SYMPTOM
Torso and Face-Shield/Lung-Bag
Cannot hear a click when
doing compressions
Chest will not rise when
breaths are given
Manikin torso will not stay
closed (when breaths are
given)
* Contact your Prestan Authorized Distributor to either purchase the replacement part or receive it at no charge if the
manikin is within Prestan's 3-year warranty period and is a qualifying warranty repair.
First-Aid-Product.com - 760-944-1048
COPYRIGHT© 2016 PRESTAN PRODUCTS, LLC
Toll Free: (800) 933-8495
TROUBLESHOOTING GUIDE
(Continued)
POSSIBLE ISSUES OR COMMENTS
Hand position and/or compression
angle may be incorrect
Compressions may not be deep enough
to activate the clicker
Clicker may be broken
Head is not tilted back enough to open
airway
Face Shield is not attached
Lung Bag may be damaged or torn
(Note: there are holes in the Infant Lung
Bag necks by design to allow deflation)
Lung Bag is not lying flat or properly
positioned over the compression plate
and/or neck of bag is twisted
Lung Bag has accumulated moisture in
the neck, which may cause the sides to
stick together and prevent air passage
Front torso skin may not be securely
latched
Confirm correct hand placement and
compress straight down from directly
above the manikin
Compress between 2 and 2.4 inches
depth
Replace Clicker in back of manikin
(RPP-CCLICK-1)*
Push on forehead and/or lift at chin to
tilt head back enough to open airway
Attach Face Shield at "buttons" on the
side of the manikin's head below ears
Blow into the Face-Shield/Lung-Bag
while it is outside the manikin to make
sure it inflates properly – if not, discard
and use another Face-Shield/Lung-Bag
Revisit instructions for installing a Face-
Shield/Lung-Bag earlier in this manual
A new Face-Shield/Lung-Bag should be
installed after each class, or more often
if needed due to excessive moisture
buildup
Secure skin onto shoulder tabs
8
SOLUTIONS
11155-RevD