Solaxx SafeDip Operating Instructions page 2

Digital chemistry reader

Advertisement

Calibration Procedure
Calibration should be performed once a month. In order to calibrate,
a buffer solution is used. This solution is easily made using the buffer
packets provided with the original purchase. The buffer solution
insures the accuracy of the Safedip digital reader.
To create a buffer solution, follow these steps:
1.Empty entire contents of 1 buffer packet into a cup with 8 oz tap
water.
2.Stir the solution using a spoon for 20 seconds or until powder is
completely dissolved.
To calibrate, follow these steps:
1. Turn ON the Safedip.
2. Immerse the probes completely into the cup with buffer solution.
3. Gently swirl the probes to remove possible gas bubbles on the
probe tip.
4. With the probe in the solution, press and hold the CAL button for
more than 3 seconds and the 7.0 value will be displayed.
Release the button and the 7.0 value will blink. Press CAL button
again and the
will then be displayed until calibration is
finished.
5. After the calibration is completed, the
with the calibration value of 7.0. While the 7.0 value is blinking,
press
at anytime to abort.
Maintenance
Probe Care and Storage
1. Rinse the probe in distilled or tap water.
2. Store the Safedip with the probe sleeve over the
electrode. Always keep the sponge in the cap soaked with
tap water or pH 7 buffer solution.
3. Always rinse the probes in distilled or tap water between
measurements to avoid cross contamination. Double rinsing
is recommended when high accuracy is required.
4. Do not touch the probes. Touching the surface of the
probes may damage and reduce the life of the probes.
pH Probe Replacement
1. Unscrew and remove the pH probe collar. Twist collar
counter-clockwise.
2. Gently pull the probe away from the meter until it
disconnects from the probe socket.
3. To attach a new probe, align the slots and carefully twist
the probe clockwise into the meter socket.
4. Firmly tighten the probe collar to create a seal with the
rubber gasket between the probe and the meter.
Probe Cleaning Recommendations
Do not soak the probe in any cleaning solution. To do so may
cause a reference potential shift which will cause
degradation in performance. When cleaning the probe, use
tap water and be careful not to scratch or damage the glass
sensing surface.
Battery Replacement
1. Use a coin to twist off the battery compartment cap.
2. Replace the two AAA batteries. Observe polarity.
3. Replace the battery compartment cap and make sure it is
tightly secured to avoid water intrusion.
Go to www.safedip.com for more detailed instructions.
Troubleshooting
will be displayed along
Go to www.safedip.com for more detailed instructions.
WARRANTY
This instrument is guaranteed to be free from defects
in material and workmanship for a period of one year
from the date of original purchase. The probes are
guaranteed to be free from defects in material and
workmanship for a period of six months from the date
of original purchase. Damages brought on by abuse,
operator negligence, accident, misapplication,
mishandling, or acts of God will void this warranty.
This warranty is limited to the repair or replacement
of the product at the discretion of the manufacturer
and will not include transportation costs. This
warranty does not cover batteries.
This product cannot be returned without a return
authorization number from Solaxx. For warranty
support or a Return Authorization number, log onto
www.solaxx.com and click on support or email:
service@solaxx.com
Digital Chemistry Reader

Advertisement

loading
Need help?

Need help?

Do you have a question about the SafeDip and is the answer not in the manual?

Questions and answers

nelson Garcia
April 7, 2025

I need my Safe Dip to show me all the parameter values

User image 67f42df118617
1 comments:
Mr. Anderson
April 9, 2025

To view all parameter values on your Solaxx SafeDip:

1. Collect a water sample using the onboard cup.
2. Press the START button to begin measurement.
3. During measurement, the hourglass icon will blink until data stabilizes.
4. Once complete, the check mark icon appears, indicating measurement is done.
5. Display rows will show values for TDS, ORP, temperature, and other parameters.

This answer is automatically generated