Ultimate Healthcare Tamora III Advance Quick Reference Manual

Ultimate Healthcare Tamora III Advance Quick Reference Manual

Hide thumbs Also See for Tamora III Advance:

Advertisement

Quick Links

Tamora III Advance
Quick Reference Guide
Please refer to the product User Manual for full instructions for use.
Control Panel Lock Out
To unlock panel:
Press & hold for 3 seconds to unlock control panel.
To lock panel:
To lock out the control panel press the Lock Out
button and all functions will be locked.
If the control panel is not used for a period of 30
seconds it will automatically lock out and inhibit
the use of the functions.
Comfort Control
The Soft and Firm buttons allow adjustment of
pressures within a safe pre-set range.
When pressing the Firm button, the output
pressure will increase to provide a higher
pressure output and thus increased support.
When pressing the Soft button, the output
pressure will be decreased to provide a lower
pressure output and thus increased comfort.
Auto-Detection -
detecting patient weight (distribution)
Simultaneously pressing the Soft and Firm buttons on the
control panel will manually activate Auto-Detection.
will start detecting the patients weight
internal pressures appropriately. The system will
continuously check this every 2 hours.
Auto Firm Mode
Pressing the Auto Firm button quickly inflates
all cells to a single pressure setting.
The system will automatically return to
Last Mode of Operation after 20 minutes.
Function Mode Button
Pressing the Mode button allows you
to cycle between the different modes
available such as Alternating, Static and
Constant Low Pressure.
to set
Low Pressure Alarm
Upon detection of low pressure, an audible alarm will be
heard and the Low Pressure indicator will illuminate.
Power Failure Alarm
If mains power is removed from power unit or the power cable is
unplugged without turning the power unit OFF, an audible alarm
will sound and the Power Failure indicator will illuminate.
Service Alarm
Should your system develop a fault condition whilst in use, an
audible alarm will be heard and the Service indicator will illuminate.
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
Alternating Mode
Alternating mode is the default mode for the
and deflate in an alternating 1-in-3 cell cycle.
The Alternating cycle will continue at the selected
cycle time until another mode is selected.
Static Mode
Pressing the Mode button until the Static indicator
illuminates puts the system into Static mode.
The system will automatically return to
Alternating mode after 20 minutes.
Constant Low Pressure (CLP) Mode
Pressing the Mode button until the CLP indicator
illuminates puts the system into Constant Low Pressure
mode.
Within this mode the cells are not alternating and the
internal cell pressure within the mattress is set at 15mmhg
to provide a comfortable pressure reducing surface.
Alternating Cycle Time selection
Choice of 10/15/20/25/30 minutes.
To select press Cycle button until required cycle
time is displayed in LCD screen.
Alarms
Alarm Mute
The Alarm Mute button temporarily resets the
audible Low Pressure/Power Failure/Service
alarms. Should the situation not be resolved and
the fault condition continues the alarm will resume.