Ventilation Beyond Limits fabian +nCPAP evolution and Auxiliary Systems Instructions for Use, Software Version 5.1.x fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 3
Disclaimer Disclaimer ACUTRONIC Medical Systems assumes no responsibility for the use or reliability of its software on equipment that is not furnished by ACUTRONIC Medical Systems. ACUTRONIC Medical Systems makes no warranty of any kind with regard to software applications that are created by the user.
Page 7
20 Electromagnetic compatibility statement ........129 20.1 Devices with SN from AN-04000 ................ 129 20.2 Devices with SN up to AN-03999 ............... 133 Appendix A: Glossary.................... 137 Appendix B: Index ....................139 fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Notes provide additional information to clarify an explanation or instruction. 1.3 Applicable product versions This Instructions for Use is applicable for fabian +nCPAP evolution devices running software version 5.1.x, where (x) can be any number. fabian +nCPAP evolution | SW V5.1.x...
DO NOT use when patient is connected to the device Expiration date Explosion Hazard warning External power supply input Flammability / Fire hazard warning. Flow sensor connection Sensor Flow sensor connection. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 11
Non-Sterile NOTE symbol Nurse Call signal output. Nurse Call signal output. Potential equalisation connection. Protective Earth ground. Single use. Storage Temperature This way UP. Type BF applied part. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 12
1 Introduction Symbol Description Type B applied part Unplug power before opening housing. Video Output USB connection. Warning regarding operation in explosive areas. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
WARNING: Connect only electrical devices which are IEC60601-1 approved. WARNING: NEVER operate the ventilator while covered or set up in a way to negatively impact the operation or function. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 14
WARNING: DO NOT modify the equipment. WARNING: Before applying non-original accessories, ensure that they are biocompatible. All accessories supplied by ACUTRONIC Medical Systems for use on fabian ventilators are biocompatible. WARNING: When connected to a patient DO NOT simultaneously touch the external power supply cord and the flow sensor connector cable.
Page 15
WARNING: DO NOT use the ventilator in association with HF (High Frequency) electrosurgical equipment. WARNING: Connect SpO and etCO sensor cables to the machine before the patient is connected. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
2 Warnings, cautions and notices 2.2 Maintenance The fabian +nCPAP evolution device is a ventilator classified as Class IIb (for prolonged use more than 24 hours and less than 30 days) according to the European Medical Device Directive, as such: •...
2 Warnings, cautions and notices 2.4 Intended use The fabian +nCPAP evolution is intended for premature infants, new-borns as well as children weighing up to 30 kg. The fabian +nCPAP evolution is intended for “in-patient use” in professional healthcare environment including intra-hospital patient transport.
3 System overview 3 System overview 3.1 Scope of delivery The fabian +nCPAP evolution product includes the following items: • One fabian +nCPAP evolution Ventilator • One Accessory kit: • One Flow Sensor (reusable) • One Flow Sensor Cable •...
Connection for future use, covered with a faceplate Oxygen sensor Connection for future use, covered with a faceplate Connection for oxygen “O ” supply 2.0 – 6.5 bar / max 40 L/min fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Ethernet : N/A (Disabled) Flow Sensor : Maximum load is one Flow sensor (Only ACUTRONIC Medical Systems Flow sensors can connect) Max. Voltage: 5 V Max operating current: 300 mA per hotwire -> 600 mA total. Max Voltage: 3.3 V...
Waves / Loops …switches to wave-/loop display. Home … displays the main screen for selecting the respiration mode. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Subsequent alarms with higher priority are only displayed optically during muting. 4.1.2 Rotary push knob The Rotary Push knob combines a push button with a rotary knob for executing various settings, selections and confirmation options. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Selection or display of the Operating mode Depending on the Display mode selected, individual fields can be shown / hidden. Display field options will be described later in section “4.3: Ventilation menu”. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
A blinking Information bar also alerts the operator of active alarms. If screen lock is in use, the symbol for Locked Screen is displayed. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
To confirm the setting, either touch the Parameter button again, or push Rotary Push knob again. To automatically set the alarm limits, select parameter and touch the AUTOSET button. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 27
The respiratory rate and minute volume monitors are calculated as averages using an update period of 6 seconds. In the event that the monitored breath rate falls below 11 bpm, these values are not reported. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
This LED illuminates Green in Mains operation and switches OFF in battery operation. When powered by the external 24 VDC connector, the Mains LED and Battery LED will NOT illuminate. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Ventilation menu indicates the configuration parameters (4) and measurements (5) relevant in the mode. The Ventilation mode may be Figure 4-8: Operation screen changed in the menu. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Press any blank area or another parameter to NOT accept the parameter and keep the previous value Alternating alarms active at the time the mode is switched, will automatically be reset. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
“Notice signal” will sound. Figure 4-11: Parameter locked The high value must be confirmed by pressing the Rotary push knob again. (see the following table) fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
The Graphics menu can be accessed from the Waves button on the Keypad. The Graphics menu The following graphics are available in the Graphics menu: Waves Figure 4-12: Waves Loops Figure 4-13: Loops Trends Figure 4-14: Trends fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Press the key again to continue updating the wave data. Figure 4-16: Freeze Waves Depending on the Ventilation mode selected, the numeric block will display the parameters that are relevant in this mode fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Switch to loops display: Freeze current loop. Current loop is shown in Blue, the saved loop is shown in Black. Figure 4-19: Compare saved Loops & current loops fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
A confirmation message will always be displayed before Trend Data is deleted. Figure 4-20: Trends screen Parameter selection for graphic waves Graphics Parameter values Timeline Adjust timeline (max. 5 days – min. 30 minutes) Update data (refresh) fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
CGA/ISO standards, including fill indicators. The inputs are coded to prevent wrong connection. Medical grade oxygen and medical grade AIR (dust-free, oil-free and dry, 2.0 bar to 6.0 bar) are required. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Figure 5-2: Membrane holder Pay attention to the correct installation position of the Expiration Membrane. The labeling must be legible upon installation. (See illustration). Figure 5-3: Expiration membrane fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Connect the machine end of the Proximal Pressure tubing to the Proximal Pressure port (Prox) on the ventilator. Connect the Inspiratory and Expiratory limb heater connectors with the corresponding connectors on the humidifier chamber heater base. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Place temperature probe outside the incubator and use Incubator extension line. If the patient is in the open warmer unit, the following setup must be used: fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
If the Humidifier does NOT feature an Inspiratory line heater, Water Traps must be installed in the Inspiratory and Expiratory Limbs. Figure 5-5: Dual limb circuit with 2 water traps fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 42
When using an Inspiratory heater, a Water Trap must be installed at the Expiratory end. Figure 5-6: Single heated limb circuit with 1 water trap Ensure that the Flow-sensor IS NOT exposed to excessive Humidification. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Preparing the nCPAP InfantFlow/InfantFlowLP/Inspire Tubing System • Connect Flow sensor to exhalation tube as shown in the picture. Recommendation: • Make the Exhalation tube as short as possible. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
The O will automatically be calibrated after leaving the Calibration menu and repeated once every 24 hours. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 45
For more information on SpO module, refer to the following sections in this manual: section 12.3: “SpO2 module” All audible alarms are silenced for 2 Minutes after startup of fabian fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Confirm that the audible tone can be test. heard Self-test successful. Calibration Calibrate Flow sensor. Calibration successful Calibrate O sensor (occurs automatically after leaving Calibration menu). fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 47
Fill level Water fill level sufficient. Humidifier Function Function OK CAUTION: Ensure that Pmanual and other pressures are reduced to safe levels prior to connecting to patient. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
When the ventilation has stopped the Information bar will display the following message “Ventilation stopped”. NOTE: The pause time will count down the two minutes and then the ventilator will resume ventilation fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
OFF, a notice tone will sound. Press the Alarm Silence button to confirm system switch OFF. Hold the button for at least three seconds until the WARNING LED stops. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
After the device has powered OFF, a notice tone will sound. Press the Alarm Silence button to confirm system switch OFF. Hold the button for at least three seconds until the WARNING LED stops. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
“Open menu” push button. Figure 6-1: Configurations menu Calibration (Flow & O2) Date / Time Display Tools Ventilation Information Patient Data PDMS (optional) Language Service Mode fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Calibration menu is displayed as follows. Figure 6-2: Calibration menu From the Calibration screen, you can calibrate the following Sensors: • Flow sensors: Neonatal and Pediatric • sensor • module fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
When centered, these lines will turn Green. Following successful Calibration, the Date of the last calibration will be displayed in the corresponding area. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 54
Flow Sensor Leakage (loose flow insert or leak) • Flow Sensor Contamination (single-use sensor recommended) • Rain-out (contact ACUTRONIC Medical Systems sales for recommended Patient Circuits and Humidifier to prevent Rain-out) fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
After a successful calibration, the Date of the last calibration is displayed in the corresponding area. Automatic Sensor Calibration runs every time the equipment Restarts and every 24 hours. The gas concentration to patient is not altered during this calibration procedures. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Rotary Push knob is pressed. Automatic lock activated / deactivated. Touchscreen automatically locks after a set time. Display Graphs and trends as filled or regular lines. Autoscale ON / OFF fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Return button and the chosen time will be displayed. Figure 6-6: Lock screen time settings The touch screen can also be locked by holding the Rotary push knob for 3 seconds fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Scaling can be performed manually with the (− / ) keys on the keypad for the timeline. For Wave height, use the Manual inspiration and Flush keys on the keypad fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Factory Settings optional: The “factory defaults” button is only active if it is enabled in service mode. Resets the device to the factory defaults. Pressure Rise I-Flow or Rise-Time fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 60
Use BTB for Vt Select if tidal volume measurement should be updated for every breath. WARNING: Only Reset to hospital and factory settings when the device is NOT connected to patients. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
For data input, just push Change data and choose the field for input For capital letters, push Accepting data with “Apply” Figure 6-11: Data entry keyboard fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
After the time or date has been changed, the confirmation check mark changes colors. Figure 6-13: Change Date & time screen The new information is applied after confirmation. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
SW and HW Configuration with licenses. Only allowed when patient is NOT connected. 6.9 Information Displays the Information screen (System Information) containing Equipment data. Figure 6-15: Information screen fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
After using service mode and before connecting fabian to a patient, restart Figure 6-16: Service mode password screen the device. Only allowed when patient is NOT connected. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 66
After the Alarm Limit has been adjusted or there are no more active alarms, the bell symbol will turn Grey. The Visual Alarm message must be reset by pressing the Alarm Silence button. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
In case alarm list or alarm log files reach their limit (100 / ca. 70000 entries respectively), the oldest entries are deleted first. 7.3 Nurse call relay The delay from the nurse call relay is < 1 second. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Contact your local checksum System failure System error ACUTRONIC Medical conductor PIC Systems Distributor Contact your local checksum System failure System error ACUTRONIC Medical monitor PIC Systems Distributor fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 70
Remaining battery Immediately switch to time < 1 min. Depleted Electrical Mains supply or Ventilation can no battery! power Access alternative longer continue on ventilation battery power fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 71
Oxygen sensor sensor defective. Replace O sensor. Sensor alarms defect Replace O sensor as Oxygen sensor quickly as possible. sensor worn. Sensor alarms used up Calibration still available. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
If the device software encounters an unexpected ERROR, a window appears in the middle of the screen with the following text in English: “Application fabian.exe encountered a serious Error and must shut down.” Figure 7-4: Application error displayed fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
In case a WDST 4/16/32/64 error is resolved (only a temporary error) WDST#8 will be triggered and displayed and the device can continue to be used. NOTE: WDST#8 is not an alarm and only shows that all watchdogs are working correctly. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
When storing the ventilator for long term storage, please remove the Battery pack. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
After 60 minutes on Battery power, the Battery is drained, and fabian will power OFF. Depleted Battery – Connect to Main – – Charge battery for approx. 5 hours – fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
External power connection on the back panel. 1/2 minus 3/4 plus • NO charging when connected to 12VDC external power supply • NO monitoring of remaining time on external power supply fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 82
I-Flow or Rise Time • SIPPV Control Will allow for use of either I-flow or Rise time based pressure • SIMV rise control during inspiration • SIMV + PSV • fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 83
Volume Guarantee/ Volume Limit Guarantee/ • SIPPV Volume Limit • SIMV Will allow either volume guarantee or volume limit to be • SIMV + PSV enabled in the user interface. • fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
A higher I-flow can be requested during NIPPV mode, because of potential higher leakages. See Section 10: “Ventilation parameters” See Section 11.9 : “Ventilation Additives” fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
To prevent Auto-Triggering, another breath cannot be triggered for 180 ms following the end of the preceding breath. See Section 10: “Ventilation parameters” See Section 11.9: “Ventilation Additives” fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
TI and TE frequency. The synchronization window for the next mechanical breath is maximum ½ See Section 10: “Ventilation parameters” See Section 11.9: “Ventilation Additives” fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
SIMV+PSV, the ventilation mode is automatically ch anged to SIMV and can be switched back only manually after the activation of the flow sensor. See Section 10: “Ventilation parameters” See Section 11.9: “Ventilation Additives” fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
OFF the ventilator and remove the flow sensor from the Y-piece of the dual limbed breathing circuit and connect the proper interface with a 15-millimeter adaptor. See Section 10: “Ventilation parameters” See Section 11.9: “Ventilation Additives” fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
11.9: “Ventilation Additives” If Inspiratory Time is set lower than 1 second; the maximum Pressure may NOT be reached. Depending on the Tubing Set, nCPAP Generator and Humidifier. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
The display will show the message: “Volume Limited”. In the event of Flow Measurement Outage (for example: changing the sensor) “V-limit” will automatically be deactivated. After Flow Measurement is available again “V-limit” is automatically reactivated. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
NOTE: Manual Breath is NOT available in O Therapy mode. A Manual breath however can lead to lung over distension. If activated, some clinicians use this for a manual sustained lung inflation after suctioning. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Flush button in a Ventilation mode to initiate the O Flush. The Flush Concentration can now be set with the Rotary push knob. Pressing the O Flush button again will end the Flush immediately. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
12.1 Accessories List WARNING: The items shown in this list have been approved for use with the fabian +nCPAP evolution ventilator. Use of non-approved items is NOT recommended and will NOT be recognized or supported by the manufacturer. If a system malfunctions with non-approved parts and / or accessories, the user is responsible and liable for any and all issues associated with the system malfunction.
Page 95
NIV generator – Vyaire InfantFlow LP For Invasive modalities, only Altera breathing circuits (REF 154308) are validated, for nCPAP modalities only Medijet, InfantFlow, InfantFlow generators and associated sets are validated. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
12.2 Spare parts list Description Membrane holder, 22mm OD / 15mm ID IMPORTANT: There is a risk of damage due to nebulizing dugs or aggressive cleaning solutions. Membrane, expiration Oxygen sensor fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
In addition, the following general messages for the sensor state can appear in the information bar in the Configurations menu: • "checking " • "processing active " fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
This mode is clinically applicable during procedures when detecting rapid changes in oxygen saturation is paramount such as induction, intubation, and sleep studies. Figure 12-4: Fast SAT fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
There are different averaging times that can be selected (2 to 4, 4 to 6, 8, 10, 12, 14, 16). Figure 12-6: SpO averaging time fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
USB device port on the rear panel of the fabian ventilator. The total length of the USB HPL/MIN, patient cable, and sensor cannot exceed 5 meters. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 101
7-pin LEMO device port on the rear panel of the fabian ventilator. The total length of the LEMO HPL/MIN, patient cable, and sensor cannot exceed 5 meters. Figure 12-9: LEMO HPL/MIN module fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Graphics menu. PRICO will adjust the FiO , to try to keep the SpO within the selected range. Figure 12-10: PRICO in fabian Therapy & fabian +nCPAP evolution devices PRICO has the following parameters: PRICO Parameters Min FiO...
Page 103
If FiO is in upper half decrease FiO by 1%. • adjustments are made up to the pre-set FiO limits (Min FiO to Max FiO fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
When PRICO is ON, this is indicated by the PRICO icon in the Status bar and the Green field with the current FiO labelled PRICO above the O parameter. Figure 12-12: PRICO “ON” fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
• There is arterial occlusion proximal to the sensor • There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or direct sunlight fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 106
The actual alarm limits and state (OFF) are also displayed in the graphical view next to the numeric values with the high/low alarm limit symbol. Figure 12-14: Alarm limits fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
PDMS interface is called acuLink. A license is needed to activate this feature, which can be ordered through your local distributor. fabian supports acuLink with three different protocols: Acutronic proprietary RS232 protocol without wave data Acutronic proprietary RS232 protocol with wave data IVOI protocol from Philips for Intellibridge/Vuelink via RS232 There are two major version of our acuLink protocol available: version 3 and version 4.
When installing a brand new sensor a second calibration can be necessary after about 30 minutes. If possible, expose the new sensor for about 30 minutes to the ambient air BEFORE installing fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
WARNING: The user of the device is responsible for performing routine maintenance when scheduled, and is also responsible for notifying ACUTRONIC Medical Systems in a timely basis when service needs to be performed. Maintenance Interval Indicator. The symbol will appear in the Information bar 30 days before the next maintenance has to be performed.
Accessories and consumables must be disposed of in accordance with the relevant instructions for use. For information, contact your local environmental or regulatory agency, or an appropriate waste disposal company. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Unfold the CaviWipe and wipe the surface of the enclosure completely. NOTE: The surface treated must remain visibly moist for three minutes. Use more wipes if necessary, to keep the surface moist for three minutes. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Switch OFF the fabian before cleaning and unplug at the mains in order to prevent damage due to penetration by liquid. Remove the Flow sensor from the flow sensor cable. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
NOTE: Prior to sterilization, ensure that the flow sensor and flow sensor cable have been replaced during the 12 month maintenance or after 30 Sterilization cycles, whichever comes first. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 114
NOTE: Articles must be individually single pouched in a qualified pouch (for example: Cardinal Health self-sealing pouch CAT# 92713). Each of the non-critical devices in- scope are thermostable and have been validated for steam sterilization. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Disassemble the membrane holder from the expiratory limb port by rotating the holder a quarter of turn in counter-clockwise direction and then pull it out from the port. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 116
Tap water (Set Point) Neutralization 02:00 Warm tap water Neodisher Z 1.0 ml per Liter Wash Rinse 1 01:00 Warm tap water Dry Time 07:00 90°C / 194°F fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 117
Dry cycle for 30 minutes Individually single pouch the exhalation membrane and membrane holder separately. Autoclave the exhalation membrane and membrane holder using the cycle parameters in Step 1 above. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 118
Duration of use The exhalation membrane and membrane holder must be replaced after 30 sterilization cycles and also in case of any detected defect that is affecting the accessory performance. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
If the flow sensor measurements become unavailable, the VG function will be disabled and t he pre-set Pinsp is used for ventilation. VG function will restart once flow sensor measurements become available again. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Select the appropriate VTe to be guaranteed by setting Vguarant to a target volume. A typical infant tidal volume target is 4 to 6mL/kg. Figure 17-1: Set Vguarant parameter fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 125
Pinsp setting and High PIP alarm setting. NOTE: To avoid unintentional breath abortions by reaching / exceeding High PIP, make sure that the High PIP alarm is set to at least +2mbar above Pmax. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 126
VG can now be activated by selecting the ON/OFF button. When VG function is enabled, this button will turn Green. A line will appear in the Pressure Wave diagram to mark the Pmax. Figure 17-4: : Vguarant ON / OFF button fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Ventilation immediately after E-time, otherwise after Preset Apnea Time. Using a delay time will determine the apnea period before backup ventilation begins. The Apnea Backup breaths are Volume Guaranteed breaths. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
It is important to place the flow sensor of the NO delivery system directly at the fresh gas outlet of the fabian +nCPAP evolution. The measuring line must be placed close to patient in Inspiratory Limb.
Value 18 Vol% Breath Frequency If Alarm Limit exceeded Alarm: Volume Monitoring If the breath volume set was not supplied Alarm Lower Threshold Value: Apnea Alarm fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
± 1.75% of reading or 0.1 mbar (whichever is greater) For volume measurements: ± 1.75% of reading or 0.1 mL (whichever is greater) For oxygen level measurements: ± 1% absolute fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Safety Classification (Applied Parts) From SN AN-04000: Type BF Till SN AN-03999: Type B Electrical Safety Classification Class I (mains power) Internally powered (battery power) Sound Pressure Max. 68 dB (A) fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
The Proportional Valves, which also act as return valves, prevent the return flow of a gas into the supply line of the other gas. An Oxygen sensor measures the Inspiratory Oxygen Concentration. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 134
19 Technical specifications Figure 19-1: Internal Gas Blending function of the fabian +nCPAP evolution Proportional Gas Blending Assembly Ventilator Manifold Air 2.0 to 6.0 bar Sensor Block Manifold Block Oxygen 2.0 to 6.0 bar Manifold PCB Sensor Pressure Regulator Manifold PCB Cover...
• Error of Delivered Volume (V ) of individual breaths greater than 35% and error of the DELIVERED VOLUME averaged over a one minute interval greater than 25%. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 136
± 3 bpm (no motion conditions); or ± 5 bpm (motion conditions); 201.11.8.101.1 201.12.4 Low and High Pulse Rate alarm conditions shall occur, as appropriate; 201.13.101 a technical alarm condition shall occur. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Guideline and Manufacturer Declaration – Electromagnetic Emissions The device fabian +nCPAP evolution is intended for operation in the environment described below. The customer or operator of the fabian +nCPAP evolution apparatus should ensure it is operated in this type of environment.
Page 138
Guideline and Manufacturer Declaration – Electromagnetic Immunity The device fabian +nCPAP evolution is intended for operation in the environment described below. The customer or operator of the fabian +nCPAP evolution apparatus should ensure it is operated in this type of environment.
Page 139
Guideline and Manufacturer Declaration – Electromagnetic Immunity The device fabian +nCPAP evolution is intended for operation in the environment described below. The customer or operator of the fabian +nCPAP evolution apparatus should ensure it is operated in this type of environment.
Page 140
The operator of the fabian +nCPAP evolution can help avoid electromagnetic interference by maintaining the minimum distance between portable and mobile HF telec ommunication devices (transmitters) and the device fabian +nCPAP evolution – depending on the output rating of the communication device, as listed below.
Guideline and manufacturer declaration – electromagnetic emission The device " fabian +nCPAP evolution " is intended for operation in the environment described below. The customer or user of the " fabian +nCPAP evolution " apparatus should ensure it is operated in this type of environment.
Page 142
Guideline and manufacturer declaration – electromagnetic immunity The device "fabian +nCPAP evolution" is intended for operation in the environment described below. The customer or user of the " fabian +nCPAP evolution " apparatus should ensure it is operated in this type of environment.
Page 143
HF communication devices and the device " fabian +nCPAP evolution " " fabian +nCPAP evolution " is intended to be operated in an electromagnetic environment with controlled HF interferences. The operator of the " fabian +nCPAP evolution " can help avoid electromagnetic interference by maintaining the minimum distance between portable and mobile HF telecommunication devices (transmitters) and the device "...
Page 144
20 Electromagnetic compatibility statement Recommended safety distances between portable and mobile HF communication devices and the device " fabian +nCPAP evolution " For transmitters with a maximum power output not listed in the above table the recommended safety distance d in metres (m) can be determined using the equation from the corresponding column, with P being the transmitter’s maximum power output Watts (W) per the transmitter’s manufacturer specifications.
Infrared radiation [wavelength greater than 700 nanometers (nm)]. Leaks Leakage from patient or system tubings /connections. Light Emitting Diode. LNCS Low Noise Cable Sensors (Masimo trade mark). Magnetic Resonance Imaging. Mechanical Ventilation. Non-invasive Ventilation fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...
Page 146
Link Power Management. Video Graphics Array. Tidal Volume is the lung volume representing the normal volume of air displaced between normal inhalation and exhalation when extra effort is not applied. fabian +nCPAP evolution | SW V5.1.x Ref: 122003.EN / Date : 26Jan2021...