Laser powered electric field probe with emsense 10/40 plug-in card for emcenter (46 pages)
Summary of Contents for ESCO Technologies Labculture LB2 Series
Page 1
User Manual Esco Technologies, Inc. 2940 Turnpike Drive, Units 15-16 • Hatboro, PA 19040, USA Toll-Free USA and Canada 1-877-479-3726 Tel 215-441-9661 • Fax 215-441-9660 us.escoglobal.com • usa@escoglobal.com Esco Micro Pte. Ltd. 21 Changi South Street 1 • Singapore 486 777 Tel +65 6542 0833 •...
Table of Contents INTRODUCTORY PAGES Table of Contents Warranty Terms and Conditions Introduction 1. Products Covered 2. Safety Warning 3. Limitation of Liability 4. European Union Directives on WEEE and RoHS Declaration of Conformity USER SECTION Chapter 1 – Product Information 1.1 Quick View Chapter 2 –...
Warranty Terms and Conditions Esco products come with either a 1, 2 or 3 year limited warranty, depending on the product purchased, beginning on the date of shipment from any Esco international warehousing location. To determine which warranty applies to your product, refer to the appendix below. Esco's limited warranty covers defects in materials and workmanship.
Page 6
These Terms and Conditions shall be governed by and construed in accordance with the laws of Singapore and shall be subject to the exclusive jurisdiction of the courts of Singapore. Technical Support, Warranty Service Contacts USA: 1-877-479-3726 Singapore: +65 65420833 Global Email Helpdesk: support@escoglobal.com Visit http://www.escoglobal.com/ to talk to a Live Support Representative Distributors are encouraged to visit the Distributor Intranet for self-help materials.
Page 9
Declaration of Conformation In accordance to EN ISO/IEC 17050-1:2010 Esco Micro Pte. Ltd. 21 Changi South Street 1 Singapore, 486777 Tel: +65 6542 0833 Fax: +65 6542 6920 declare on our sole responsibility that the product: Category : Class II Type B2 Biological Safety Cabinet Brand : Labculture Model...
Chapter 1 - Product Information 1.1 Quick View 1. Esco Sentinel Control System 3. Interior Work Area 2. Sash Window 4. Armrest Class II, Type B2 Biological Safety Cabinets...
Chapter 2 – Sentinel Control System 2.1 Sentinel Control System Menu Socket Display Button Button Button Button Set/Mute Down Lamp Visual Button Button Button Button Alarm 1. Fan Button Turns on and turn off the fan. 2. Lamp Button Turns on and turn off the fluorescent lamps. 3.
2.2 Menu Options Please refer to the following diagram for complete reference to all menu options available. SET TIME HH:MM UV TIMER HH:MM MEASUREMENT UNIT METRIC | IMPERIAL SETTINGS SET EXP. TIMER HH:MM:SS DATA LOG PERIOD HH:MM:SS STOPWATCH TIMER EXPERIMENT TIMER NORMAL MODE SET MODE QUICKSTART MODE...
2.2.1.3 Measurement Unit Using this option, the user can select the unit in which air velocity is measured and displayed. The user can choose between metric (m/s) and imperial (fpm). METRIC MENU SETTINGS MEASUREMENT UNIT IMPERIAL 2.2.1.4 Set Experiment Timer Experiment timer is a countdown timer that can be used for critical experiment.
2.2.3 Field Calibration The purpose of calibration is to ensure the accuracy of the airflow display and alarm (if present). This involves measuring airflow with reference instrumentation and establishing reference between airflow sensor(s) on the BSC to the standard reference. Calibration should only be carried out by trained personnel. This section is presents a brief overview of the calibration menu function.
Page 17
2.2.4.3 Airflow Monitor Whenever the air velocity falls below the fail point, the air fail alarm will be triggered. This option is used to enable/disable alarm. The alarm is enabled by default. If the ambient temperature is outside of 18-30 C (which is the cabinet working temperature), the Airflow Monitor is automatically disabled ENABLED...
2.2.4.10 Set Sash Cycle (Only for Motorized BSC) To reset the sash cycles count to zero or to input the sash cycle count manually. The maximum of value of sash cycles is 16000. The cycles will raise every sash move to up and down. Warning message will be showed after the cycle value reach 15000.
2.5 Diagnostic Mode Diagnostic mode can be accessed by pressing the SET button. The diagnostic mode allows the user to know the condition of the BSC or help the service engineer during maintenance and troubleshooting. On Screen Explanation MODE Shows which mode is active: NORMAL MODE, QUICKSTART MODE or MAINTENANCE MODE VERSION Shows the version of the firmware TEMPERATURE...
Chapter 3 – Basic Cabinet Operation 3.1 Sash Window Operation 3.1.1 Sash Window State Sash is fully open Sash is in safe position Sash is fully closed Blower can be active Blower can be activated Blower can’t be activated Fluorescent lights can be used Fluorescent lights can be used Fluorescent lights can’t be used Unsafe working condition...
3.1.3 Using Sash Window The sash window should be fully closed when the cabinet is not in use. This helps keep the work zone interior clean. The sash window should always be in the normal operating height at all times when the cabinet is in use.
Minimize room activity (personnel movements, closing and opening of doors, etc.) since these external airflow disturbances may adversely affect the BSC’s internal airflow, thereby possibly impairing the containment capabilities of the BSC. Ensure that the sash is at normal operating height (READY state) before starting any experiment. ...
3.5 UV Lamps The UV light is germicide and virucide. Even at the minimum acceptable irradiance in a BSC – 40 μW/cm Department of Health and Human Services et. al., 2000), it would only take 12.5 minutes to reach 30,000 μW/cm2 (1 W = 1 J/sec), which has been listed as germicidal for spore forming organism.
Chapter 4 – Maintenance 4.1 Scheduled Maintenance Proper and timely maintenance is crucial for trouble free functioning of any device and your Esco BSC is no exception to this rule. We strongly recommend that you follow the maintenance schedule suggested hereunder in order to obtain optimal performance from your Esco BSC.
Page 28
Accessing the paper catch The purpose of accessing the paper catch is to remove any retained materials that might cause obstructions. Care must be taken as the area is contaminated. Before opening the paper catch: Use appropriate personal protective equipment (PPE). ...
Sash cleaning procedure for motorized sash BSC 1. Remove armrest 2. Press MENU button, select SET MODE then choose MAINTENANCE MODE - this mode allows you to lower the sash further than usual 3. Lower the sash by pressing the DOWN button continuously until the sash is in the lowest position and create a 10 cm opening between the sash and the...
Page 33
LOG RECORD Cabinet : _______________________________________________________________ Serial Number : _______________________________________________________________ Person in Charge : _______________________________________________________________ 1. This log record should be used by the operator to record any new agent/virus/bacteria/germs that has been introduced to the cabinet during its operation, problems encountered, etc. 2.