Dear Customers, By purchasing a care bed from Malsch care & clinic design ® you have obtained a long-lasting medical device with func- tions that meet all the requirements of everyday care while maintaining the highest standards of safety. We would like to thank you for the trust you have placed in us.
C O N T E N T S Product overview Description of function Specific function Back rest (BR) Environmental sustainability Upper leg section (US) Note regarding the Height adjustment (HA) instructions for use Auto contour ('Nicole' / 'Andrea') Pictograms/symbols Trendelenburg-/ Safety instructions anti-Trendelenburg position ('Melanie') Technical data...
Page 5
IV drip holder Warranty and service Bedside light Declaration of conformity Reclining surface extension Certificates Urine bottle basket / urine bottle Hand controller holder Mattresses Side rail height extension/ triple side rail Troubleshooting Product safety Cleaning and disinfection Disinfection Care of wooden parts Maintenance Storage...
E N V I R O N M E N T A L S U S T A I N A B I L I T Y Betten Malsch GmbH care beds are used in the care sector Betten Malsch GmbH care beds are manufactured in ac- and in senior citizen facilities.
You will find these in tion manual is the basis for this. the Service area at www.bettenmalsch.de The instruction manual is designed for the Malsch homecare beds 'Melanie', 'Nicole' and 'Andrea'. The images, graphics and text it contains may differ from the equipment supplied.
In the event of any fault or failing which could endanger any person, the bed must not be used. 11. The Malsch homecare beds are equipped with an emer- gency mode powered by batteries. This allows the one-off Only trained staff may use the electrically operated care lowering of the reclining surface in the event of a power beds.
Page 10
ensure that there are no sharp edges, mechanical stresses Moving the bed on a sloping or unsafe surface or pinch/shear points. It is important that the following safety instructions are observed 15. The design of the hand controller means it can be hung to prevent risks and to avoid any damage to the bed: on the bed in such a way that the buttons are not acti- vated inadvertently between two objects.
Page 11
In addition, the safe working load should not be may only be carried out by an electrical specialist exceeded. or someone with electrical training using a special measurement and inspection device. Inspections If an immobile occupant remains in the same posi- in accordance with BGV A2 can be carried out by tion for a long period without the use of positioning specialist staff trained by the manufacturer as part...
T E C H N I C A L D A T A Low voltage drive system Height adjustment: approx. Melanie 40 up to 80 cm* (SMPS switching power supply) Nicole 40 up to 80 cm* Electrical connection: 240 V ≈ 50 Hz Andrea 33 up to 78 cm* Output voltage: 35 V = 2 A...
R A T I N G P L A T E Betten Malsch GmbH | Rohbergstraße 9 | D-36208 Wildeck-Obersuhl Tel.: +49 (0)6626 / 915 100 | Fax: +49 (0)6626 / 915 116 PFLEGEBETT Nicole 0815 1234567 180 kg 2016 Eingang: 100-240 V ~ 50/60 Hz 2.1-0.9 A...
H A N D C O N T R O L L E R H O L D E R - ' M E L A N I E ' Locking function on the back of the hand controller for restricting operation by occupants.
H A N D C O N T R O L L E R H O L D E R - ' N I C O L E ' / ' A N D R E A ' Locking function on the back of the hand controller for restricting operation by occupants.
F U N C T I O N S Mounting brackets Adjustable on both sides back rest Adjustable Mounting brackets Adjustable back rest on both sides upper leg section Adjustable lower leg section Bed panel Erectable Erectable side rails (full side rails (full length) length)
Adjustable Mounting brackets Adjustable back rest on both sides upper leg section Adjustable Adjustable lower leg section upper leg section Adjustable lower leg section Bed panel Lifting frame Lifting frame Adjustable Castors feet 'Andrea'...
D E S C R I P T I O N O F F U N C T I O N Depending on the design and type of care bed, the bed can The lower leg section (LS) can be put into a horizontal posi- be put into the following positions by adjusting the back rest, tion by care staff using an adjustable notched support.
sections are adjusted equally so that the occupant can be Caution! Please consider occupant safety brought into an upright sitting position. when adjusting the bed inclination. No for- eign objects should be located in the area of This function should only be used with mobile occupants and the lifting mechanism.
'Andrea' Braking and moving The 'Andrea' model features a non-moveable frame insert. 'Melanie'/'Nicole' Adjustment is achieved via height-adjustable plate stands which can be screwed and unscrewed through a thread. The 'Melanie' and 'Nicole' care beds are equipped with cas- tors with single castor fixing. All four castors can be locked Caution! The 'Andrea' frame insert must not (fig.
Side rail adjustment Full length side rail In the down position, the side rails are located on top of each other beside the reclining surface frame. Raising function: Lift up the upper side rail until you hear the safety bolt lock into place.
D E L I V E R Y A N D A S S E M B L Y Betten Malsch GmbH care beds are generally delivered fully assembled, or are assembled on site by company technicians or contractual partners.
A C C E S S O R I E S ( S E L E C T I O N ) Trapeze bar The trapeze bar can be inserted to the left and right of the head end in the designated mounting sockets on the reclin- ing surface frame.
Reclining surface extension The bed can be extended by up to 20 cm via the optional "reclining surface extension" feature. In this case, the existing side rails must be replaced. Caution! The reclining surface extension should only be assembled at the factory or by authorised specialist staff.
If this results in the and must not be used as an aid to standing up mattress depth increasing to more than 16 cm, Betten Malsch or as a handle. GmbH offers a flexible side rail height extension or the use of a triple side rail.
Hand controller operator, retailer hand controller retailer or Betten or motor is faulty or Betten Malsch buttons Malsch GmbH GmbH Locking of func- Release of func- Check all moving tions on the hand tions on the hand...
P R O D U C T S A F E T Y P R O D U C T S A F E T Y The product bears the CE marking and thus meets the requirements of German and European safety standards applicable to that product.
Care of wooden parts Malsch care beds only use wooden parts that are finished in compliance with the DIN 68861-1A standard. The aim is to produce a comfortable design, maximum functionality and a high level of practical use. To ensure you are able to enjoy this product for as long as possible, we recommend cleaning with commercial furniture cleaning products and polishes.
(temperature, humidity, heat, etc.) bed is used again and must be reported to the relevant person. Only original parts from Betten Malsch GmbH The manufacturer’s transport aid should be should be used. used to transport the beds.
A N D S E R V I C E C O N F O R M I T Y By purchasing a care bed from Betten Malsch GmbH, you have chosen a premium, high-quality product. The care beds have a 24-month warranty beginning from the purchase date.