Quick Links

HIGH VACUUM CORRESPONDENCE
SMALL SIZE PRESSURE GAUGE
PG-35L
  marking
(Compliance with EMC Standards)
INSTRUCTION MANUAL   
①The type of pressure media for PG-35L should be used liquid and gas that don't corrode for SUS316L.
②For stability, use a regulated direct current power supply.
 Surge  absorbing  devices(diodes,varistors,  etc.)are  necessary  if  inductive  loads  such  as  relays  or  solenoids  are  connected  to  the  same 
circuit as the PG-35L.
 If using a DC power supply unit such as a switching power supply, the FG terminal should be earthed. Do not wire in parallel to high tension 
cables or power lines, or use cable ducts which contain high tension cables or power lines.
③Be careful not to crimp any wires during handling, or put any pressure on the display area of the main body while assembling piping.
④Use pH neutral detergents to clean the body. Do not use solvents such as thinners.
⑤This product is dust proof and drip proof(IP65 of IEC standards)and is not suitable for use in environments requiring higher standards.
 Also, do not use this product in an environment with a possibility of product being covered by liquids other then water(Such as oil, solvent, 
and etc.)and outdoor.
⑥Do not use pointed objects such as pens to press the setting buttons on the display panel, as this may push holes in the setting buttons and 
damage them.
⑦Do not insert wires, etc. in the pressure port, as this may damage the internal diaphragm and cause malfunctioning.
⑧gasket type:
 Do not touch or scratch the edge of the fitting, as this may damage the sealing and cause leakage.
⑨The  PG-35L  do  not  have  an  explosion  proof  structure.  Do  not  use  it  for  the 
detection of flammable gases.
⑩When  analog  output  is  supplied  to  a  noise-sensitive  device,  a  low-pass  filter  is 
requested in a customer's circuit.
⑪Countermeasures for noise interference:
 Please  connect  either  the  shield  wire  or  the  metal  part  of  the  product  to  frame 
ground(FG)of the power source.
⑫In case a wire extension is needed, please use a shielded wire.
Ver.2.2
Important Information and Warnings
Thank you for purchasing a 
NIDEC COPAL ELECTRONICS CORP. product.
In order to use the product correctly and most 
appropriately, please completely read this manual before 
use and keep it for future reference.
For more detailed information please ask for the nearest
distributor or the following sales center.
Nishi-Shinjuku Prime Square., 7-5-25  
Nishi-Shinjuku Shinjuku-ku Tokyo 160-0023, Japan
https://www.nidec-copal-electronics.com
Power supply unit
Surge absorbing circuit
AC input
DC output
F G
F G←The shield wire
+ V
PG35L
1

Advertisement

loading

Summary of Contents for Nidec PG-35L

  • Page 1 ⑥Do not use pointed objects such as pens to press the setting buttons on the display panel, as this may push holes in the setting buttons and  damage them. ⑦Do not insert wires, etc. in the pressure port, as this may damage the internal diaphragm and cause malfunctioning. ⑧gasket type:  Do not touch or scratch the edge of the fitting, as this may damage the sealing and cause leakage. ⑨The  PG-35L  do  not  have  an  explosion  proof  structure.  Do  not  use  it  for  the  detection of flammable gases. Power supply unit ⑩When  analog  output  is  supplied  to  a  noise-sensitive  device,  a  low-pass  filter  is  Surge absorbing circuit + V requested in a customer's circuit.
  • Page 2: Specifications

    Specifications PG-35L Model 102R 103R Type Gauge pressure Rated pressure range −100〜100kPa −100〜1000kPa Maximum pressure 200kPa 1500kPa Break-down pressure 300kPa 2000kPa Acceptable media Liquids or gases that do not corrode SUS316L 10.8〜30VDC(including ripple) Operating voltage Current consumption 50 mA maximum Two outputs NPN/PNP:Transistor open collector Switch rating:  30VDC100mA maximum Residual voltage: 1.2V maximum(NPN)/2.2V maximum(PNP)at 100mA. Hysteresis 0〜300 counts setting(adjustable) Switch Repeatability ±0.2%FS±1digit outputs Response Approx.5ms(Digital Fillter:“F-0”) Short circuit protection Exists Output voltage 1〜5V /Pin ( L)〜Pin ( H), Output impedance:10kΩ , Resolution:1/204 Analog output...
  • Page 3: Function Names

    Function Names Display unit − LED (red) Full 3 digit LED     Mode button SW1 LED (green) +/−     Down button SW2 LED (red)     Up button C O P A L E L E C T R O N I C S Operating Procedures Set magnification Setting1 (eleven selections) Initial Pressure Setting2 Operations Set analog output Settings Settings Hysteresis Mode  ...
  • Page 4 ■Analog Output Setting The analog output setting is determined by the value of the second digit :  the green SW1 LED should be flashing during the setting. −Pr           0           +Pr Mode (Vzero) R mode ( Compound pressure output) G mode ( Positive pressure output) V mode  (Negative pressure output) An example of setting the R mode in the 102R range. In the Operations Mode, press the Press the  button until the Set the value of the second Press the  button for more    and  buttons simultaneously to SW1 LED under the second digit to “1” using the  and than one second to return to  enter the Initial Settings Mode. digit flashes.   buttons. the Operations Mode. Notes:V mode for 102R are set prior to shipment.R/G mode for 103R can only be selected. ■Switch Output Setting The switch output setting is determined by the value of the first digit :  the red SW2 LED should be flashing during the setting. Output SW1 output SW2 output There are four operation modes. These are shown in the diagrams below. Mode Separate Window comparator Separate Window comparator Separate Mode Window Comparator Mode Operation...
  • Page 5: Troubleshooting

    ■Hysteresis setting To set hysteresis the LED should be flashing.  “−” An example of setting a hysteresis value of 8.0kPa with the 102R (kPa) . Press the  button for  Set the value to “8.0” Press the  button until  more than one second to  Press the  and  buttons. using the  and   the −LED flashes. return to the Operations  buttons. Mode. Note:20digits is set prior to delivery. ■Digital Filter setting To set Digital Filter all LED should be not flashing. An example of setting 2.5s response time by Digital Filter. Press the  button for  Set the value to “F-3” Press the  button until  more than one second to  Press the  and  buttons. using the  and   all LED do not flashe. return to the Operations  buttons. Mode. Note: “F-0”is set prior to delivery. Troubleshooting ■If the following error messages are displayed, follow the procedures in the table. Display and problem Cause Solution Turn off the power and verify the load  Output current is exceeding 100mA. connected switch output 1 and 2. Pressure was applied at the zero point ...
  • Page 6 ■Non-Display Mode <Non-Display[Temporary]Mode> ・When the keys are not operated for more than 10 seconds during Operation Mode, the system will automatically select Non-Display [ Temporary]Mode and  the display will turn off. ・Decimal point LED shown in the figure below will blink during Non-Display[Temporary]Mode. ・Using the EEPROM, the PG-35L can retain preset values even if the power is turned off. ・If an error message is detected, the display will comeback and show the error message. ・You can change any functions during Non-Display[Temporary]Mode. (How to set) Blink +/− +/− +/− +/− C O P A L E L E C T R O N I C S C O P A L E L E C T R O N I C S...
  • Page 7: Model Numbers

    Fitting part:G2 type Fitting part:VC type COPAL ELECTRONICS COPAL ELECTRONICS G1/4 ( PF1/4) 9/16-18UNF (M5 screws included) Brackets(Option) ■PG-35L Holder cover set  (sold separately) (□42.6) (5.7) □40 +0.5 24.7 COPAL ELECTRONICS panel cut dimensions  (Panel thickness: 1〜3.6mm) Panel holder cover Panel holder Main body ■Accessories  (Sold separately) Product name Model no. Description Applicable model Holder cover set PG-30/35 Panel holder cover,  ACPG-004 (for protection of  panel holder PG-35L gauge sides) (Note)Since this product contains small components,  please handle this product carefully. ...