New-Flow DPG3000T Instruction Manual

New-Flow DPG3000T Instruction Manual

Digital pressure gauge

Advertisement

DIGTIAL PRESSURE GAUGE 3"- DPG3000T
Instruction Manual

1.0 PUSH-BUTTON DEFINITION

: Low Pressure Set Point / Display Peak Hold
: Set Different Pressure Unit / Pressure Setting Mode
UNIT/SET
: High Pressure Set Point / Reset Peak Hold

2.0 OPERATION METHOD

1. Press and hold down the UNIT/SET button for 3 seconds. The DPG3000T will go into the pressure setting mode to set the
high or low pressure set points.
2. Press and release the
for 3 seconds.
3. Press and release the
button to check the value of the high pressure setting, and the high pressure set point will be displayed
for 3 seconds.
4. Press and release the
reset zero. This is in order to make sure the input port is eigher disconnected or connected to a "0" psi input.
(NOTE.: Do Not operate the step when the DPG3000T in the airtight pipeline which has pressure. And plase reset zero under
the atmospheric pressure.)
5. To set the high or low pressure set points:
(a.) Press and hold down the UNIT/SET button for 3 seconds, the screen shows
high pressure setting, and the first digit will be flashing.
(b.) Set the first digit value by press the
(c.) Then pressure the
by the
button to complete the third and fourth digits.
(d.) After all 4 digits of the high pressure setting are down press the
the top, left hand side indicating low pressure setting. Repeat steps (b) and (c) above to set the 4 digits of the low pressure
settings.
(e.) After all 4 digits of the low pressure setting are down press the
indicating your pressure settings have been saved into memory.
(f.) To make sure that the settings are correct, you can press the
press the
button to check the value of the high pressure setting.
6. Display and reset peak hold:
(a.) To see the maximum peak pressure the unit has measured:
Press and hold down the
the screen.
(b.) To reset the maximum peak pressure:
Press and hold down the
(c.) To return to the current pressure reading:
Press and hold down the
measurement will be displayed and will not blink.
button to check the value of the low pressure setting, and the low pressure set point will be displayed
button and the
button at the same time for 3 seconds, the LCD will display "8888" and then
button (select from 0 to 9).
button, which shifts to the second digit. Again use the
button for 3 seconds. The maximum peak pressure value will be displayed and will blink on
button for 3 seconds, and the peak hold value will reset.
button for 3 seconds and the gauge will be back to the measurement mode, the current pressure
on the top, left hand side indicating
button to select 0 thru 9 followed
button one more time, the screen will show
button one more time, the screen will show "LOAD",
button to check the value of the low pressure setting and
on
DPG3000T_1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DPG3000T and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for New-Flow DPG3000T

  • Page 1 : High Pressure Set Point / Reset Peak Hold 2.0 OPERATION METHOD 1. Press and hold down the UNIT/SET button for 3 seconds. The DPG3000T will go into the pressure setting mode to set the high or low pressure set points.
  • Page 2 3.0 ELECTRICAL INTERFACE White Wire: Power Lead for supply input (10~24 VDC) Black Wire: Return Lead Brown Wire: Limit Switch output (NPN transistor output) Blue Wire: 4~20mA output for meter (must be loop grounded with black wire) 10~24 VDC White wire 75.5 Brown wire Black wire...

Table of Contents