CellPath JAY-0100-00A User Manual

Slimline slide drying hotplate

Advertisement

Quick Links

Slimline Slide Drying Hotplate
User Manual
www.CellPath.com | T: +44 (0)1686 611333

Advertisement

Table of Contents
loading

Summary of Contents for CellPath JAY-0100-00A

  • Page 1 Slimline Slide Drying Hotplate User Manual www.CellPath.com | T: +44 (0)1686 611333...
  • Page 2: Table Of Contents

    Specification Location Operating Instructions Cleaning Instructions Calibration and Offset Instructions Setting Controller Offset Parameters Routine Inspection Recommendations Spare Parts Wiring Diagram Troubleshooting Guide Warranty Terms and Conditions Non Warranty Information EC Declaration of Conformity www.CellPath.com | T: +44 (0)1686 611333...
  • Page 3: General Notes

    In the event of this instrument malfunctioning, all specimens within the device should be checked to ensure no harm or damage to the specimen has been caused. Amendments Issue 4: May 2021 www.CellPath.com | T: +44 (0)1686 611333 www.cellpath.com | T: +44 (0)1686 611333...
  • Page 4: User Safety

    Where the mains supply or controls. The instrument requires no regular plug connection differs refer to servicing, but CellPath Ltd do recommend an local regulations or consult an annual inspection, as detailed in the manual electrician.
  • Page 5: Specification

    The Slimline Slide Drying Hotplate features a digital temperature control, This product should be returned to over temperature cut out and miniature CellPath Ltd for disposal and recycling. circuit breakers for safety. Dimensions: Width 183mm Depth 450mm Height 85.5mm...
  • Page 6: Operating Instructions

    D. Exiting the Set mode is achieved by pressing the P key or automatically if no key is pressed for 15 seconds. After that time the display returns to the normal function mode. www.cellpath.com | T: +44 (0)1686 611333...
  • Page 7: Cleaning Instructions

    In the event of a fault, push back in to reset. applied with a soft cloth. If fault situation continues, please contact  your Service Engineer or CellPath Ltd. The heated surface of the hotplate will require cleaning at regular intervals, using a minimal quantity of mild...
  • Page 8: Calibration And Offset Instructions

    ± 1°C (as shown in the image above), then the instrument is within calibration. If the temperatures are not in agreement i.e. a difference greater than ± 1°C, then the instrument will need to be re-calibrated. www.cellpath.com | T: +44 (0)1686 611333...
  • Page 9: Setting Controller Offset Parameters

    (offset) applied to the temperature display. This was the adjustment (offset) which was used when the slide drying hotplate was manufactured to calibrate the instrument. www.cellpath.com | T: +44 (0)1686 611333...
  • Page 10: Routine Inspection Recommendations

    User Manual Routine Inspection Recommendations CellPath Ltd recommend that a simple annual inspection be made for all CellTec laboratory equipment in order that any malfunction can be identified and rectified as early as possible. This is to ensure user safety and prolong instrument life span.
  • Page 11: Wiring Diagram

    Slimline Slide Drying Hotplate Wiring Diagram www.cellpath.com | T: +44 (0)1686 611333...
  • Page 12: Troubleshooting Guide

    4. Temperature of hot plate A: Temperature control A: Instrument should be seems to be fluctuating more circuit fault. checked by a competent than expected. person. (Expected temperature variation ±1°C from the set temperature.) www.cellpath.com | T: +44 (0)1686 611333...
  • Page 13 B: Actual hotplate temperature is lower than the set B: Check the set temperature. temperature C: Instrument should be C: Temperature control circuit checked by a competent fault. person. www.cellpath.com | T: +44 (0)1686 611333...
  • Page 14: Warranty Terms And Conditions

    Any goods or parts thereof, which may require repair or replacement, shall be and Conditions repaired or replaced (at the election of CellPath Ltd) at the works of CellPath Ltd. The product to be repaired shall be ΠCellPath Ltd warrants to the Customer...
  • Page 15: Non Warranty Information

    It is the customers responsibility to ensure that the loan equipment is packed in the packaging provided by CellPath Ltd, in order that CellPath Ltd can arrange collection of the loan instrument. If the loan instrument is not packed and ready for collection within...
  • Page 16 User Manual Notes www.cellpath.com | T: +44 (0)1686 611333...
  • Page 17 Slimline Slide Drying Hotplate Notes www.cellpath.com | T: +44 (0)1686 611333...
  • Page 18 User Manual Notes www.cellpath.com | T: +44 (0)1686 611333...
  • Page 19: Ec Declaration Of Conformity

    Slimline Slide Drying Hotplate EC Declaration of Conformity We herewith confirm the following products: Slimline Slide Drying Hotplate - JAY-0100-00A Conforms with requirements outlined by the following European Directives: Low Voltage Directive 2014/35/EU EMC Directive 2014/30/EU We confirm the declaration:...
  • Page 20 CellPath Ltd, 80 Mochdre Enterprise Park, Newtown, Powys, SY16 4LE, UK T: +44 (0)1686 611333 | E: sales@CellPath.com | CellPath.com...

Table of Contents