GmbH or an authorised centre, and if the devices have been used in accordance with the instruction manual.
  • Page 13: Scope Of Delivery

    Scope of Delivery The following items should be included: CETONI NEMESYS M  Star nut for syringe fixation Other accessories such as syringes, tubing material, pressure sensors etc. have to be purchased separately. CETONI Nemesys M Hardware Manual...
  • Page 14: Technical Data

    DIMENSIONS (L x W x H) 310 x 56 x 108 mm WEIGHT 2,3 kg Electrical Data SUPPLY VOLTAGE (POWE R SUPPLY UNIT) 24 V DC POWER CONSUMPTION 24 W Interfaces 1 Mbit/s RS-232 Section 5.6 ACCESSORY PORT Section 5.5 CETONI Nemesys M Hardware Manual...
  • Page 15: Dimensional Drawing

    Dimensional Drawing CETONI Nemesys M Hardware Manual...
  • Page 16: Dosing Performance

    Dosing Performance The following table gives an overview of the minimum and maximum dosing speeds and the resulting flow rates for the CETONI stainless steel syringes. Below the speeds and flow rates referred to as pulsation-free, the dosing precision slowly decreases.
  • Page 17: Materials (Syringes)

    3 ml 10x1,5 5,5x1,5 5,5x1 7x1,5 5 ml 17x1,5 12x1,5 7x1,5 7x1,5 10 ml 17x1,5 16x1,5 11x2 11x1,5 7x1,5 25 ml 34x1,5 24x1,5 18x2,5 18x2 10x1,5 50 ml 34x1,5 34x1,5 28x2,5 28 x 2 10x1,5 CETONI Nemesys M Hardware Manual...
  • Page 18 When used for the first time, sliding rings require a few syringe strokes until they achieve their full sealing effect. This can also be done manually and with an empty syringe. Please also take this into account after a seal change. CETONI Nemesys M Hardware Manual...
  • Page 19: Operating The Hardware

    System Manual. However, there are the following deviations when configuring the CETONI Nemesys M and when separating: The CETONI Nemesys M is equipped with a rotary switch. This allows the module to be switched off so that it can remain connected to the system while other modules are being configured. The...
  • Page 20: Mounting A Syringe

    Mounting a Syringe Before mounting a syringe on the CETONI Nemesys M, it must be configured and selected in the software. The procedure is described in the software manual. For this you need the Volume (Scale Volume), the nominal stroke (Scale Length) and the maximum stroke (Piston Stroke).
  • Page 21 Scale Volume Scale Length Proceed as follows to mount a syringe on the CETONI Nemesys M Use the software to move the piston holder to the foremost position. Push the syringe piston into the syringe as far as it will go and screw the star nut onto the syringe by six to seven turns:...
  • Page 22 Furthermore, pull the fixing screw back until its tip disappears in the adjusting screw. Now insert the syringe into the syringe holder and tighten the star nut hand-tight. Star Nut CETONI Nemesys M Hardware Manual...
  • Page 23 If the piston holder is set for a particular syringe, you can remove this syringe and, for example, insert it again when it is filled. IMPORTANT. Syringes, and particularly the seals, are wear parts. Check them on a regular basis and replace them if necessary. CETONI Nemesys M Hardware Manual...
  • Page 24: Fluidic Connection

    The required tightening torque is 45 Nm.  9/16" for compression fittings for 1/16" and 1/8" pipes  14 mm for compression fittings for 2, 3 and 4 mm pipes Metal Sealing Disc 16 mm Open-End Wrench Compression Fitting CETONI Nemesys M Hardware Manual...
  • Page 25: First-Time-Tube-Installation

    12 mm for compression fittings for 2, 3 and 4 mm tubes CAUTION. Only use fittings and capillaries specified for the anticipated pressure levels. After connecting, check the tightness of all fluidic connections on a regular basis. CETONI Nemesys M Hardware Manual...
  • Page 26: Tube Disassembly

    Fitting Body Marker Line Cap Nut Pull out the tube. The cap nut and the ferrules remain on the tube. CETONI Nemesys M Hardware Manual...
  • Page 27: Tube Reassembly

    For easier bleeding of the syringes it is often helpful to operate the syringe pump upright. To increase the standing safety in this case, the CETONI Nemesys M is equipped with a stand foot. This is magnetically located in a recess on the side. Take it out, turn the module upright and put the stand foot back in place rotated by 90°.
  • Page 28: Accessory Port

    The accessory port allows the integration of accessories such as valves and pressure sensors. To do this, plug the cable connector of the accessory component into the socket of the CETONI Nemesys M until it engages. Make sure that the plug can only be inserted when the coding nose is pointing upwards.
  • Page 29: Rs-232 Connection

    Now, you can reactivate your system and send or receive data through RS232. Since every module contains a gateway from RS232 to the system’s internal CAN bus, you can now address each module of your system with only one RS232 cable. CETONI Nemesys M Hardware Manual...
  • Page 30 Nemesys interface and the 9-pin D-Sub: Nemesys Interface (socket) 9-pin D-Sub socket Cable RS232 RX Orange TXD Transmit Data RS232 TX Brown RXD Receive Data Signal GND Black GND Signal GND D-Sub housing = shielding Shielding CETONI Nemesys M Hardware Manual...
  • Page 31: Overload Shut-Off

    Overload shut-off The CETONI Nemesys M has a nominal thrust force of 1000 N, which enables pressure to be built up in the syringe as well as the associated application. To protect the device and the application from damage due to overload, it is equipped with a force measuring device. If the nominal force of 1000 N is exceeded by 30%, the drive is switched off.
  • Page 32: Syringe Maintenance

    To remove the piston, the rear cap must be removed. For this purpose, the piston end piece of the 3-, 5- and 10-ml syringes must first be removed. Use a 5.5 mm and an 8 mm open-end wrench for this to be done. Piston End Piece (3, 5 und 10 ml) CETONI Nemesys M Hardware Manual...
  • Page 33 Then pull the rear cap, and for the 25- and 50-ml syringes also the adapter plate, off the piston rod. Adapter Plate (25 and 50 ml) Rear Cap (25 and 50 ml) Rear Cap (3, 5 and 10 ml) CETONI Nemesys M Hardware Manual...
  • Page 34: Piston (Dis)Assembly

    To disassemble the piston, unscrew the grub screw located on the side of the piston by about two and a half to three turns. Use a 1.3 mm Allen key for the 3-, 5- and 10-ml syringes and a 1.5 mm Allen key for the 25- and 50-ml syringes. Grub Screw CETONI Nemesys M Hardware Manual...
  • Page 35 When used for the first time, sliding rings require a few syringe strokes until they achieve their full sealing effect. This can also be done manually and with an empty syringe. Please also take this into account after a seal change. CETONI Nemesys M Hardware Manual...
  • Page 36: Front Cap Removal/Installation

    O-ring if necessary. The dimensions of the required O-rings can be found in section 4.7. Adapter Plate (25 and 50 ml) Front Cap (3, 5 and 10 ml) O-Ring O-Ring Rear Cap (25 and 50 ml) CETONI Nemesys M Hardware Manual...
  • Page 37: Mounting The Seal Wash Set

    10 ml syringes, the end of the piston as well. Then insert the O-ring into the cylinder (for dimensions, see section 4.7) and push the connection piece on until it rests against the cylinder. Connection Piece O-Ring CETONI Nemesys M Hardware Manual...
  • Page 38: Maintenance And Care

    A lubrication of the O-rings for example with silicone grease increases their service life considerably and should be made if your application allows. Replacement seals can be obtained from CETONI GmbH. However, O-rings from other suppliers may also be used, provided they meet the specifications in section 4.7.
  • Page 39: Transport And Storage

    Please do not lift and transport the devices in the plugged together state, unless you use the original packaging. The CETONI Nemesys M contains a lithium metal battery. A consignment may consist of a maximum of two packages, each containing a maximum of four devices. No further markings are required for shipping.
  • Page 40: Maintenance And Care

    CETONI GmbH to coordinate further actions. The device may only be opened by CETONI GmbH or thereby authorized service staff. Otherwise, the warranty claims are void.
  • Page 41: Disposal

    Disposal Please send your old devices back to CETONI GmbH. We will take care of proper disposal. The device contains a battery and must not be disposed of with domestic waste. Please follow the instructions in chapter 6 for the return shipment.