Electromagnetic Immunity and Emissions Declaration
and Guidance
The Reader is intended for use in the electromagnetic environment
specified in the next table . The customer or the user of the Reader
should ensure that it is used in such an environment .
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Magnetic Field (50Hz)
IEC 61000-4-8
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips and
Interruptions
IEC 61000-4-11
IEC 60601-1-11
Conducted Fields
Disturbance
IEC 61000-4-6
Radiated Fields
Disturbance
IEC 61000-4-3
Radiated and
Conducted Fields
Aircraft use
16
Reader Compliance Level
± 8 kV Contact
± 15 kV Air
30 A/m
± 2 kV for power supply lines
± 1 kV line(s) to line(s)
0% 230V for 1 cycle
0% 230V for 0 .5 cycle at 8 phase angles
70% 230V (30% dip in 230V) for 25 cycles
0% 230V for 250 cycles
6 Vrms
150 kHz to 80 MHz
10 V/m at 80 MHz to 2700 MHz (AM
Modulation)
FAA RTCA /DO-160 edition G Section 20
Category T .
Can be used on aircraft according to the
directions provided by the operator of
the aircraft