Summary of Contents for DRUCK & TEMPERATUR Leitenberger LR-Cal Pyros Basic 375
Page 1
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 1 of 24 Opera ng Manual Dry Block Temperature Calibrator PYROS Basic 375 + PYROS Basic 650 DRUCK & TEMPERATUR Leitenberger GmbH Bahnhofstr. 33 72138 Kirchentellinsfurt GERMANY Tel.: +49-7121-90920-0 Fax: +49-7121-90920-99 E-Mail: DT-Export@Leitenberger.de...
Page 2
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 2 of 24 INDEX 1 - INTRO DUCTION ......................4 1.1 - Purpose and summary of instru c ons ...............4 2 - SCOPE OF SUPPL Y......................5 2.1 - Name : ......................
Page 3
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 3 of 24 WARNING Hazardous voltages are present in this electrical equipment during opera on. Non-observance of the safety instruc on can result in severe personal injury or property damage.
Page 4
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 4 of 24 1 - INTRODUCTION 1.1 - Purpose and summary of instruc ons This manual contains the use and maintenance instruc ons valid for the following equipment: PYROS Basic 375 and PYROS Basic 650 Portable Temperature Calibrator model:...
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 5 of 24 2 - SCOPE OF SUPPLY 2.1 - Name: Portable Temperature Calibrator, complete of accessories as listed (reference to paragraph 2.7). 2.2 - Technical data: TECHNICAL DATA PYROS Basic 375 PYROS Basic 650...
Page 6
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 6 of 24 2.3 - Service (func on): The portable temperature calibrator PYROS Basic 375 PYROS Basic 650 has been designed for: Control andcalibra on of temperature sensors, in the laboratory, on board and in the field, in conformity with ISO 9000 standard.
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 7 of 24 3 - GENERAL RECOMMENDATIONS Î ATTENTION The configura on of the instrument is protected by a password and by the protec on code. To accede at the configura on parameters reset the password & the protec on code: contact our technical office.
Page 8
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 8 of 24 - Advises: The me the probes takes to reach the Set point is much more higher as the difference in diameter from the probes and the holes is bigger. Do not insert the probe when the instrument has already reached the set temperature;...
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 9 of 24 4 - SAFETY INSTRUCTIONS ATTENTION: Due to the fact that the calibrator is a portable instrument to be used in the field, it is very important to ensure that the socket has been earthen correctly when connec ng it to the electricity supply.
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 10 of 24 5 - PREPARATION OF OPERATION Remove the calibrator from the packaging and place it on a flat surface. Make sure that the instrument has been correctly earthen. Supply the oven with line 230V, 50Hz (3,15A) or 115 V (6,3A) + earth, 3,15A (6.3A for 115V).
Page 11
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 11 of 24 COMMANDS LIST POS. DESCRIPTION SUPPLY SOCKET MAIN SWITCH PROTECTION FUSES SWITCH TEST SWITCH TEST ON/OFF LIGHT TEMPERATURE REGULATOR VOLTAGE SWITCH/INDICATOR...
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 12 of 24 6 - OPERATION PROCEDURE 6.1 - Opera on descrip on The PYROS-650 calibrator consist of a metal dry well with one hole of 26x150mm into which, the inserts available for almost any sensor size to be calibrated, are inserted.
Page 13
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 13 of 24 6.2.5 - Temperature sensors The temperature sensor used for the reading and thermoregula on is a Thermocouple probe; the probe is inserted directly into the equalising block so as to supply a temperature value close to the real value in the block.
Page 14
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 14 of 24 Before any calibra on follow the general recommenda on: Switch on the calibra on with the main switch (2); wai ng for the end of self-test procedure. Put the probe to check into the equaliser block: reference to chapter 3.
Page 15
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 15 of 24 Enter the pass. ‘1020’ using the arrows and press the “P” bu on again (if the password is entered correctly, the display will show Sp 1, if not, the instrument returns to the condi on it was in previously).
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 16 of 24 For example: If you want to calibrate the Pyros to the values 50 and 500°C proceed in this steps: insert a standard thermometer in the oven and set the set point of 50°C and wait un l the oven is stable then read the value of the display of Pyros and the value of the standard probe.
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 17 of 24 9 - TYPICAL FAULTS Before carrying out these opera ons the instrument must be disconnect from the electricity supply and the equaliser block must be at ambient temperature. N°...
Page 18
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 18 of 24 10 - APPENDICES 10.1 -Thermoregulator Descrip on of the func ons OUT1 = indicates that the hea ng is ON OUT2 = indicates that the fan coil is ON at low speed DISPLAY This is used to increase the...
Page 19
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 19 of 24 10.2 - Spare parts list LR-Cal PYROS-650 (Reference numbers referring to the enclosed drawings) CUP SOCKET 3SCH28366 FUSES 5x20 3,15A for 230V 5x20 6,3A for 115V SWITCH TEST CO NNECTION 3B&BPAN10A SWITCH TEST LAMP...
Page 20
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 20 of 24 "Declara on of conformity" DRUCK & TEMPERATUR Leitenberger GmbH, Bahnhofstr. 33, 72138 Kirchentellinsfurt, GERMANY THERMOSTATIC CALIBRATOR LR-Cal PYROS-375 / PYROS-650 Declares that the: is conforms with the requirements of the following European direc ve: Low voltage direc ve 73/23/EEC amended by 93/68/EEC and by 2006/95/CE EMC direc ve EMC 2004/108/CE and that it has been designed in accordance with the following harmonized regula on:...
Page 21
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 21 of 24...
Page 22
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 22 of 24...
Page 23
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 23 of 24 NOTES:...
Page 24
DRUCK & TEMPERATUR Leitenberger GmbH: Manu PYROS Basic 375 + PYROS Basic Page 24 of 24 www.lr-cal.net DRUCK & TEMPERATUR Leitenberger GmbH Bahnhofstr. 33 72138 Kirchentellinsfurt GERMANY Tel.: +49-7121-90920-0 Fax: +49-7121-90920-99 E-Mail: DT-Export@Leitenberger.de www.LR-Cal.net...
Need help?
Do you have a question about the LR-Cal Pyros Basic 375 and is the answer not in the manual?
Questions and answers