1. Introduction
The content of this PERFORM Operating Document (POD) provides guidelines for use
and the cleaning procedure for the ImpediMed SFB7 bio-impedance analysis unit. This
unit assesses body composition (i.e. fat-free mass, fat mass, intra-/extracellular fluid) of
participants. This POD applies to all users of the unit.
PC-POD-CP-010-v02
PERFORM Operating Document
PC-POD-CP-010-v02
Printed copies are not controlled.
PC-POD-CP-010-v02
Date
May 7, 2020
Page 1 of 10

Advertisement

Table of Contents
loading

Summary of Contents for ImpediMed SFB7

  • Page 1 1. Introduction The content of this PERFORM Operating Document (POD) provides guidelines for use and the cleaning procedure for the ImpediMed SFB7 bio-impedance analysis unit. This unit assesses body composition (i.e. fat-free mass, fat mass, intra-/extracellular fluid) of participants. This POD applies to all users of the unit.
  • Page 2: Unit Components

    PC-POD-CP-010-v02 2. Unit Components 1= BIA electrodes 2= Lead set 3= Alligator clips (4) 4= Test cell for calibration check 5= Power module for recharging the battery 6= Main power cable 7= Crossover network cable 8= CD-ROM 9= Stylus pen 10= Microfiber cloth 3.
  • Page 3: Contraindications For Use

    PC-POD-CP-010-v02 This mode measures bioimpedance parameters over a frequency range of 4-1000 kHz with 256 data points defining the BIS mode as a true BIS. On-screen graphs display the measured data in the form a cole-cole plot. In addition, the characteristic frequency for the subject is determined as well as total cell membrane capacitance.
  • Page 4: Preliminary Instructions

    PC-POD-CP-010-v02 6. Preliminary Instructions Participant should respect the following instructions in order to minimize changes in the body’s hydration status (as this can over or underestimate values): Excessive exercise should be avoided 2 hours prior to measurement to avoid changes in vascular perfusion, temperature, cutaneous blood flow, vasodilation, and fluid losses.
  • Page 5 NOTE: Ensure that the metallic part of clip is in direct contact with the conductor side (underside) of the electrode tab and that the clips are aligned to the centers of the electrode tabs. The ImpediMed SFB7 should be unplugged from the recharging unit before use to avoid possible noise contamination of the measurement.
  • Page 6: Data Retrieval

    2. Your computers operating system will automatically detect and install the drivers for the adapter. 3. Restart your computer to allow the adapter to be fully installed. 4. Once your computer has restarted insert your ImpediMed software CD-Rom into your computer’s CD-Rom drive. PC-POD-CP-010-v02 Printed copies are not controlled.
  • Page 7: Safety And Cleaning

    NOTE: BIS and SFBI measurement data is stored on the unit, but only BIS measurement data is retrievable via Ethernet. NOTE: Uploaded files are automatically deleted from the Imp SFB7. Sophisticated file transfer ensures that each file is transferred, confirmed as received, before deletion.
  • Page 8 Do not plug the unit leads to any mains power outlet/point. iv. Do not use or operate unit in the presence of strong electromagnetic fields. v. The ImpediMed SFB7 has yet to be clinically validated for use on pregnant patients. vi. Only use ImpediMed electrodes.
  • Page 9: Troubleshooting

    5. Do not wind the leads tightly, crinkle, or twist the leads, as this may cause the fine wires inside to break. 6. If the leads appear damaged, contact ImpediMed or a licensed distributor for replacements. See website information https://www.impedimed.com/products/sfb7/ iii.
  • Page 10 Problem Possible Cause Solution 1. Attach external power pack and try again. No display when SFB7 is 1. Battery power is low. 2. If external power pack is attached, switched ON. 2.No external power. check that power pack is plugged in and switched on.