6. Operating

6.1 Getting Started
1) The main voltage must correspond to the voltage given on the name-plate
2) Open the front door and remove packing materials.
3) Install shelves in the chamber.
4) Connect tubing to the drain.
Put the other end of the tubing to sink-hole which is lower than the valve.
5) Turn the circuit breaker on.
6) Turn the power switch on.
6.2 Over Temperature Protection
User must set the over temperature protection before operation.
Set this 10~20% higher than the set temperature.
6.3 How to set temperature
1) If you press the MODE button once, "TEMP" will be indicated on the display.
2) Press MODE button once again. Set temperature will be indicated on the display, waiting for
input.
4) Press LEFT SHIFT(◀) button to move to digit you want change value.
5) Press INC(▲) button to change value
6.4 How to set time
1) Press MODE button after temperature set finished. "TIME" will be indicated on the display.
2) Press MODE button once again.
3) The first digit of the LED display blink and prompt user input. (00:00)
4) Press LEFT SHIFT(◀) button to move to digit you want change value.
5) Press INC(▲) button to change value from 0 to 9.
6) Set four digits of time value you want to operate.
7) You can set timer up to 99 hours 59 min maximum.
If you do not want timer function, set the value 00:00.
8) After setting is done, press the MODE button
6.5 Operating
1) Temperature is controlled when the START/STOP button is pressed
2) If you press the START/STOP button during operation, the equipment stops
- 10 -