The shipping components list and actual material received should be compared and any shortages reported to Envirco immediately. Assembly The Hospi-Gard® lsoClean® is shipped completely assembled from the factory and requires no assembly. Every unit is tested prior to shipment to verify airflow and that the unit is leak-free.
Hospi-Gard IsoClean 800 & 400 ® ® HEPA Filtration System I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l ■...
Page 6
The air should not be recirculated into the general ventilation. ln certain cases in existing facilities, recirculation of air into the general ventilation system is unavoidable and, in such cases, CDC guidelines state that HEPA filters should be installed in the exhaust duct leading from the room to the general ventilation system. Innovators in Clean Air Technology | www.envirco.com...
NOTE: IT IS THE RESPONSIBILITY OF THE PURCHASER, AND NOT ENVIRCO, TO ENSURE THE ROOM IN WHICH THE UNIT IS USED IS IN COMPLIANCE WITH CDC GUIDELINES FOR A NEGATIVE PRESSURE ISOLATION ROOM AND MONITORED TO ENSURE ITS CONTINUING PERFORMANCE.
Page 8
Exterior cabinet cleaning must be performed with the unit running. Wipe the exterior of the unit with a damp sponge or cloth, paying particular attention to the prefilter intake grille. Rutala WA, APlC guideline for selection and use of disinfectants, Am J lnfect Control 1996 Aug ; 24(4): 313-42. Innovators in Clean Air Technology | www.envirco.com...
Page 9
Hospi-Gard IsoClean 800 & 400 ® ® HEPA Filtration System I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l Prefilter Replacement The prefilter should be replaced every 60 to 90 days.
Page 10
Remove the six (6) screws fastening the motor ring to the panel and remove the motor assembly from the unit. CAUTION: WHEN REINSTALLING THE MOTOR ENSURE THAT IT DOES NOT TOUCH OR PUNCTURE THE HEPA FILTER. Innovators in Clean Air Technology | www.envirco.com...
Hospi-Gard IsoClean 800 & 400 ® ® HEPA Filtration System I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l Relocating a Contaminated Unit NOTE: USE PROPER PROTOCOL, GOWNING, AND PROTECTIVE MEASURES.
Adjust the speed control down to the next lowest setting. Verify all filters are properly in place. Excessive Contamination Replace HEPA filter. Verify the HEPA filter is installed properly. ■ Wiring Diagrams IsoClean® 800 Wiring Diagram Innovators in Clean Air Technology | www.envirco.com...
Page 13
Hospi-Gard IsoClean 800 & 400 ® ® HEPA Filtration System I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l ■...
Hospi-Gard IsoClean 800 & 400 ® ® HEPA Filtration System I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l ■...
Page 16
Germicidal Lamp 61528 Minihelic Gauge 63360 Keylock 61528 Minihelic Gauge 63360 Keylock For Isoclean 400 units purchased prior to February 2021, contact the factory for verification of correct motor, motor/blower, and capacitor parts. Innovators in Clean Air Technology | www.envirco.com...
Contact Envirco ■ For warranties, replacement parts, technical support, or any other questions please contact Envirco at 1-800-884-0002. For tech support related issues including troubleshooting and warranty, please email techsupport@envirco.com. For customer service related issues inlcuding parts ordering, please email customerservice@envirco.com.