5. REMOTE OPERATION:

The following paragraphs provide setup and general guidelines for operating the Model 8310 using an
external controller.
5-1. IEEE-488 INTERFACE BUS CONNECTOR: Joining the Model 8310-1 to a system controller requires the
connection of IEEE-488 control bus cable to the IEEE-488 INTERFACE BUS connector located on the rear panel.
Figure 3 shows the connector's contact pin numbering scheme and lists the signal designator for signal present at
each contact pin.
PIN No.
5-2. GPIB ADDRESS/SERIAL COMMUNICATIONS SETTINGS: The GPIB Bus Address and Serial Communications
options are programmed via an internal 8 position DIP switch SW1 which is located on the rear panel. The switch is
shared between the two functions, with SW1-1 controlling the selection. When SW1-1 is OFF, the remaining switches
set the GPIB bus address. Likewise, when SW1-1 is ON, the switches are used to select the various serial options,
including baud rate, parity, and handshaking. Refer to Figure 4 for switch location.
Aeroflex / Weinschel
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SIGNAL LINE
1
DIO 1
2
DIO 2
3
DIO 3
4
DIO 4
5
EOI (24)**
6
DAV
7
NRFD
8
NDAC
9
IFC
10
SRQ
11
ATN
12
SHIELD
* GND (N) refer to the signal ground return of the referenced pin.
** Return pin on pin 24.
Figure 3. IEEE-488 Interface Bus Pin Locations
PIN No.
SIGNAL LINE
13
DIO 5
14
DIO 6
15
DIO 7
16
DIO 8
17
REN (24)**
18
GND (6)*
19
GND (7)*
20
GND (8)*
21
GND (9)*
22
GND (10)*
23
GND (11)*
24
GND, LOGIC
IM-289A