BioLife ThawSTAR CFT2 Instructions For Use Manual

BioLife ThawSTAR CFT2 Instructions For Use Manual

Automated cell thawing system

Advertisement

ThawSTAR
Automated Cell Thawing System

1. ThawSTAR Automated Cell Thawing System Function and Description

ThawSTAR
Automated Cell Thawing System (Item No. AST-601) is designed to rapidly thaw live biological contents of 1.8-2.0 mL cryogenic
®
vials taken directly from liquid nitrogen (LN2) storage, a -80°C freezer or vials equilibrated and held at dry ice temperature (-78.5°C) in a
container such as the ThawSTAR
vial receives a customized thaw cycle based upon multiple built-in detection parameters. The intended result is a thermal profile equivalent
to a water bath thaw, but without the water, and comparable post-thaw cell recovery and viability.
LEDs
Base light

1.1 General Safety

ThawSTAR Automated Cell Thawing System is intended for the exclusive use by trained and experienced laboratory personnel.
1.1.1. Review instructions before installing, using or cleaning the unit.
1.1.2.
DANGER: To prevent risk of severe or fatal electrical shock, use only the supplied AC power supply with cables for
the US, UK and EU packaged with the unit.
1.1.3.
WARNING: Do not place near flammable materials. This unit is not designed to be explosion-proof or spark-proof.
1.1.4.
WARNING: Avoid prolonged exposure to ultraviolet light such as sunlight or germicidal lamps unless covered by
light resistant plastic bag or a large piece of unbroken aluminum foil placed over the unit when not in use.
1.1.5.
WARNING: Do not clean unit with ketone solvents; use recommended cleaning protocol in section 5.
1.1.6.
CAUTION: Do not place near or immerse in water or other liquids.
1.1.7.
WARNING: Removal of shell without authorization will void your limited warranty. Do not remove exterior shell of
the unit unless instructed to do so by an authorized BioLife representative.
1.1.8.
WARNING: Unit has a covered mini-USB port on the bottom; do not insert any USB plug into this port or attempt to connect unit
with any other device, (eg. a computer), unless instructed to do so by an authorized BioLife representative.
1.1.9.
WARNING: Do not use flag labels or labels that have a tab extending beyond the vial body or with folds or ridges with the
unit as these features may interfere with heat transfer or cause vial to become lodged in the unit.
1.1.10.
CAUTION: The unit is designed for stationary indoor laboratory bench top use; outdoor "field" use or use on a
laboratory bench subject to strong vibrations such as near a centrifuge, shaker, or vortex mixer is not advised.
1.1.11. Ensure the unit is placed securely on a level surface during operation.
1.1.12. Recommended for use with standard 1.8 to 2.0 mL cryogenic vials containing cryopreserved cells; the ThawSTAR
system is optimized for 1.o mL volume of aqueous solution at a starting temperature of <- 70°C.
1.1.12.1. Ensure vials were frozen in an upright and vertical position and that frozen contents are at the bottom of the vial.
1.1.12.2. Vials should be gently inserted into the top opening of the unit;
should be inserted into the unit at any time.
1.1.13.
CAUTION: Keep liquids out of the unit; ensure cryogenic vials have clean exteriors and are tightly capped to prevent leaks.
1.1.14.
CAUTION: Do not touch the vial with hands or any other object until the thawing process is finished (after
cryogenic vial is inserted into the opening of the unit and the thawing process has been initiated).
1.1.15. Thawing time is typically 2-3 minutes from the time of vial insertion. Laboratory personnel are advised to be in close
proximity to the unit while in use so that vials may be immediately retrieved upon thaw completion.
1.1.16. Before moving the unit, ensure that any cryogenic vial has been removed from unit opening.
1.1.17. Keep unit opening covered while not in use.
1.1.18.
HOT SURFACE WARNING: The heating elements inside the vial opening are heated while the unit is turned on. Avoid
inserting body parts such as fingers or any object except a 1.8 to 2.0 mL cryogenic vial into the vial opening at any time.
1.1.19. Ensure the power cord can be easily disconnected from the power supply when placing the unit on a laboratory bench.
CFT2
®
CFT2 Transporter (Item No. AST-602). Utilizing patent-pending STAR
®
Power
connection
Unit
opening
On/Off
switch
CAUTION: NO other objects (e.g. forceps)
Instructions for Use
adaptive sensing technology, each
Mini USB port (to be
used only by authorized
service personnel)
Speaker

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ThawSTAR CFT2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BioLife ThawSTAR CFT2

  • Page 1 WARNING: Unit has a covered mini-USB port on the bottom; do not insert any USB plug into this port or attempt to connect unit with any other device, (eg. a computer), unless instructed to do so by an authorized BioLife representative.
  • Page 2 1.2 Use As Intended If the ThawSTAR Automated Cell Thawing System is used in a manner not specified by BioLife in these Instructions for Use, operators are ® regarded as proceeding under their own risk and warranty terms and conditions may be voided.
  • Page 3 Note: If unit does not reach a “ready” state within 10 minutes of powering up and the power supply was properly connected to the unit and plugged into an electrical outlet with a provided cable, please contact BioLife customer service.
  • Page 4 Push the top of the vial down gently until the unit Two-tone, rising emits a deep dual tone. The vial will now remain in a depressed position. The unit will begin phase change detection using STAR™ sensing technology. Front: Three LED light triangle rotating clockwise Base: Pulsing pattern Beginning of solid to liquid phase change is...
  • Page 5 Ensure vial is within specifications outlined in section 1.5. (4) Try to thaw frozen ThawSTAR Confirmation Vial. If ® situation repeats itself with same error code 5, hardware issue (latching fault) may be the root cause; contact BioLife customer service.
  • Page 6 Note: If thermal pads are 9 (nine LEDs at the top of the not intact and/or damaged, contact circle: 12,13,14,15,1,2,3,4,5) BioLife customer service. Avoid sticking forceps or other foreign objects into the (If in step 8, error code 7 opening.
  • Page 7 Automated Cell Thawing System for any reason, it is required that you ® decontaminate the unit prior to sending. Contact BioLife technical support or a BioLife authorized representative for decontamination instructions and a Return Material Authorization (RMA) Number. 8.2 Ordering Information Item No.
  • Page 8 ©2019. BioLife Solutions, Inc. All rights reserved. Patents pending. info@biolifesolutions.com | www.biolifesolutions.com BioLife, ThawSTAR, STAR designations are trademarks owned by BioLife Solutions, Inc. IG-AST-19052R01...