If any damage is discovered, file a claim with the responsible carrier immediately. Assembly Note: The Hospi-Gard AirCeil unit is completely assembled at the factory with the exception of the ¼-inch-20 eyebolts that can be used when hanging the unit from an overhead structure.
The button to the left of the up arrow quickly increases to the airflow index by increments of ten (see Figure 4). • The button to the left of the down arrow quickly lowers to the airflow index by increments of ten. Innovators in Clean Air Technology | www.envirco.com...
Page 6
The green Setting Indicator Light will flash, showing the last programmed airflow setting. The motor will ramp up temporarily and then turn “OFF.” When the AirCeil is not in operation, both the red and green indicator lights will be turned “OFF.” Innovators in Clean Air Technology | www.envirco.com...
Hospi-Gard AirCeil ® ™ HEPA Filtration System I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l Motor Speed or Airflow Adjustment Session Point the Flow-Set Transmitter toward the Infrared Sensor.
10% or 50 CFM greater than the room supply air, whichever is greater. Achieving Negative Pressure Note: It is the responsibility of the purchaser and not ENVIRCO Corporation to ensure the room in which the AirCeil unit is used to create the negative pressure is in compliance with CDC guidelines for a negative pressure isolation room and monitored to ensure its continuing performance.
Hospi-Gard AirCeil ® ™ HEPA Filtration System I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l Difficulty Achieving Negative Pressure Note: The AirCeil has the capacity to exhaust up to 600 ft3/min of air, however the actual capacity will depend on several factors unique to each facility and may include, but shall not be limited to the length and number of bends in the flexible...
Page 10
HEPA filter pleats. Even a light touch on the filter surface may damage the filter. A quantitative leakage and filter performance test should be performed each time the filter is changed. Motor/Blower Lubrication The AirCeil motor/blower comes equipped with sealed for life bearings and requires no lubrication. Innovators in Clean Air Technology | www.envirco.com...
Perform a leakage and filter performance test prior to using the relocated unit. Note: CDC guidelines require that a quantitative leakage and filter performance test should be performed each time the unit is relocated. A list of licensed testing organizations is available on request from ENVIRCO Corporation. Unit Decontamination ■...
With the unit in the Run Mode, depress the Clear button once to ensure that the airflow index is properly set from 1-100. Press the Enter button to accept the setting if correct or reset the index is set too high. For further help with troubleshooting, please contact ENVIRCO Technical Support: ENVIRCO...
Move the aerosol generator hose to all possible leak locations (check the perimeter of the filter seal, the entire filter face and the back panel of the unit). Acceptance: DOP penetration not to exceed 0.03%. Innovators in Clean Air Technology | www.envirco.com...
The obligation of ENVIRCO under this warranty is limited to repair or replacement of any parts that are defective for a period of one year after invoice date, provided that ENVIRCO receives written notice of such defect. For...
Page 18
» MAC-T Ceiling Grid System » VOR Vertical Flow Surgery Isolator ENVIRCO MAC 10 is a registered trademark of ENVIRCO, USA. Specifications subject to modifications or changes without notice. US patents 4,560,395 and 5,470,363. Other patents issued and pending in foreign countries. ENVIRCO ®...