Siemens ULTRAMAT 6F Operating Instructions Manual page 40

Field devices of series 6
Hide thumbs Also See for ULTRAMAT 6F:
Table of Contents

Advertisement

Gas inlet conditions
Permissible sample gas pressure
Purging gas pressure
Sample gas flow
Sample gas temperature
Unheated devices
Heated devices
Sample gas humidity
7.4
Technical specifications CALOMAT 6F
Design, enclosure
Weight (analyzer)
Electrical characteristics
Power consumption
Fuse values
100 to 120 V
200 to 240 V
Gas path
Gas connections
Gaskets (O-rings)
Sensor
Measuring cell body
Internal gas path (gas lines)
Leak-tightness
Gas inlet conditions
Permissible sample gas pressure
Sample gas flow
Sample gas temperature
Temperature of measuring cell
Sample gas humidity
Purging gas pressure
40
500 to 3000 hPa (7.3 to 43.5 psi) absolute
< 165 hPa (2.4 psi) over ambient pressure
18 to 60 l/h (0.3 to 1 l/min)
0 to 50 °C (32 to 122 °F)
Max. 15 °C (27 °F) above temperature of analyzer part
< 90% RH (relative humidity), non-condensing
Approx. 25 kg (55 lbs.)
Approx. 30 VA
1T/250
0.63T/250
Stainless steel 1.4571, pipe diameter 1/4" or 6 mm
FFKM (e.g. Kalrez)
Silicon, SiO
N
, gold, epoxy resin, glass
x
y
Stainless steel 1.4571
Stainless steel 1.4571
Leakage loss < 1 μl/s
800 to 1100 hPa (11.6 to 16 psi) absolute
30 to 90 l/h (0.5 to 1.5 l/min)
0 to 50 °C (32 to 122 °F)
60 °C (140 °F)
< 90% RH (relative humidity), non-condensing
< 165 hPa
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779389002-01, 09/2018