Page 6
Removal and cleaning of valve head 1 . Release both of the black hose couplings GL-14 between The pump heads. 2. Using the allen wrench provided, loosen crosswise with a quarter-turn each and remove the 4 screws on the valve head top.
Page 7
Replacement of the pump head and assembling 1.Before replacing the valve head check if the valves are not stuck together by holding the head between two fingers. You should hear a soft clicking while shaking it carefully. 2. Place the valve head into the valve head top. Check if the rubber sealing is symmetrical placed in the metal part of the valve head top.
Page 8
5. Now connect the connecting hose, tighten the hose con- nections, and then assemble the second pump head the same way as the first one. 6. If the vacuum is not achieved, it may be necessary to replace or clean the valve head seals. To do this, remove the valve heads as indicated in 6.2.
Page 9
● Technical consulting. ● Working with strong acids / bases en Buchi VAC V-500 - 513 If strong acids or bases are removed with a Vacuum pump, protecting clothing such as goggles and laboratory coats must be worn during operation of the device.
Page 10
Putting out of order Before the device can be moved, the power cable must be disconnected and all hoses taken off. Storage / Transport Clean the device thoroughly (see 6.1). Any chemicals remaining in the device must be removed completely and the glass pans must be washed.
Page 11
Spare parts Only original spare parts and accessories from BUCHI guaran- tee that the device will function properly and safely. Spare pans and accessories from other manufacturers may be used only with the express permission of BUCHI AG. General safety regulations and page 10 must be observed when assembling or dismantling the device.