AMPTEC 620VN Explosive Safety Igniter Tester OPERATION/MAINTENANCE MANUAL Preliminary Manual Revision A 10APR2013 AUSTIN, TEXAS 78759 USA CAGE CODE 1CRL2 PHONE +001 (512) 858-4045 or 1-800-350-5105 (from the USA) FAX +001 (512) 3858-4340 or 1-800-430-5440 (from the USA) website http://www.amptec.com...
Page 2
The specification, operation, drawing and schematic sections of this manual contain the unique detail that define the AMPTEC 620VN. Due to continuing product refinement, due to possible parts going obsolete and other component manufacturer changes, AMPTEC RESEARCH reserves the right on rare occasions to change any of its products specifications.
SECTION A: RECEIVING AND INITIAL INSPECTION Introduction to the AMPTEC 620VNS Igniter Tester Receiving, Unpacking, and Initial Inspection Setup and Use SECTION B: 620VN FAILSAFE OHMMETER SPECIFICATIONS Specifications SECTION C: REPLACEMENT, OPTIONAL AND ACCESSORY ITEMS Available Accessories and Options SECTION D: OPERATION, FUNCTION AND USE...
Page 4
R109 and 620VL-070 Current (I source - 2 of 2) Source} 620VN Main PCB Bill Of Materials (components parts list) page 1 of 2 620VN Main PCB Bill Of Materials (components parts list) page 2 of 2 Drawing 620VN-600 620VN Main PCB Assembly (Silkscreen Ref. Design Dwg. ) rev. J...
A-2. Receiving, Unpacking A-1. Introduction to the AMPTEC 620VN and Initial Inspection Some of the devices the 620VN Igniter Tester may be used on include: fuses, squibs, igniters, Should the AMPTEC shipping box appear damaged upon arrival, request that the carrier's explosive bolts, rocket motor squibs, automobile air-bag initiators and many others.
Page 7
Option "403" Option "519S or D" - "S” for a single unit, “D” for dual/side by Kelvin Mini-probe side two unit install. The AMPTEC 620VN Igniter Tester (this Test Lead Set option comes with a flush panel mount faceplate cutout and support tray) can be installed/secured into a standard 19"...
Page 8
RS232C Interface. PCB and the unit’s RS232C interface. When the 620VN main power switch is placed in the Gold Plated Five way power “ON” position, the unit powers up all Input Jacks...
V high and I high command. terminals should be shorted together, and the V low and I low terminals should be shorted. The 620VN Voltage Read - Commands RS232C I/O to output a High, Voltage low, Current high and Current low wires single data string ( 1 resistance reading).
Page 10
These instructions should provide a step-by-step how to operate a AMPTEC 620A series (620ES, 620BH, 620EH, 620VL, 620VN etc.) Explosive Safety Igniter Tester via the optically Isolated RS232C interface. An important note - all manual selected push button front panel ranges must first be de-selected (popped out) prior to any RS232C remote operation.
Page 11
6. If the commands above were entered and the correct prompt was not received, then type “V”. The correct prompt from typing “V” is “620VN version 1.0 (C)opyright Amptec Research 3/16/2013”. If this prompt was not received, then the firmware of the unit is out of date and must be updated. If the correct prompt was received, but the commands in bullet 5 did not yield the correct prompt, then there is a defect in the unit.
D-4. 4-Wire Resistance Measurement Jacks. All AMPTEC 620VN Kelvin leads have In many resistance measurement applications the 4 wire current high, current low as well as the contact resistance and can exceed the value the voltage sense high and voltage sense low of the test resistance by several orders of magni- routed to the end of the leads.
Page 13
Sect. D - Operation, Functional Self-Test and Use By using a 4-wire Kelvin type lead set or by shorting the V and I terminals together and and I terminals together the instrument is in the 2 wire resistance mode. Resistance Offset in 2 Wire mode All wiring including harness wires from the two wire test connection out, that are "in series"...
Troubleshooting Explosive Safety Igniter Tester, servicing personnel should become familiar with this Since the 620VN Tester is used to test potential section to gain a complete understanding of the deadly explosive force detonators and warheads internal workings of the ohmmeter.
Page 15
- i.e. connect to an earth ground to avoid a static discharge to a static sensitive component. Handle all 620VN internal components as if they are static sensitive if you are not sure.
Sect. E - General Operation and Design E-4. Analog to Digital Conversion E-3. Circuit Descriptions The A to D conversion is done with a ICL8068 /ICL71C03 chip set. The ICL8068 takes care of The circuit descriptions which follow are the analog part and the ICL71C03 takes care of referenced to Figures E-1, E-2, E-3 and the the digital part of the 4 ½...
Page 17
Sect. E - General Operation and Design R INT C INT BUFFER BUFFER INTEGRATOR COMPARATOR 1 µF ZERO CROSSING DETECTOR C REF C AZ C STRAY V IN FIGURE E-2A. PHASE l AUTO-ZERO C INT R INT BUFFER BUFFER INTEGRATOR COMPARATOR 1 µF ZERO CROSSING...
Sect. E - General Operation and Design E-4-1. Reference Voltage Detailed Description The precision reference voltage required to do the A/D conversion is developed by IC201. The Analog Section zener voltage is attenuated to approximately - 0.5V . This voltage is applied to IC2-7 . Figures E2 diagrams A thru D shows the equivalent circuit of the analog section in 3 E-4-2.
Page 19
IC201 is equal to the non-inverting input of IC202. Again, this condition occurs when the voltage drops across R212 and R222 620VN Failsafe Current Calculation - Worst are equal to the voltage drops across R213 and Case Component Failure R221.
Page 20
The 620VN Tester measurement circuitry is also failsafe current limited, even under worst case component failure. For the 620VN Tester the normal or typical operating current level is less than 5 mA, and <8mA on as a Failsafe Level .
Page 21
The locations of the continued instrument accuracy or following a adjustments are shown on drawing number main PCB electronic component 620VN -600 at the back of this manual. repair/replacement. The recommended calibration interval is 1 year. F-3-1. Zero Offset Adjustment All AMPTEC 620 series Igniter Testers are four 1.
Page 22
620VN Meters fitted with "OPT 247" Optically Isolated Continuous operating power The 620VN Igniter Tester display has a LOW BATTERY LED indicator (dot) just to the left of the meter's main 4 ½ digit numeric display. The LED dot when lit is used as a Low Battery indicator. This status is also available over the meter's optically isolated RS232C serial interface.