Manual Supplement
In the Specifications section on page 2, change the Warranty,
From:
One year
To:
Three years limited
Change #4
On page 6, under the Current-Input Fuses section add the following Table:
F1
F2
On page 6, under the Warning replace step 3 with:
Press to set the range to 200 .
On page 7, under the Warning replace step 5 with:
5. Remove the defective fuse and replace it with a fuse of an appropriate rating. See the Current Input
Fuse Table.
Change #5,
64137, 64189
On page 4, add the following to the Symbols, table:
On page 2, in the Specification section, add
Safety and EMC with:
Safety
................................................................ IEC 61010-1, 600 V CAT II / 1000 V CAT I, Pollution Degree 2
Electromagnetic Environment
EMC
.................................................................... Applies to use in Korea only. Class A Equipment (Industrial Broadcasting &
Current Input Fuse Table
F1 - Fuse, 11 A, 1000 V, Fast blow or equivalent
F2 - Fuse, 440 mA, 1000 V, Fast blow or equivalent
Conforms to relevant South Korean EMC Standards.
Conforms to CAN/CSA-C22.2 No. 61010-1, second edition,
including Amendment 1.
.............. IEC 61326-1, Basic
Electromagnetic
[1]
Communication Equipment)
[1]
This product meets requirements for industrial (Class A)
electromagnetic wave equipment and the seller or user should take
notice of it. This equipment is intended for use in business
environments and is not to be used in homes.