loading

Summary of Contents for Aeroflex Weinschel 8310

  • Page 1 (217) 352-9330 | Click HERE Find the Cobham / Aeroflex / Weinschel 8310-36-2-F at our website:...
  • Page 2 Operation & Service Manual Model 8310 Programmable Attenuator Units This documentation may not be reproduced in any form, for any purpose unless authorized in writing by Aeroflex / Weinschel, Inc. © Aeroflex / Weinschel, Inc. Frederick, Maryland 1999-2006 Manual Rev. 9-08 Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ...
  • Page 3 SAFETY SUMMARY SAFETY SUMMARY • DEFINITIONS. To minimize shock hazard, the instrument chassis must be connected to an electrical ground. Using The following definitions apply to WARNINGS, the supplied three-conductor power cable ensures CAUTIONS, and NOTES found throughout this manual. that the instrument can be firmly connected to the ac power source and electrical ground at a grounded power outlet.
  • Page 4 SAFETY SUMMARY NOTE CAUTION DO NOT return any instrument or component to Weinschel Corporation without receiving prior • All procedures and/or steps identified as factory authorization. must be followed exactly as written and according to industry accepted ESDS device handling procedures. Failure to comply WILL SAFETY SYMBOLS.
  • Page 5: Table Of Contents

    5-6.2. SYNTAX OF COMMANDS ................................ 20 5-6.3. OUTPUT DATA FORMAT ................................21 5-6.4. NOTATIONAL CONVENTION ..............................21 5-7. 488.2 COMMON COMMANDS ..............................22-23 5-8. GENERAL COMMANDS ................................24-25 Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 6 8310-38-F & -R, ATTENUATOR UNIT ASSEMBLY REPLACEABLE PARTS LIST (P/N 193-7038-1) ....42 8310-38-2-F & -R, ATTENUATOR UNIT ASSEMBLY REPLACEABLE PARTS LIST (P/N 193-7038-2) ....42 Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 7 8310 BASE (+5V) ATTENUATOR UNIT ASSEMBLY REPLACEABLE PARTS LIST (P/N 193-8034-1) ....55 9. APPLICATIONS ..................................56 10. ACCESSORIES ..................................56 11. CONTACTING AEROFLEX / WEINSCHEL........................56 12. AEROFLEX / WEINSCHEL WARRANTY ........................57 13. ASSEMBLY/WIRING DIAGRAMS MODEL 8310-1-X ASSEMBLY DRAWING ..........................193-7001 MODEL 8310-2-X ASSEMBLY DRAWING ..........................
  • Page 8: General Information

    RF signal is routed through either the front or rear mounted Ports A & B connector (Ports A-A if the RF is routed front to rear) to a single or multiple Aeroflex / Weinschel programmable attenuators thus creating CH1 (Channel 1).
  • Page 9: Unpacking And Inspection

    Weinschel Corporation, giving equipment model and serial numbers, so that necessary action can be taken. Under U.S. shipping regulations, damage claims must be collected by the consignee; DO NOT return the equipment to Aeroflex / Weinschel Corporation until a claim for damages has been established.
  • Page 10: Related Manuals

    Programmable Attenuator Unit. 1-12. POWER REQUIREMENTS: Aeroflex / Weinschel supplies a detachable power cable (P/N 068-21) to connect an 100 to 240 Vac power source with a frequency between 50 to 60 H to the Attenuator Unit.
  • Page 11: Specifications

    5. Refer to Individual data sheet (Appendix C) for detailed specifications on internal programmable attenuators. 2-2. PHYSICAL DIMENSIONS: NOTE: All dimensions are given in mm (inches) and are maximum, unless otherwise Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 12: Configurations\Rf Specifications

    3.25 dB max 1.95 150T-62 Sma/F Front 193-7204-1 8310-204-2-R 62/2 dc-18 3.25 dB max 1.95 150T-62 Sma/F Rear 193-7204-1 Table 1, Continued on next page Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 13: Installation

    Setup the IEEE-488 bus address or RS-232/422 Communications options for your application using paragraphs 5-2. If using the RS-422 serial configure the selectable signal pair terminations using paragraph 5-4. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 14: Input/Output Options

    These connectors provide a input and output port where various types of RF signals can be applied to the devices internally mounted in the Model 8310 (Connector location specified by customer when ordering). Some special configurations could contain Aeroflex / Weinschel’s Model 1568 SMA Panel Adapters or other types of crowns (see accessories for other types).
  • Page 15: Front Panel Controls & Indicators

    ® For more information about the PLANAR CROWN connectors contact the Sales Department at Aeroflex / Weinschel. 4. FRONT PANEL CONTROLS & INDICATORS: The following paragraphs provide setup and general guidelines for operating the 8310 Series Programmable Attenuator Unit and its different bus configurations. Also provided is a general description of the internal circuitry of the 8310.
  • Page 16: Remote Operation

    GPIB bus address. Likewise, when SW1-1 is ON, the switches are used to select the various serial options, including baud rate, parity, and handshaking. Refer to Figure 4 for switch location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 17 This switch is factory set to 10. IEEE-488 Address Truth Table Switch Number Decimal Weight Address: Figure 4. Internal Dip Switch Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 18: Ieee-488 (Gpib) Bus Operation

    The 8310 program message units may be divided into two syntax groups: commands and queries. Refer to the section on command syntax for more information. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 19: Serial Operation

    For those systems incorporating local front panel controls, the serial port can lockout local users, providing a Remote/ Local function similar to that of GPIB operation. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 20: 5-4.1. Rs-232 Operation

    The DTR signal is asserted when power is on, indicating that the unit is ready. RS232 9-pin DB9 Pinout Null Modem Cable Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 21: 5-4.2. Rs-422/485 Operation

    CTS- Clear To Send input RTS+ Request To Send output RS422 9-pin DB9 Pinout RXD+ CTS+ RXD- RTS - TXD- CTS - TXD+ RTS+ Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 22: Status Reporting

    Request Control: The device is requesting control (not implemented) Operation Complete: This bit is generated in response to an *OPC command. It indicates that the ITS 2000 has completed all pending operations. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 23: General Syntax Structure

    ’0’-’9’, and the letters ’A’-’F’, preceded by the characters ’#H’ or the C language style prefix ’0x’. Examples: 123 (integer) #H55 (hex) 0xAA (hex) #B1010 (binary) Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 24: 5-6.3. Output Data Format

    Ellipses indicate that the preceding argument or parameter may be repeated. [,...] The preceding item may be repeated, but each repetition must be separated by a comma. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 25: 488.2 Common Commands

    Return Value: *IDN? returns the following ’Weinschel,8310 Series, 123, 1.00A’ Example(s): Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 26 This function prevents the8310 Series from executing any further commands or queries until there Remarks: are no pending operations. The8310 Series executes all commands sequentially, and does not allow overlapping commands. none Return Value: *WAI Example(s): Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 27: General Commands

    This function reads the reference setting of the currently selected channel. Remarks: refatten real reference attenuation value, in dB Return Value: REF? returns ’30.00’ Example(s): Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 28 This function decrements the attenuation setting of the current channel by the value of the Remarks: attenuator’s programmed stepsize (see STEPSIZE command). none Return Value: DECR Example(s): Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 29: General Circuit Description

    Standard 8310 Series Programmable Attenuator Units contain internal circuitry to control a variety of Aeroflex / Weinschel programmable attenuators (3200T, 150T, and 4200 Series) via front panel controls or standard communications interfaces including GPIB (IEEE-488) and RS-232/RS-422 /RS485. Figure 5 provides a simplified block diagram of the 8310’s internal circuitry.
  • Page 30: Rf Section

    Typically Aeroflex / Weinschel programmable attenuators are bi-directional and the RF signal can be applied to either Port connector. Refer to the individual data sheets located in the Appendix C for detailed and specific specifications pertaining to the internal mounted attenuators.
  • Page 31: Maintence

    When parts are thoroughly clean, dry parts using 5 psi of clean moisture free compressed air or preferably dry nitrogen (pressurized spray will work well). Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 32: 7-3.3. Machined Surfaces And Hardware

    MIL-C-83112 is recommended. Exercise care to ensure no metal filing or residue remains inside the connector and the connector is thoroughly dry. Where rust, corrosion, and/or oxide deposits are present in large quantities, replace the connector. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 33: Line Voltage Fuse Replacement

    Remove defective fuse and replace with the correct fuse (Refer to applicable parts list for fuse part number). e. Snap the Fuse Drawer shut and re-connect ac power cord. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 34: Replacable Parts List

    (this information is on the component side of the PCB). 8-3 DRAWING NUMBER: The Aeroflex / Weinschel part number consists of a basic number with a dash number. The basic number should cross reference to a drawing number for most of the items. For those items that do not have a drawing number, the manufacturers part number is provided.
  • Page 35: 8-4.6. Assembly And Componet Location

    WIRING DIAGRAM 089-3249 ICD, ATTEN UNIT, MODEL 8310, 3250 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Weinschel Drawing 193-7001 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 36: 8310-1-3-T, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7001-3)

    WIRING DIAGRAM 089-3249 ICD, ATTEN UNIT, MODEL 8310, 3250 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Weinschel Drawing 193-7001 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 37: 8310-2-F & -R, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7002-1)

    WIRING DIAGRAM 089-3249 ICD, ATTEN UNIT, MODEL 8310, 3250 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Weinschel Drawing 193-7002 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 38: 8310-35-F & -R, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7035-1)

    ICD, ATTEN UNIT, MODEL 8310, 3200 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7035 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 39: 8310-35-3-T, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7035-3)

    ICD, A TTEN UNIT, MODEL 8310, 3200TY PE IM-289 O & S MA NUA L, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7035 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 40: 8310-35-F-E & -R-E Attenuator Unit Assembly Replaceable Parts List (P/N 193-7035-5)

    089-3244 ICD, ATTEN UNIT, MODEL 8310, 3200TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7035 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 41: 8310-36-F & -R, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7036-1)

    ICD, ATTEN UNIT, MODEL 8310, 3200 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7036 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 42: 8310-36-3-T, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7035-3)

    ICD, ATTEN UNIT, MODEL 8310, 3200 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7036 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 43: 8310-37-F & -R, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7037-1)

    WIRING DIAGRAM 089-3244 ICD, ATTEN UNIT, MODEL 8310, 3200 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Weinschel Drawing 193-7037 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 44: 8310-37-3-T, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7037-3)

    089-3244 ICD, ATTEN UNIT, MODEL 8310, 3200TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7037 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 45: 8310-38-F & -R, Attenuator Unit Assembly Replaceable Parts List (P/N 193-7038-1)

    ICD, ATTEN UNIT, MODEL 8310, 3200 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7038 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 46 089-3244 ICD, ATTEN UNIT, MODEL 8310, 3200TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7038 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 47 ICD, ATTEN UNIT, MODEL 8310, 3200 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7038 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 48 ICD, ATTEN UNIT, MODEL 8310 (+5 V) IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7136 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 49 ICD, ATTEN UNIT, MODEL 8310 (+5 V) IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7137 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 50 ICD, ATTEN UNIT, MODEL 8310 (+5 V) IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7138 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 51 089-3422 ICD, ATTEN UNIT, MODEL 8310 (+5V) IM289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7138 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 52 ICD, ATTEN UNIT, MODEL 8310, 150 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7201 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 53 ICD, ATTEN UNIT, MODEL 8310, 150 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7202 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 54 ICD, ATTEN UNIT, MODEL 8310, 150 TYPE IM-289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7204 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 55 ICD, ATTEN UNIT, MODEL 8310-352-X-X DC - 6 GHz IM289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7403 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 56 ICD, ATTEN UNIT, MODEL 8310-352-X-X DC - 6 GHz IM289 O & S MANUAL, MODEL 8310 Refer to Aeroflex / Weinschel Drawing 193-7403 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 57 NUT HEX #4 MS51957-15 SCR PAN HD 4-40 X 3/8 LG MS51859-2 WASHER FLAT #4 NYLON Refer to Aeroflex / Weinschel Drawing 193-8034 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 58 NUT HEX #4 MS51957-15 SCR PAN HD 4-40 X 3/8 LG MS51859-2 WASHER FLAT #4 NYLON Refer to Aeroflex / Weinschel Drawing 193-8034 for parts location. Aeroflex / Weinschel Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 59 11. CONTACTING AEROFLEX / WEINSCHEL: In the event of a malfunction, contact Aeroflex / Weinschel. An apparent malfunction of an instrument or component may be diagnosed over the phone by first contacting the Customer Service Department at Aeroflex / Weinschel.
  • Page 60 12. AEROFLEX / WEINSCHEL WARRANTY: PRODUCTS - Aeroflex / Weinschel warrants each product it manufactures to be free from defects in material and workmanship under normal use and service anywhere in the world. Aeroflex / Weinschel’s only obligation under this Warranty is to repair or replace, at its plant, any product or part thereof that is returned with transportation charges prepaid to Aeroflex / Weinschel by the original purchaser within ONE YEAR from the date of shipment.
  • Page 61 APPENDIX A APPENDIX A CARE AND HANDLING OF MICROWAVE COAXIAL CABLE ASSEMBLIES A-1 CARE AND HANDLING OF ASSEMBLIES. A-3 MAKING INITIAL CONNECTIONS. To ensure accurate measurements and optimal Exercise caution when mating cables. Poor connections performance of Weinschel products, the microwave coaxial lead to poor system performance.
  • Page 62 CARE & HANDLING OF COAXIAL CABLES A-4.2 HEX-NUT TYPES. To mate a connector of the A-5.1 SECURING CABLES. Use toothed, rubber-lined hex-nut type, always use a torque wrench set to the correct "P-clamps" to hold cables in place. If it is necessary to use torque value.
  • Page 63 All commands are sequential Functional criteria met with an operation complete Command is finish executing message is generated in response to that command Aeroflex / Weinschel Revised 8-30-03 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 64 Appendix C Internal Device Data Sheets Click on link below to get internal device data sheet... • 150 Series Programmable Attenuators...dc-18/26.5 GHz • 3200 Series Programmable Attenuators...dc-2/3 GHz • 3250 Series 75Ω Series Programmable Attenuators...dc-1.5 GHz • 4226 & 4228 Series Solid-State Attenuators Artisan Technology Group - Quality Instrumentation ...
  • Page 65 5/04 Incorporated Aeroflex name change. All Units 9/08 Incorporated Aeroflex / Weinschel ECN 08-295 All units built after 9-4-08 1/10 Incorporated Aeroflex / Weinschel ECN 10-009 8310-204-F Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 66 EC DECLARATION OF CONFORMITY This is to certify that the: SmartStep Attenuator Unit 8310 Series Manufactured by: Weinschel Corporation Conforms with the protection requirements of Council Directive 89/336/EEC and the Amending Directive 93/68/EEC, relating to Electromagnetic Compatibility and Council Directive 73/23/EEC relating to LVD (Safety), by the application of: EN50081-1: 1992 Generic Emissions Standard.

Table of Contents