Where to send a Kit for Re-Calibration / Repair Extent of Rosenberger Re-Calibration Services 6. Use, Maintenance and Care Connector Care Interface Dimensions and Gauging Handling and Storage 7. Connections Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260 D-84526 Tittmoning Germany www.rosenberger.com...
Tighten the connector coupling nuts of the RPC-2.92 interfaces to the recommended torque with a torque wrench (e.g. Rosenberger 03W021-000) while holding the RPC-2.92 female counter- part with a wrench (e.g. Rosenberger Combi Wrench 03W021-000). Do not over torque! Consider the power applied to the Load calibration standards. Maximum power limits for the sum of RF and DC power are defined in the specific data sheets.
User Manual 19CK010-150 Calibration Kit Follow further limitations given in the specific data sheets. 1.3 Kit Contents Table 1-1 Parts of the 19CK010-150 Calibration Kit Device Part number Remarks Quantity Open circuit plug 19S12L-000D3 Open circuit jack 19K12L-000D3 Short circuit plug...
These standard definitions are used in the calibration of the VNA to determine the systematic errors of the measurement system. Please always use the most recent source of information (Standard Definitions Cards or VNA specific files). Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
PNA family and compatible ones ‘calkit’ file can be used on Rohde&Schwarz VNAs ZVA, ZVB, ZVBT, ZNB families and compatible ones Example file name: 19CK010-150 (BH003)_mb_2020-05-30.calkit Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
OSLT is the most commonly used two-port calibration method. The Thru connections are performed by using well-known, precision calibration adapters in male-male or female- female configuration or as a direct connection of the male and the female test ports. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
5. Calibration Kit Service 5.1 Calibration interval Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use, the wear and the actual requirements a different calibration interval might be suitable. This lies in the responsibility of the user.
Small particles, usually of metal, are often found on the inside connector mating planes and on the threads. They should be removed to prevent damage to the connector surfaces. The items required for cleaning connectors and the procedure to be followed is described below. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
6.2.1 Gauging connectors A connector should be gauged before it is used for the very first time or if someone else has used the device to which it is connected. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 12
This will result in poor reflection and possibly unreliable contact. 6.2.2 Gauging There are no special Rosenberger gauges available for SMP connectors. It is recommended to use a Caliper (e.g. Mitutoyo ID-F150H) Checking devices for protrusion or excessive recession only a single measurement is sufficient.
Do not touch mating plane surfaces or the contact fingers. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove. The wearing of protective gloves is recommended. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Rev. Engineering change number Name Date Herbert Babinger 24/04/12 Markus Hantschel 21/12/20 20-1058 Lars Ramtke 21/12/20 Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260 D-84526 Tittmoning Germany www.rosenberger.com Email : info@rosenberger.com 14 / 16...
Page 15
Therefore, with every connection, care must be taken to ensure that all components are aligned with each other and remain so. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Need help?
Do you have a question about the 19CK010-150 and is the answer not in the manual?
Questions and answers