Silberbauer Compact-Laser Series Instructions For Use Manual

Silberbauer Compact-Laser Series Instructions For Use Manual

Compact - laser

Advertisement

Quick Links

         
Compact - Laser
CL plus
Instructions for use
Type: CL plus 60-660 L
0120
Serial number: 60660 L 601 J
assembled by
Silberbauer India
with the European know how of
Silberbauer Medical Electronics, Vienna,
Austria, Europe

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Compact-Laser Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Silberbauer Compact-Laser Series

  • Page 1           Compact - Laser CL plus Instructions for use Type: CL plus 60-660 L 0120 Serial number: 60660 L 601 J assembled by Silberbauer India with the European know how of Silberbauer Medical Electronics, Vienna, Austria, Europe...
  • Page 2: Table Of Contents

    Content: Page: 1 Intended Use ......................4 2 Safety precautions to be observed during the use of therapy lasers ..........................4 2.1 Alignment Laser Spectacles (Therapist’s Spectacles): ....5 2.2 Laser Protection Spectacles (Patient’s Spectacles): ....6 2.3 Indications ....................6 2.4 Contraindications ..................
  • Page 3  Compact - Laser Series CL plus Instructions for use Delivery :  2 ea. NiMH – Batteries „Green“, 2050mAh,  Charger for NiMH- Batteries,  Laser – Storage, Acrylic Glass,  lockable Casket, Aluminum,  Instructions for use  Applications of the Soft–Laser ...
  • Page 4: Intended Use

    1 Intended Use The Compact – Laser plus is a Soft Laser (Low Level Laser) for radiation of skin, mucous membrane and dental applications in human and veterinary medicine (LLLT = Low Level Laser Therapy). It is a “Medical Product“ in accordance with the EU Medical Product Law and with the Essential Requirements 93/42 EWG and 2007/47 EC.
  • Page 5: Alignment Laser Spectacles (Therapist's Spectacles)

    CL plus until 100 mW: for all other CL plus models: and admitted for laser mode D (continuous operation) Silberbauer - laser alignment spectacles offer only protection against accidental  irradiation through laser up to these levels! They are NOT ALLOWED to be used for the deliberate view of beam! ...
  • Page 6: Laser Protection Spectacles (Patient's Spectacles)

    CL plus: Protection level at the wavelength of the laser : Laser mode: D (continuous operation) The Silberbauer laser protection spectacles offer only protection against  accidental irradiation through laser up to these powers! They are NOT ALLOWED to be used for the deliberated view into the beam! ...
  • Page 7: Contraindications

     Compact - Laser Series CL plus Instructions for use 2.4 Contraindications The laser treatment has a low risk only. Different authors have controversial positions because the laser is especially successful in applications where you for forensic reasons might have concerns. However, if no sufficient tests are available, some contraindications should be observed: ...
  • Page 8: Compact - Laser Cl Plus : View

    7 Rotary Switch 2 (Dose) 2 Aluminium Body 8 Type Plate and Details 3 Lamp multi colour 9 Socket for (performance monitoring) Silberbauer Point Detector PS3 4 Push-button (Start / Stop) 10 Battery Cap 5 Lamp 11 Laser Warning Label (ready / battery low)
  • Page 9: Initial Use

    Now your Silberbauer Compact - Laser mini is ready to use! 4 Use of the Compact – Lasers CL plus 4.1 Commissioning...
  • Page 10  Compact - Laser Series CL plus Instructions for use  ( see chapters 4.2 and Place the laser perpendicular to the area to be irradiated 4.3). The built-in timer will start immediately after pressing the start button. The Lamp (3) will lit yellow and shows that the laser power has the desired value. A short warning tone resounds.
  • Page 11  Compact - Laser Series CL plus Instructions for use For your information only – it calculates automatically: Therapy time depending of the position of the switches (in minutes : seconds): (for blank cells, time is over 1 hour) 60 mW- Model: Geriatrics Pain therapy new scars...
  • Page 12  Compact - Laser Series CL plus Instructions for use Kind of operation of the indicator lamps: Back lamp green dark If the laser is turned off or if the battery is wholly empty green blinks If the device is turned on slowly and the battery is ok green...
  • Page 13: Laser-Point And Distance Of Irradiation

    L a s e r B e a m The point of the Silberbauer - CL-plus Laser is made of stainless steel. It is created in the shape of a small tube that follows the cone. So it is easy to watch the point where the laser beam comes to the skin.
  • Page 14: Irradiation Angle

     Compact - Laser Series CL plus Instructions for use 4.3 Irradiation Angle To obtain optimal success during treatment, the laser must be positioned perpendicular to the skin's surface. Low reflection, wide reflection, Optimal penetration depth lower penetration depth = optimal success = less success 4.4 Use of the assessory If necessary put the bent light conductor or the power reducing tip onto the...
  • Page 15: Location Of Acupuncture Points With The Compact - Laser Plus

    Silberbauer Point Detector PS3 and a connection cable is required. Put one plug of the connecting cable into the rear socket of the laser. The other cable plug is required to be inserted into the golden socket of the Silberbauer PS3 instead of the point-search-tip.
  • Page 16: Detection Of Ear- And Skull Acupuncture Points

     Compact - Laser Series CL plus Instructions for use 4.5.3 Detection of ear- and skull acupuncture points The acupuncture points at the cranial zone because of Yamamoto and at the ear are “silent“ - in contrast to acupuncture points for the rest of the body.
  • Page 17: Calculation Of The Correct Dose

    Close to this circle you will find a number which shows the size of this area in cm² . Set the size switch at the laser in this position. 4.7 Checking the output power of the laser The Silberbauer Compact - Laser plus has a mechanism for control of the laser output power:...
  • Page 18: Charge Of The Nimh- Battery

     Compact - Laser Series CL plus Instructions for use tone if the power is too high. In this case the laser will switched off automatically and can be switched in after the Rotary Switch 1 is turned into ist off- position. 4.8 Charge of the NiMH- Battery Attention! Don’t try to charge normal one use batteries!
  • Page 19: Possible Dysfunctions Of The Laser

     Compact - Laser Series CL plus Instructions for use 5 Possible dysfunctions of the laser Attention must be paid to the existing danger involved in meddling with a damaged instrument and being exposed to dangerous laser- radiations! If the green lamp (5) doesn’t glow after switching in the laser: ...
  • Page 20: Cleaning And Maintenance

     Compact - Laser Series CL plus Instructions for use 6 Cleaning and maintenance To avoid any danger of infection, the laser point must be disinfected before and after each treatment (and the power-reducing-tip and the bent light conductor). The laser can be cleaned carefully with a cloth moistened with alcohol or surgical disinfection.
  • Page 21: Warranty

    We will provide free replacement of all parts which are defective because of material failure or mistake in production or we will repair them. Guarantee: All recently manufactured Silberbauer products: 3 years All used products: 1 year. Excluded from the warranty are...
  • Page 22: Laser Inspection

    The CL-Laser mini must be inspected at regular intervals of 2 years. The inspection will be made in Vienna by Labor Silberbauer. Please, send the Laser with all accessories and Instruments book! If you want to get a spare device during the time of inspection, pleasse contact us.
  • Page 23: Inspection Certificate

     Compact - Laser Series CL plus Instructions for use The output power should be: in position „Acup. Child“ 25% of the nominal output power, in Position „Acup. Adult“ 50% and in all other positions 100% of the nominal output power. Range of tolerance of the measured power: Nominal Power +/- 5% If the measured power is out of the mentioned range of tolerance, the laser must be calibrated or repaired.
  • Page 24: Warning Advices And Label Information

     Compact - Laser Series CL plus Instructions for use 11 Warning advices and Label Information : For the Laser there are used different warning labels depending of the model: e.g. for the model CL plus 100 - 658: All other models have the same warning labels with other values for power and wavelength.
  • Page 25: Technical Data

    Compact - Laser Series CL plus Instructions for use 12 Technical Data: Laser device, Series Compact - Laser CL plus xx - xxx : Manufacturer and distributor: Prof. Dipl.Ing. Gerhard Silberbauer Medizinische und physikalische Elektronik Hießgasse 15, 1030 Wien, Österreich 0120 Intended use:...
  • Page 26  Compact - Laser Series CL plus Instructions for use...
  • Page 27  Compact - Laser Series CL plus Instructions for use Calibration - Certificate Device: Compact - Laser Model: CL plus ....Serial number: …………………………………….. It is hereby confirmed that the instrument mentioned above, has been examined and calibrated at its time of manufacturing.