PLATINOUS SERIES CONTROLLER MANUAL Introduction This manual has been written for users of the Platinous Sterling Series Chambers and particularly for technicians. Read it thoroughly to obtain the maximum performance from the chamber. Safety Symbols The following safety symbols are used throughout this manual. Degrees of Danger This mark means extremely dangerous VERY...
Page 6
The user’s manual composition that came with your chamber contains the following reference materials. Use them as necessary. Manual Content Instruction Manual Explains the basics of the Platinous Sterling Series. SCP-220 Controller Explains operation from the instrumentation panel in greater detail. Quick Reference Reorganizes setting procedures into a simple, handy digest.
PLATINOUS SERIES CONTROLLER MANUAL 1. Overview The SCP-220 Instrumentation is a graphic interface for operating and monitoring the chamber. The user moves from menus to monitor and setup screens as he/she prepares and conducts tests. The software is divided into setup, operation and monitoring functions. This chapter gives an overview of each of them.
PLATINOUS SERIES CONTROLLER MANUAL 1.2. Operation Interface The operation interface gives the user start and stop control over the constant and program modes. Below is a function tree of this interface. OPER./STOP Stop Operation Constant Mode START PGM Selection Step No. Program Mode START PAUSE...
PLATINOUS SERIES CONTROLLER MANUAL 4. Flashes during communications between the chamber and the PMS-CA(CG) protocol converter. Also, pressing this box displays the E-BUS setup screen, unless a password has been registered with the instrumentation. 5. Displays the date and time. Also, pressing this box displays the Date and Screen Saver Screen, unless a password has been registered with the instrumentation.
Page 10
PLATINOUS SERIES CONTROLLER MANUAL Keypad: Numerical Input Use for numerical input. 0 ~ 9 keys Use to input a decimal point. . key Use to input negative numbers. Clears the value in the text box. CLR key Deletes the value left of the cursor. DEL key Cancels the displayed input and returns the display to the original QUIT key...
PLATINOUS SERIES CONTROLLER MANUAL 2. Monitoring 2.1. Chamber Monitor Target Temperature Monitored Conditions Target Humidity * In the following situations, “- - -“ is displayed in the field boxes. • If a sensor input error occurs • If either dry/wet-bulb temperature is outside the 0 ~ 100°C range (for humidity display only) •...
PLATINOUS SERIES CONTROLLER MANUAL 2.2. Trend Graph Horizontal Axis The horizontal axis determines the data sampling interval. The time scale can be selected between 10 minutes and the Max. Rec. time set on the Sampling Setup Screen. Range Control Increases the displayed scale. Expand Decreases the displayed scale.
PLATINOUS SERIES CONTROLLER MANUAL 2.3. Detail Monitor: Chamber Table 2.1. Detail Monitor: Chamber Screen Process Value Displays the value currently monitored inside the chamber. Set Point Displays the current control target setting. Heater % In temperature-controlled operations: Displays heater output. In humidity-controlled operations: Displays humidifying heater output.
PLATINOUS SERIES CONTROLLER MANUAL 2.4. Detail Monitor: Time Signal Monitor Table 2.2. Time Signal Monitor Screen Specimen PWR CNTL Indicates whether power is being supplied to specimens or not. Power supply to Standard specimen is cut when the alarm occurs. Time Signal (1/2) Indicate whether the user-set time signals have been output or not.
PLATINOUS SERIES CONTROLLER MANUAL 2.5. Detail Monitor: Program Monitor This screen can only be displayed during programmed operations. It contains information on the program being run. 1. PGM Name Displays the No. and name of the program being run. 2. Current Step Displays the No.
Page 16
PLATINOUS SERIES CONTROLLER MANUAL Displays the total time the program has run up to the present. Time is counted up and it includes time lost to alarms and time consumed while the chamber is paused. 11. End Mode Displays the operating mode the chamber will assume at program end. 12.
PLATINOUS SERIES CONTROLLER MANUAL 3. Constant Mode Setup This chapter explains how to set up the constant mode. The constant mode is the basic chamber operating mode. 3.1. Constant Mode Setup The following procedure explains how to make control settings for the constant mode. 1.
PLATINOUS SERIES CONTROLLER MANUAL 3. Turn humidity control ON/OFF as needed. (For temperature-only chamber only). Press the ON box next to Hum. Control to make the setting. When the dot is yellow: Humidity control is ON. When dot is black: Humidity control is OFF. 4.
PLATINOUS SERIES CONTROLLER MANUAL Reference You can name time signals. For details on naming, see “Name Time Signals” in “5.11. Chamber Configuration”. 3.3. Product Temp. Control (Option) Detail 1. Press the P.T.C. Detail button at the bottom of screen to get the product temp. control detail screen.
PLATINOUS SERIES CONTROLLER MANUAL 4. Program Mode Setup This chapter explains how to set up the program mode. 4.1. Program Create Programs are created by inputting or editing target temperature, humidity and time for each program step. The basic procedure is as follows. 1.
Page 21
PLATINOUS SERIES CONTROLLER MANUAL 3. Select the input mode from the editing buttons. Press the Input box to newly input program data or to change existing data. Editing Page Scroll Buttons Buttons [Editing Buttons] Press to input new settings or edit existing ones. Input Press to add new steps in-between existing steps.
Page 22
PLATINOUS SERIES CONTROLLER MANUAL [Input & Edit Screen] 4. To edit existing programs, display the step to be edited with the page scroll buttons and press the Step No. box of that step. Steps can be selected from step No. 1 to one number higher than the last step.
Page 23
PLATINOUS SERIES CONTROLLER MANUAL Turn ramp control ON/OFF as needed. Press the ON box next to Ramp Control to make the setting. When dot is yellow: Ramp control is ON. Ramp control is applied from the previous step’s control target to this step’s target. When dot is black: Ramp control is OFF.
Page 24
PLATINOUS SERIES CONTROLLER MANUAL 12. Turn soak control ON/OFF as needed. Press the ON box next to Soak Control to make the setting. When dot is yellow: Soak control is ON. The exposure time set for the step will be counted down only while the temperature (&...
PLATINOUS SERIES CONTROLLER MANUAL 4.2. Product Temp. Control (Option) Detail 1. Press the P.T.C. Detail button at the bottom of the program step edit Detail screen to get the product temp. control detail screen. The P.T.C. Detail button will not be enabled if your chamber does not have the product temp.
PLATINOUS SERIES CONTROLLER MANUAL 4.3. Other Setup From the Other Setup screen, you name programs and select end processing. Press to name programs and select end processing. 1. Get the Other Setup screen. Press the Other box on the bottom of the Program Input & Edit screen. 2.
Page 27
PLATINOUS SERIES CONTROLLER MANUAL Select the end processing. This sets chamber operating status at the end of the last step in the program. Shuts OFF power to the chamber. Power OFF (The breaker remains in the ON position.) Holds the temperature and humidity control Hold End Step targets of the last step.
PLATINOUS SERIES CONTROLLER MANUAL 6. Turns counters A/B ON/OFF as needed. Press the ON box under Counter A/B to make the setting. When dot is yellow : The set step range will be repeated the number of cycles set in Repeat Cycles.
PLATINOUS SERIES CONTROLLER MANUAL When the Save button is pressed Saves the input data and returns the display to the Yes button Program List screen. If you changed existing settings, the original data is updated with the new data. Returns the display to the Program Select Edit No button screen without saving data updates.
PLATINOUS SERIES CONTROLLER MANUAL 2. To select the step where wanting to add the new step(s), display the step with the page scroll button and press the Step No. box of that step. 3. A blank step is added in the selected position. The existing step and all subsequent steps will be pushed back one step for every new step added.
Page 31
PLATINOUS SERIES CONTROLLER MANUAL 2. Select the first in the range of steps to copy. Display the first step to copy with the page scroll buttons and press the Step No. box of that step. Range Number of steps : 1 ~ 49 First step to copy : 1 ~ last step in program Last step to copy : First step to copy ~ last step in program Copy destination step : 1 ~ Last step in program + 1...
PLATINOUS SERIES CONTROLLER MANUAL 4.6. Step Delete This function deletes steps from the selected program. Steps are deleted one at a time. Once a step has been deleted, all subsequent steps move up in the program to fill the space. 1.
PLATINOUS SERIES CONTROLLER MANUAL 4.7. Program Delete This function deletes a selected program from the Program List. 1. Get a program setup mode. Press the Program Setup box on the main menu. TEMP. Get the program delete mode. Press the Delete button on the Program List screen. 3.
PLATINOUS SERIES CONTROLLER MANUAL 4. The selected program will be highlighted and a message confirming whether to execute the operation or not will be displayed Deletes the selected program and returns the Yes button screen to the Program List screen. (The delete command will still be active.) Returns the screen to the Program List screen No button...
Page 35
PLATINOUS SERIES CONTROLLER MANUAL 3. Select the program to copy. Only registered programs can be copied. If an unregistered program is selected, a message notifying you to that effect will be displayed. 4. Select the program space where a copy in the program. Programs cannot be copied into ROM programs Nos.
PLATINOUS SERIES CONTROLLER MANUAL 4.9. Program View This function displays settings of the selected program in table format. Selected programs can be viewed but not edited. It is a handy feature for checking setup of programs that have been run. 1.
Page 37
PLATINOUS SERIES CONTROLLER MANUAL 3. Select the program to be displayed. 4. Program settings will be displayed in table format. [Buttons] Displays the previous page. Displays the next page. Displays a keypad where to input the step number. That Step step will appear at the top of the table.
PLATINOUS SERIES CONTROLLER MANUAL 5. Chamber Setup 5.1. Timer Setup 5.1.1. Start Timer Setup 1. Get the chamber setup mode. Press the Chamber Setup box on the main menu. TEMP. 2. Press the Timer Setup box on the Chamber Setup screen.
Page 40
PLATINOUS SERIES CONTROLLER MANUAL 3. Press the Setup button under the Start button. 4. Set up the start timer. Set the start mode, operation mode, date, time and the day of the week. Table 5.1. Start Timer Setup Screen Elements 1 Time Starts the chamber at the specified time on one special day.
PLATINOUS SERIES CONTROLLER MANUAL 5. Save the start timer. Press the Save button. A message confirming whether to execute the operation or not will be displayed on the screen. Saves the timer setup. Yes button Returns the display to the Timer Setup screen No button without saving the timer setup.
Page 42
PLATINOUS SERIES CONTROLLER MANUAL 2. Press the Timer Setup box on the Chamber Setup screen. 3. Press the Setup button under the End button.
Page 43
PLATINOUS SERIES CONTROLLER MANUAL 4. Set up the end timer. Set the start mode, operation mode, date, time and the day of the week. Table 5.2. End Timer Setup Screen Elements 1 Time Shuts down the chamber at the specified time on one special day.
PLATINOUS SERIES CONTROLLER MANUAL If you press the Cancel button, a message confirming whether to ignore timer setup or not is displayed. Returns the display to the Timer Setup screen Yes button without saving the timer setup. Returns the display to the End Timer Setup No button screen.
PLATINOUS SERIES CONTROLLER MANUAL 3. Press the Start and End buttons. The ball in the Start and End buttons and the CLOCK box along the top of the screen will turn yellow when the timer is turned ON. At the same time, the details of the ON timer(s) will be displayed in the display area(s) on the Timer Setup screen.
Page 46
PLATINOUS SERIES CONTROLLER MANUAL 2. Press the High/Low Limit Setup box on the Chamber Setup screen. 3. Set the absolute high/low temperature and humidity and the upper deviation temperature. The absolute high/low limits will determine the target input range in the constant and program modes.
PLATINOUS SERIES CONTROLLER MANUAL Absolute high/low limits These settings are designed to protect the chamber against thermal damage. Set them at least 10°C higher/lower than target temperature. When tripped, alarm generated and the chamber stops running. Upper deviation limit This setting is designed to protect specimens against heat damage.
PLATINOUS SERIES CONTROLLER MANUAL 3. Set the sampling cycle. This cycle determines at what interval data for the trend data is to be sampled. Cycle Sets sampling cycle. Selectable from 1 sec, 3- sec, 1 min or 5 min. Max. Rec. Time Displays the maximum amount of data that can be stored at the Cycle setting at left.
Page 49
PLATINOUS SERIES CONTROLLER MANUAL 2. Press the Protection ON/OFF box on the Chamber Setup screen. 3. Turn protection ON/OFF as needed. Pressing the boxes turns the protection ON/OFF. The dot is yellow when protection on ON from remote. See Table 5.3 for a description. * When keys or buttons locked by the protection are pressed, a warning message is displayed on the screen.
PLATINOUS SERIES CONTROLLER MANUAL 5.6. Defrost Setup (Option) This option is not available on Platinous Sterling Series Chambers. 5.7. Humidity Calculation 1. Get the chamber setup mode. Press the Chamber Setup box on the main menu. TEMP. Press the Humidity Calculation box on the Chamber Setup screen.
Page 51
PLATINOUS SERIES CONTROLLER MANUAL Set up humidity calculation. This function calculates dry-bulb temperature, wet-bulb temperature or relative humidity using the inputs from the other two of these variables. Here, it is also possible to select the method of steam pressure calculation. The air Velocity setting is available only when Pernter selected.
PLATINOUS SERIES CONTROLLER MANUAL 5.8. Alarm Report View 1. Get the chamber setup mode. Press the Chamber Setup box on the main menu. TEMP. 2. Press the Alarm Report box on the Chamber Setup screen. Displays the No. (1 ~ 100) assigned to each alarm.
PLATINOUS SERIES CONTROLLER MANUAL 5.9. ROM Information View 1. Get the chamber setup mode. Press the Chamber Setup box on the main menu. TEMP. 2. Press the ROM Information box on the Chamber Setup screen. * ROM version is displayed in the boxes.
PLATINOUS SERIES CONTROLLER MANUAL 5.10. Service Guide 1. Get the chamber setup mode. Press the Chamber Setup box on the main menu. TEMP. 2. Press the Service Guide box on the chamber Setup screen.
PLATINOUS SERIES CONTROLLER MANUAL [Period Checks] Leakage Breaker Trip Test* Overheat Protector Test Humidifying Water level Check [Maintenance] Wet-Bulb Wick Check Condenser Filter Cleaning* Aux. Water Tank Cleaning* Water Tank Filter Cleaning* Main Water Tank Cleaning* Water Pump Filter Cleaning* Humidifying Tray Cleaning Test Area Cleaning Electrical Parts/Water Circuit Comp Cleaning...
Page 56
PLATINOUS SERIES CONTROLLER MANUAL 2. Press the Configuration box on the Chamber Setup screen. 3. Input the password. A keypad will be laid over the screen to enable input. Make your input from there. If a password has not been registered, proceed to step 4 of this procedure. 4.
PLATINOUS SERIES CONTROLLER MANUAL 5.11.2. E-BUS Setup (Optional) (Not available on Platinous Sterling Series Chambers) The E-BUS is for communications between the chamber and a chamber. It requires the PMS-CA(CG) protocol converter. For more information on communications and operation, see the user’s manual provided with the PMS-CA(CG). This screen is for setting the chamber address and baud rate used in E-BUS communications.
PLATINOUS SERIES CONTROLLER MANUAL 5.11.3. Power Recovery Setup This screen is for setting the recovery mode in the event of a power failure, the backup mode to be applied when trouble occurs, and door alarm control. 1. Select the Recovery Setup item. Press the 2 box on the Chamber Configuration screen.
PLATINOUS SERIES CONTROLLER MANUAL 5.11.4. Control Attainment Range Setup This screen is for setting temperature and humidity attainment range and holding time. The chamber uses these settings to stabilize control. When the control item stays within the set Attainment Range for the set holding Time, the chamber assumes the control target to have been reached and starts the step/test timer.
PLATINOUS SERIES CONTROLLER MANUAL 5.11.5. Analog Output Setup This screen is for setting up analog data output. The signals are output from the connection terminal for temperature & humidity recorder (option). The factory setting is humidity output. The humidity terminal is labeled, therefore be careful if outputting another channel.
PLATINOUS SERIES CONTROLLER MANUAL 5.11.6. Name Time Signals This screen is for naming time Nos. 1 ~ 8. If an unregistered signal is selected, a message notifying you to that effect will be displayed. 1. Select the Name Time Signals items. Press the 5 box on the Chamber Configuration screen.
PLATINOUS SERIES CONTROLLER MANUAL 5.11.7. Date & Screen Saver This screen is for setting the chamber’s built-in calendar and clock, and the screen sleep mode. 1. Select the Date & Screen Saver item. Press the box on the Chamber Configuration screen. The same result can be achieved by pressing the date display, unless a password has been registered with the instrumentation.
PLATINOUS SERIES CONTROLLER MANUAL 5.11.8. Register User Password To prevent access to the chamber configuration mode, a password can be registered with the instrumentation. Setting a password will required that the password be input before any of the setup screens can be accessed. If it is not necessary to protect settings, then a password need not be entered.
Select for chamber at altitude between 2500 and 2500-4999 4999 feet above sea level. Select for chamber at altitude 5000 feet above 5000~ sea level or higher. Note: The altitude setting is used to automatically adjust the SCP-220 internal humidity calculation for the altitude of the chamber.
Page 65
PLATINOUS SERIES CONTROLLER MANUAL 5. Set the No Frost Expand Range This setting should not be changed by the customer. Select “Yes” for all temperature humidity chambers. Select “No” for all temperature only chambers...
PLATINOUS SERIES CONTROLLER MANUAL 5.11.11. Product Temperature Control (Option) Setup This screen is for product temperature control option settings. This screen will not be accessible if the chamber is not equipped with this option. This option allows the chamber to bring the user product (or specimen) to the desired temperature faster by varying the “internal air setpoint”...
Page 67
PLATINOUS SERIES CONTROLLER MANUAL This tuning parameter is used to dampen the affect of the P parameter on the air temp SV. The D parameter relates to how quickly the rate of temp change is for the product. For more dampening, enter a large D value. For less dampening, enter a small D value.
This mode is reserved for serviceman use and cannot be accessed or used by the user. Tampering with parameters can lead to trouble. ESPEC will not assume responsibility for equipment trouble whatsoever when resulting from unauthorized use of the service mode.
PLATINOUS SERIES CONTROLLER MANUAL 6. Operation Modes 6.1. Restrictions on Operation Control The Operation Mode Selection screen can be accessed from any screen except in the following cases. • While confirmation messages are displayed • While the keypad is displayed •...
PLATINOUS SERIES CONTROLLER MANUAL 6.2. Startup in the Constant Mode 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen. 2. Start operation in the constant mode. Press the START button under Constant Mode.
PLATINOUS SERIES CONTROLLER MANUAL 6.3. Ending the Constant Mode 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen. 2. Stop operation in the constant mode. Press the STOP button under Stop Operation.
PLATINOUS SERIES CONTROLLER MANUAL 6.4. Program Selection 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen. 2. Get the program and start step selection mode. Press the PGM No.
PLATINOUS SERIES CONTROLLER MANUAL 6.5. Startup in the Program Mode 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen. 2. Start the selected program. Press the START button under Program Mode.
PLATINOUS SERIES CONTROLLER MANUAL 6.6. Pausing in the Program Mode 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen. 2. Pause the program. Press the PAUSE button under Program Mode. 3.
PLATINOUS SERIES CONTROLLER MANUAL 6.7. Restarting in the Program Mode 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen. 2. Restart the program. Press the RESTART button under Program Mode. 3.
PLATINOUS SERIES CONTROLLER MANUAL 6.8. Jump to Next Step This function suspends the current step and starts the next step of the program. 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen.
PLATINOUS SERIES CONTROLLER MANUAL 6.9. Ending in the Program Mode 1. Get the Operation Mode Selection screen. Press either the OPER./START key on the operating panel or the chamber operating status box on the screen. 2. Stop the program. Press the STOP button under Stop Operation. 3.
Page 80
PLATINOUS SERIES CONTROLLER MANUAL PLATINOUS Series SCP-220 Controller Standard Chamber Instruction Manual Original 1.00: April 30, 1999 Edited and Published by: ESPEC NORTH AMERICA. Revision 1.01: July 25, 2001 425 Gordon Industrial Court S.W. Byron Center, MI 49315 Revision 1.02: July 18, 2003...