4 Follow the measurement instructions in Chapter 4.3.1.2 Operating Mode „Average Value"
5 After the measurement pull the probe out of the material and loosen up the material again by shaking
the bucket.
6 Follow the instructions point 2 to point 4 two more times to obtain in total three averaged
measurement values.
7 Fill the material into another bucket, as then the bottom layers are on top and vice versa. This
procedure is in particular important for material near the saturation point, as free water may build up
at the bottom layers.
Follow the instructions point 2 to point 4 three times to obtain in total another three averaged measur-
ement values.
8 In total six averaged measurements are available now. The representative value for the probe is the
average value out of these six measurements.
23