4” x 40” spiral wound ultra filter system (28 pages)
Summary of Contents for Better Water EQASSY0.75KB2.03
Page 1
Better Water LLC Cartridge Based Pyrogen Filtration System (2-4 Filters) Operator Manual rev. Apr 2017 REA 2339...
Page 2
Better Water LLC. All rights reserved. The content of this manual is the intellectual property of Better Water LLC. It is furnished for the express use by Better Water LLC, their customers and dealers, for informational use only for operation, service, and internal training.
Cartridge Based Pyrogen Filtration System Operator Manual (2-4 Filters) TABLE OF CONTENTS Our Company ……………………………………………………………………………………... 01 - Contact Us ……………………………………………………………………………………….. 01 - Technical Phone Support ………………………………………………………………………. 01 - Technical Support Info Online …………………………………………………………………. 01 - Specific Contacts ……………………………………………………….……………………….. 02 - Introduction ………………………………………………………………………………………. 02 Warnings &...
Page 4
Cartridge Based Pyrogen Filtration System Operator Manual (2-4 Filters) Visit our website to see our complete product line of water purification products! www.betterwater.com Better Water LLC; 698 Swan Dr; Smyrna, TN 37167; www.betterwater.com rev. Apr 2017...
Cartridge Based Pyrogen Filtration System Operator Manual (2-4 Filters) Better Water LLC is a leading integrated manufacturer of water treatment equipment and components for the industrial, commercial and institutional Company markets. Located in Smyrna, Tennessee, Better Water LLC continues its history...
- Order Form - Return Goods Authorization Request Form Introduction The Better Water LLC Cartridge Based Pyrogen Filtration System is manufactured to the utmost quality. With proper care, preventative maintenance, and proper use, it should provide you with a very effective means of filtration for dialysis treatments.
9. Do not operate this unit in an environment where temperatures may be below 50 F or above 10. This device should not be used for purposes outside the device’s stated applications, specifications, or limitations. Better Water LLC; rev. Apr 2017 Page 3 of 20...
3. Drain requirements: This equipment does not require a drain. 4. Floor space: This unit is normally assembled at Better Water LLC and mounted to the wall surface. Requirements: Approx. 24” (depth-from wall) x 36” (width-along the wall) x 36” (height-up the wall);...
* NOTE: Details between the different models are basically the same with the exception of the number of filters. Bypass Valve Pressure Gauge Inlet/Outlet Valves Pressure Gauge Sanitary Sample Port (second port on opposite side) 20” Slim-Line Housings Better Water LLC; rev. Apr 2017 Page 5 of 20...
8. Filters are now on-line and ready for use. DAILY OPERATION 1. This device does not require any specific daily operation procedures, other than that the valves are in the proper position for normal operation. Better Water LLC; rev. Apr 2017 Page 6 of 20...
* OPEN during normal operation Outlet Valve Open to allow water to exit the device, and closed to prevent water from exiting the device. * OPEN during normal operation Better Water LLC; rev. Apr 2017 Page 7 of 20...
6% household bleach is 60,000 ppm Sodium Hypochlorite Better Water LLC recommends using IRON FREE bleach. Care should be taken to select iron free bleach because many discount or generic brands will have high iron content. Iron will be harmful to the equipment and shorten its lifespan.
5. When ready to sample, open the port for 1 to 2 minutes to allow product water to rinse the residuals and any endotoxins from the sample port before sampling. Better Water LLC; rev. Apr 2017 Page 9 of 20...
- Avoid sample cup cap contamination by coming in contact with any external surfaces. - Replace the Luer Lock Cap on the Sanitary Sample Port. 10. Follow appropriate procedures for collection of samples with Accu Vials for colony count and LAL. Better Water LLC; rev. Apr 2017 Page 10 of 20...
RO is running Perform Chemical, Microbial, and Endotoxin Testing on Schedule established by the Submit samples to a qualified feed and product water as Medical Director testing laboratory per AAMI requirements Better Water LLC; rev. Apr 2017 Page 11 of 20...
Operator Manual (2-4 Filters) SYSTEM MAINTENANCE, Changing the Filters Better Water LLC, recommends that the Pyrogen filters be changed every 6 months, unless otherwise dictated by, bacteria/endotoxin testing, and/or higher Delta-P (ΔP) pressure across the system. 1. Start with all valves, sample ports, and drain ports closed.
1. Filters are fouled 1. Perform a disinfect greater than 15 psi 2. Filters exceed recommended 2. If disinfection does not correct change out time the problem, the Pyrogen filters must be replaced. Better Water LLC; rev. Apr 2017 Page 14 of 20...
Buyer shall be responsible for all travel expenses and freight charges at all times. f. Better Water LLC shall have no obligation to repair or replace any defective part if buyer fails to follow the procedure set forth in “HOW TO OBTAIN A REPLACEMENT PART UNDER LIMITED WARRANTY”.
Page 20
Damage to or caused by filters/membranes or other replacement parts not purchased from Better Water LLC or damage caused by modification, alteration, repair or service of the equipment or any of its parts by anyone other than Better Water LLC or its expressly authorized representatives.
QUALITY ASSURANCE CHECK LIST (example) ITEMS TO BE CHECKED Date Time of Day Technician’s Initials Cartridge Based Pyrogen Filters PSI Pre-Filter, psi PSI Post-Filter, psi ΔP Across Filters (<15), psi Better Water LLC; rev. Apr 2017 Page 17 of 20...
The above flow rates should be set with all dialysis machines running. c. If bicarbonate or acid makeup water is required, both should have the proper flow controls to maintain flow velocity. Better Water LLC; rev. Apr 2017 Page 19 of 20...
Page 24
Cartridge Based Pyrogen Filtration System Operator Manual (2-4 Filters) Better Water LLC; rev. Apr 2017 Page 20 of 20...
Need help?
Do you have a question about the EQASSY0.75KB2.03 and is the answer not in the manual?