Suez Sievers 500 RLe Operation And Maintenance Manual

Online toc analyzer
Table of Contents

Advertisement

www.sieversinstruments.com
Water Technologies & Solutions
e
Sievers 500 RL
Online TOC Analyzer
operation and maintenance manual
2.04 firmware or later
DLM 74300-05 EN Rev. A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sievers 500 RLe and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Suez Sievers 500 RLe

  • Page 1 www.sieversinstruments.com Water Technologies & Solutions Sievers 500 RL Online TOC Analyzer operation and maintenance manual 2.04 firmware or later DLM 74300-05 EN Rev. A...
  • Page 2 Printed in the USA © 2018 6060 Spine Road Boulder, CO 80301 USA www.sieversinstruments.com For technical support, contact Technical Support or your local service provider. In North America, contact Technical Support at +1 888 245 2595. In Europe, the Middle East, and Africa, contact Technical Support at +44 161 864 6800.
  • Page 3: Identification Records

    Default Administrator Password: SIEVERS Analyzer serial number: _______________________ (This is located on the side of the Analyzer.) Date of receipt and installation of Analyzer: _______________________ (This is the warranty start date.) SUEZ ©2018 3 of 162 DLM 74300-05 EN Rev. A...
  • Page 4: Table Of Contents

    Step 3: Select a Location for the Analyzer ..........43 SUEZ ©2018 4 of 162 DLM 74300-05 Rev.
  • Page 5 Printing Data History ............. . 70 SUEZ ©2018 5 of 162 DLM 74300-05 Rev.
  • Page 6 Disabling Password Protection ............94 SUEZ ©2018 6 of 162 DLM 74300-05 Rev.
  • Page 7 Cleaning the Analyzer ............. . 133 SUEZ ©2018 7 of 162 DLM 74300-05 Rev.
  • Page 8 Step 5: Return the Analyzer to SUEZ ........
  • Page 9: List Of Tables

    Analyzer 3-Year Maintenance Worksheet ........134 SUEZ ©2018 9 of 162 DLM 74300-05 EN Rev.
  • Page 10: List Of Figures

    UV Lamp Ferrule Orientation Detail ......... . 125 SUEZ ©2018 10 of 162 DLM 74300-05 EN Rev.
  • Page 11 Left-Side Analyzer Dimensions ..........155 SUEZ ©2018 11 of 162 DLM 74300-05 EN Rev.
  • Page 12: Revision History

    Confidentiality The information contained in this manual may be confidential and proprietary and is the property of SUEZ. Information disclosed herein shall not be used to manufacture, construct, or otherwise reproduce the goods disclosed herein. The information disclosed herein shall not be disclosed to others or made public in any manner without the express written consent of SUEZ.
  • Page 13: Standard Limited Warranty

    (Mountain Time), excluding all company and legal holidays. Telephone support is provided for troubleshooting and determination of parts to be shipped from SUEZ to the customer in order to return the product to operation. If telephone support is not effective, the product may be returned to SUEZ for repair or replacement. In some instances, suitable instruments may be available for short duration loan or lease.
  • Page 14: Limitation Of Remedies And Liability

    The Operation and Maintenance Manual is believed to be accurate at the time of publication and no responsibility is taken for any errors that may be present. In no event shall SUEZ be liable for incidental or consequential damages in connection with or arising from the use of the manual and its accompanying related materials.
  • Page 15: Beschränkte Ansprüche Und Haftung

    être engagée pour les éventuelles erreurs qu’il pourrait contenir. Le groupe SUEZ ne saurait en aucun cas être tenu pour responsable des préjudices accidentels ou de quelque nature que ce soit, dus à l’utilisation du manuel ou de la documentation connexe. La garantie ne s’applique qu’à l’acquéreur d’origine.
  • Page 16: 限定責任および救済

    限定責任および救済 前述の保証は、お客様による不適切または不十分な取り付け、保守、調整、校正、あるいは操作によって 生じた不具合には適用されないものとします。 取り付け、保守、調整、校正、または操作は、 『SUEZ ( 操 作取扱説明書 )』に記載されている指示に従って行う必要があります。 推奨外の保守資材を使用すると、 保証請求が無効になる場合があります。 ここで提供される救済は、お客様の唯一の排他的救済となります。 SUEZ は、いかなる場合においても、直 接的、間接的、特別的、付随的、または派生的損害(利益の逸失を含む)に対し、それが契約、不法行為、 またはその他の法的理論に基づくものであるかどうかにかかわらず、 一切責任を負いません。 『Operation and Maintenance Manual ( 操作取扱説明書 )』は、出版された時点で正確であるものと考えられており、 万が一発生した誤りに対する責任は一切負いません。 SUEZ は、いかなる場合においても、マニュアルまた はそれに付属の関連資料の使用に関連して、またはその使用が原因で発生した付随的または派生的損害に も一切責任を負いません。 保証は最初の購入者に対してのみ有効です。 本限定保証を、 SUEZ の書面による 同意なしに、最初の購入者から第三者に譲渡することはできません。 SUEZ は、商品性および特定の目的 に対する適合性の黙示の保証を一切拒否します。 赔偿与责任限制 上述保证不适用于因客户不正确或不恰当的安装、维护、调整、校准或操作导致的故障。安装、维护、调 整、校准或操作必须遵循操作与维护手册中的说明进行。使用非推荐的维护材料可能会导致保证失效。...
  • Page 17: Warnings

    This symbol on the instrument indicates that the user should refer to the manual for instructions on filling the DI water cartridge, as found in the Installation and Maintenance chapters. SUEZ ©2018 17 of 162 DLM 74300-05 EN Rev. A...
  • Page 18 Always “clean-up” the Analyzer by running low-TOC DI water after running high TOC or salt samples. Warning If this instrument is used in a manner not specified by SUEZ, the protection provided by the instrument may be impaired. SUEZ ©2018 18 of 162 DLM 74300-05 EN Rev.
  • Page 19 Advertencia Este símbolo indica el terminal de protección a tierra (masa) del analizador. SUEZ ©2018 19 of 162 DLM 74300-05 EN Rev. A...
  • Page 20 Como protección frente a la exposición accidental a la radiación ultravioleta, no maneje la lámpara UV fuera de su alojamiento protector. Advertencia Si este instrumento se utiliza de una manera no especificada por SUEZ USA, la protección ofrecida por el instrumento puede verse reducida. Advertencia Este producto es de clase de seguridad I.
  • Page 21 Para proteger el Analizador de deterioros, siempre drene el agua del mismo antes de su envío. Français Avertissement Ce symbole placé sur l’instrument indique que l’utilisateur doit se rapporter au manuel pour les instructions de fonctionnement. SUEZ ©2018 21 of 162 DLM 74300-05 EN Rev. A...
  • Page 22 Avertissement Si la lampe UV venait à être cassée ou endommagée, elle devrait être remplacée conformément à la procédure en vigueur dans votre entreprise pour le remplacement de produits toxiques. SUEZ ©2018 22 of 162 DLM 74300-05 EN Rev. A...
  • Page 23 éviter tout risque de blessure, s'assurer que l'analyseur est connecté à une ligne d'alimentation à la tension adéquate. Avertissement Si cet instrument est utilisé de manière non conforme à ce qui est spécifié par le groupe SUEZ aux USA, la protection fournie par l’instrument risque d’être réduite. Avertissement Arrêtez toujours l’analyse avant de mettre l’analyseur hors tension ou de le débrancher.
  • Page 24 Jeder Betrieb, der Zugriff auf das Innere des Analysegeräts erfordert, einschließlich Installation von Wartungsteilen, kann zu Verletzungen führen. Um gefährliche Verletzungen durch einen Stromschlag zu vermeiden, schalten Sie den Strom ab und trennen Sie, wenn möglich, das Analysegerät vor dem Öffnen vom Stromnetz. SUEZ ©2018 24 of 162 DLM 74300-05 EN Rev. A...
  • Page 25 Schäden am Analysator und die Gefahr einer Verletzung zu verhindern, achten Sie darauf, dass der Analysator an eine Steckdose mit der richtigen Spannung angeschlossen wird. Warnung Wenn dieses Instrument in einer Art und Weise verwendet wird, die nicht von SUEZ USA festgelegt ist, kann der durch dieses Instrument gebotene Schutz beeinträchtigt werden. Warnung Beenden Sie stets die Analyse, bevor Sie das Analysegerät ausschalten oder den Stecker herausziehen.
  • Page 26 Questo simbolo indica il terminale di messa a terra protettivo dell'Analizzatore. Avvertenza Questo simbolo indica che per conformarsi alla Direttiva europea 2002/96/CE sui rifiuti provenienti da apparecchiature elettriche ed elettroniche (WEEE), l'analizzatore deve essere smaltito separatamente dai normali rifiuti. SUEZ ©2018 26 of 162 DLM 74300-05 EN Rev. A...
  • Page 27 Avvertenza Se lo strumento viene utilizzato in modo diverso da quello specificato da SUEZ USA, la protezione fornita dallo strumento può risultare compromessa. Avvertenza Per proteggersi da esposizione accidentale a radiazioni ultraviolette, non utilizzare la lampada UV al di fuori del relativo alloggiamento di protezione.
  • Page 28 日本語 警告 機器上のこの記号は、ユーザーがマニュアルの操作手順を参照すべきであることを示します。 警告 iOS システムおよびバイアルポートには、サンプルバイアルの隔膜に刺し通すための 2 本の先の鋭い ニードルが含まれています。 指や不適切な物質を iOS システムやバイアルポートに入れないようにし てください。 警告 500 オンライン分析装置およびポータブル分析装置では、iOS システム内の水が加熱している場合が あります。 バイアルを iOS システムに挿入する前に、ドアをスライドして開き、30 秒間待ち、サン プルが完全に排水されるようにしてください。 排水前にバイアルを挿入すると、iOS システムから熱 水が噴出するおそれがあります。 SUEZ ©2018 28 of 162 DLM 74300-05 EN Rev. A...
  • Page 29 保守部品を取り付ける際など、 分析装置の内部にアクセスする操作では、 負傷につながるおそれがあります。 感電 を避けるために、分析装置を開く前に電源を切り、可能であれば、電源コードを抜いてください。 警告 UV ランプおよび表示画面には水銀が含まれており、地域によっては危険物質と見なされる場合があります。 これ らの部品は、連邦、州、または地方自治体の規制に従って破棄してください。 警告 UV ランプが壊れたり、損傷したりした場合には、組織の有毒廃棄物処理手順に従って処理し、連邦、州、または 地方自治体の規制に従って破棄してください。 警告 不測の紫外線放射被爆を防ぐために、UV ランプは保護ハウジングから出して操作しないでください。ランプのハ ウジングは熱くなる場合があるため、ハウジングの扱いには十分に注意してください。 警告 これは安全基準クラス I の製品です。 アース付き電源に接続する必要があります。 分析機器の破損と人体傷害を回 避するため、分析機器を正しい電圧の電力線に接続するようにしてください。 警告 SUEZ USA が指定した方法以外の方法でこの装置を使用すると、装備されている保護機能が作動しないおそれがあ ります。 SUEZ ©2018 29 of 162 DLM 74300-05 EN Rev. A...
  • Page 30 TOC 値の誤った読み取りおよび分析装置の損傷を避けるために、 サンプル注入口が開いており、 DI 貯水槽が満杯に なっていることを必ず確認してから分析を開始してください。 バイアルからサンプル吸入を行っていないときは、 標準 iOS システムのドアを必ず締めてください。 警告 適切な廃棄方法については、各標準セットや試薬についている製品安全データシートをご覧ください。 警告 ランプが正常に動作しているかどうかは、UV 反応器のガラス製のウィンドウで確認できます。目を近づけすぎた り (15cm 以下 )、長時間 (1 分以上 ) 見つめたりしないでください。保護用眼鏡の装着をお勧めします。 警告 適切な廃棄処理方法については、各標準液セットに付随する MSDS を参照してください。 中文 警告 仪器上标有此符号表示用户应参考手册上的操作说明。 警告 试剂瓶槽含有锐利的针头,用于刺穿样品试剂瓶的封口膜。请不要将手指或其它不适当的物品放入试剂瓶 槽。 SUEZ ©2018 30 of 162 DLM 74300-05 EN Rev. A...
  • Page 31 在 500 联机分析仪中,此符号表示分析仪的保护性接地端(地线)。 警告 这个符号表示根据欧盟废电机电子设备 (WEEE) 指令 2002/96/EC,分析仪应以不同于普通废品的方式 处理。 警告 任何需要接触分析仪内部的操作,包括安装维修件,均可能导致人身伤害。为避免可能的电击伤害,在打开分析仪 之前,请关闭电源开关并断开仪器与电源的连接(如果可能)。 警告 仪器上的此符号表示,用户应按手册上的说明灌注去离子水盒,请参阅手册的 “ 安装 ” 和 “ 维护 ” 章节。 警告 紫外线灯和显示屏含有水银,在您所在的地区可能被视为危险材料。处理这些材料时,请遵循国家、州/省或地方 政府的相关规定。 警告 如果紫外线灯破损或损坏,应根据您所在组织机构的有毒废料处理程序以及国家、州/省或地方政府的相关规定进 行处理。 警告 为预防在紫外线放射下暴露导致伤害,请勿将紫外线灯置于保护罩之外。紫外线灯罩可能烫手,接触灯罩时应格外 小心。 SUEZ ©2018 31 of 162 DLM 74300-05 EN Rev. A...
  • Page 32 警告 本产品为 I 类安全产品。本产品必须连接具有接地端的电源。电源电压应正确无误,以免损坏分析仪或对人体造成 伤害。 警告 如果本仪器未按照 SUEZ USA 的规定使用,仪器提供的保护措施可能会失效。 警告 在关闭或断开分析仪电源之前,必须停止分析。 警告 请确保 DI 水容器中装满水,特别是在测试高 TOC 或高盐浓度的试样时尤其如此。在测试完高 TOC 或高盐浓度的试样 后,请务必使用低 TOC 的 DI 水清洗分析仪。 警告 本产品为 A 类产品。在家庭环境中,本产品可能导致电磁干扰,用户可能需要采取适当措施以减少干扰。 警告 为避免 TOC 读数错误或损坏分析仪,在开始分析前必须保证试样进口打开且 DI 水容器已满。当不从样瓶取样时, 应关闭标准 iOS 系统的门。 警告 紫外线反应器中的玻璃窗显示灯的正常工作,请勿过近距离 (离眼睛小于 6 英寸)或过长时间 (超过 1 分钟)观察...
  • Page 33: Chapter 1: Introduction

    The Sievers 500 RL On-Line TOC Analyzer from SUEZ is a high-sensitivity Analyzer used to measure the concentration of total organic carbon (TOC), total inorganic carbon (TIC), and total carbon (TC = TOC + TIC) in water samples (for patent information, see the section called “Trademarks and Patents”...
  • Page 34 Special conditions in operating environments may require modifications to the recommendations in this manual. Specific instructions provided by Technical Support or your local service provider supersede the information provided in this manual. SUEZ ©2018 34 of 162 DLM 74300-05 EN Rev. A...
  • Page 35: Chapter 2: System Description

    Chapter 2. System Description System Specifications System specifications for the Sievers 500 RL On-Line TOC Analyzer are on the following page. SUEZ ©2018 35 of 162 DLM 74300-05 Rev. A...
  • Page 36: System Overview

    Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors. System Overview The 500 RL On-Line TOC Analyzer consists of five major subsystems: 1. Sample inlet system and sample pump, including the Integrated On-Line Sampling (iOS) System 2. Oxidation reactor SUEZ ©2018 36 of 162 DLM 74300-05 Rev. A...
  • Page 37: Figure 1 Analyzer Schematic (Shown With Standard Ios)

    DI water pump 5. Electronics subsystems, comprising: • Microprocessors and circuit boards • Data outputs Figure 1: Analyzer Schematic (Shown with Standard iOS) A brief description of each major Analyzer component follows. SUEZ ©2018 37 of 162 DLM 74300-05 Rev. A...
  • Page 38: Sample Flow Path

    The CO forms carbonic acid upon reaction with water, and the carbonic acid disassociates into hydrogen ions and bicarbonate ions:   H + HCO SUEZ ©2018 38 of 162 DLM 74300-05 Rev. A...
  • Page 39: Additional System Components

    The Analyzer cannot be directly connected to a computer via the USB port. Rather, the Analyzer’s USB port can only be connected to a USB storage device, such as a USB flash memory drive. SUEZ ©2018 39 of 162 DLM 74300-05 Rev. A...
  • Page 40 Chapter 2: System Description SUEZ ©2018 40 of 162 DLM 74300-05 Rev. A...
  • Page 41: Chapter 3: Installation

    There will be a charge if the Analyzer must be repackaged for return shipment. Insure the Analyzer for the return shipment. Note: Do not start the Analyzer until the DI water cartridge has been filled with DI water. SUEZ ©2018 41 of 162 DLM 74300-05 EN Rev. A...
  • Page 42: Step 1: Unpack And Inspect The Analyzer

    Sievers 500 RL On-Line TOC Analyzer • Sievers 500 RL On-Line TOC Analyzer Operation and Maintenance Manual, downloaded from our website or print copy purchased separately from SUEZ. • Sievers 500 RL On-Line TOC Analyzer Quick Start Guide • Warnings Sievers 500 RL and 500 RL On-Line TOC Analyzers document •...
  • Page 43: Additional Installation Equipment

    (10 °C) can cause errors in the measurements. The Analyzer withstands the hazards of industrial process environments, and the Analyzer enclosure is rated IP 45. SUEZ ©2018 43 of 162 DLM 74300-05 EN Rev. A...
  • Page 44: Step 4: Install Power And Control Wiring

    After connecting the wire, pull on each connection gently to make sure the connection is secure. SUEZ ©2018 44 of 162 DLM 74300-05 EN Rev. A...
  • Page 45: Installing The Analog Outputs And Alarms

    Completely loosen the screw for each pin you are connecting, insert the wire, tighten the screw, and then gently pull on each connection to make sure the connection is secure. Then, replace the terminal blocks. SUEZ ©2018 45 of 162...
  • Page 46: Table 1 Alarm Outputs (Tb2)

    Output Reserved Reserved Alarm 4 (NO*) Alarm 4 (NC*) Alarm 4 (Common) Alarm 3 (NO*) Alarm 3 (NC*) Alarm 3 (Common) * NC = normally closed, NO = normally open SUEZ ©2018 46 of 162 DLM 74300-05 EN Rev. A...
  • Page 47: Wiring The 4-20 Ma Connection

    Max. Load 600 ohms 4-20 mA Output No external power required Isolated self power Voltage compliant to 20 V Rload Recorder Figure 3: Wiring Diagram for the 4-20 mA Connection SUEZ ©2018 47 of 162 DLM 74300-05 EN Rev. A...
  • Page 48: Wiring The Remote Start (Binary Input) Connection

    9 (+24 V) 8 (-) 7 (+) Conduit Internal to Analyzer Local Control External to Analyzer K1 = Custom Computer Terminal Figure 4: Wiring Option for Binary Input Using Analyzer’s Internal Supply SUEZ ©2018 48 of 162 DLM 74300-05 EN Rev. A...
  • Page 49: Wiring The Ethernet Cable

    Before installing the printer cable, USB flash drive, or serial cable, you must first remove the cover plate from the Analyzer. Loosen the two thumb screws on the cover plate and remove the plate. SUEZ ©2018 49 of 162 DLM 74300-05 EN Rev. A...
  • Page 50: Installing The Printer

    ® ® must be IBM compatible or Epson FX 850 compatible. SUEZ sells Citizen and Epson (for China and Japan regions) printers for use with the Analyzer. To install the printer, follow these steps: 1. Connect the printer’s power converter to a grounded power source and plug the cable in to the back of the printer.
  • Page 51: Installing The Serial Connection

    The serial communication is supported only for computers running the Microsoft Windows operating system. Note: For long distance connections (up to 1,000 feet or 305 meters), SUEZ recommends use of either a RS-232 current loop converter or a RS-422 & converter. Companies such as B B Electronics manufacture both kinds of converter.
  • Page 52: Step 7: Fill The Di Water Cartridge

    2. Open the Analyzer’s front panel (you will need the key from the accessories kit to open the bottom latch). 3. Locate the DI water cartridge at the center of the Analyzer (see Figure 7). SUEZ ©2018 52 of 162 DLM 74300-05 EN Rev. A...
  • Page 53: Figure 7 Locating The Di Water Cartridge

    To loosen the fitting, use the John Guest fitting removal tool from the Analyzer Accessories Kit to push up on the fitting (not like a wrench to turn the fitting). Place one end of the SUEZ ©2018 53 of 162...
  • Page 54: Figure 9 Removing The John Guest Fitting

    DI water cartridge. If large air bubbles are present, tap or manipulate the tubing to work the air bubbles out. 12. Keep the Analyzer powered on for 2 minutes. SUEZ ©2018 54 of 162 DLM 74300-05 EN Rev. A...
  • Page 55: Step 8: Configure Basic Analyzer Settings

    On-Line TOC Analyzer offers two levels of security, one included with the Analyzer and the other available for purchase as an upgrade from SUEZ. The Password Protection feature is included with the Analyzer and provides a basic level of security. The Sievers DataGuard feature is available as an upgrade from SUEZ and provides support for the electronic signature regulation 21 CFR Part 11.
  • Page 56: Enabling Password Protection

    Chapter 3: Installation If you purchased the DataGuard firmware feature from SUEZ or will be using basic Password Protection, enable the feature now. If you will not be using DataGuard or Password Protection, proceed to page 58. Enabling Password Protection If you will be using basic Password protection rather than DataGuard, activate the password now by following the steps below.
  • Page 57: Establishing A New Administrator Account For Dataguard

    If you have purchased the optional System Suitability and Linearity Protocols, follow these steps: 1. Insert the USB flash drive containing the Suitability activation key into the Analyzer’s USB port. SUEZ ©2018 57 of 162 DLM 74300-05 EN Rev. A...
  • Page 58: Setting The Clock And Time Zone

    4. Press the Location button. Use the keypad to specify a name for the Analyzer and press Enter . Press the Number button and then the Alpha button to toggle between numbers and letters. 5. Press the Back button twice to return to the Maintenance tab. SUEZ ©2018 58 of 162 DLM 74300-05 EN Rev. A...
  • Page 59: Setting The Analyzer Mode

    If you installed a printer in “Step 8: Configure Basic Analyzer Settings,” configure the printer port to match the printer. 1. Select the I/O tab and press the Printer button. SUEZ ©2018 59 of 162 DLM 74300-05 EN Rev. A...
  • Page 60: Exporting And Printing System Settings

    Note: If the TOC level in your sample water is typically below 50 ppb, schedule a regular TOC Autozero. See “Programming the TOC Autozero” on page 86 for details. SUEZ ©2018 60 of 162 DLM 74300-05 EN Rev. A...
  • Page 61: Step 9: Rinsing The Analyzer

    1 sets the Autozero for once every 24 hours at the specified time. 5. Use the date and time buttons to specify the start date and time for the first TOC Autozero. SUEZ ©2018 61 of 162...
  • Page 62 Chapter 3: Installation SUEZ ©2018 62 of 162 DLM 74300-05 EN Rev. A...
  • Page 63: Chapter 4: Basic Analyzer Operation

    System, not all menu options may be relevant or available on your Analyzer. When the Analyzer is first powered on, the Startup screen displays basic SUEZ contact information and the firmware version number. The Analyzer then begins initialization to check the status of consumables and prepare for taking measurements.
  • Page 64: Figure 10 The Sievers 500 Rl E On-Line Toc Analyzer Main Screen

    3. The Status area displays information about the current operation mode, the Analyzer’s current status (Initializing, Idle, Analyzing, etc.), a countdown timer showing the time remaining in the current analysis, and the repetition and rejection count, if applicable. SUEZ ©2018 64 of 162 DLM 74300-05 EN Rev. A...
  • Page 65: Taking Toc Measurements

    On-Line TOC Analyzer can operate in up to four different modes, depending on your configuration. To set the Analyzer mode, follow these steps: 1. Select the Setup tab (see Figure 11). SUEZ ©2018 65 of 162 DLM 74300-05 EN Rev. A...
  • Page 66: Grab Mode Measurement Settings

    When selecting Grab mode (available on configurations with an iOS), you need to specify the number of replicate measurements and rejects as follows: 1. On the Setup tab, press the Configure Mode button. SUEZ ©2018 66 of 162 DLM 74300-05 EN Rev. A...
  • Page 67: Managing Data History

    Figure 12: The Data Tab Setting up Data History The data history can accomodate approximately 33,000 entries (at least 90 days of usage). 1. Select the Data tab. 2. Press the Setup History button. SUEZ ©2018 67 of 162 DLM 74300-05 EN Rev. A...
  • Page 68: Archiving Data History

    4. Press the Values button and select one of these options: • TOC/Mode — Displays TOC data and the mode only. • TOC/IC/TC — Displays TOC, IC, and TC data. SUEZ ©2018 68 of 162 DLM 74300-05 EN Rev. A...
  • Page 69: Graphing Data History

    3. Press the Setup button to specify the scale for the graph. 4. Press the Type button to specify which data are displayed on the graph: TOC only or TOC/IC/TC. SUEZ ©2018 69 of 162 DLM 74300-05 EN Rev. A...
  • Page 70: Printing Data History

    3. Press the Range button to specify a preset time range: 1 Day , 1 Week , All , or Custom . • If you select a Custom range, press the To Date and From Date buttons and enter the desired time range values. SUEZ ©2018 70 of 162 DLM 74300-05 EN Rev. A...
  • Page 71: Exporting Data History

    Setting Up Analyzer Input and Output Menu selections under the I/O tab allow you to configure data for alarms, analog output, binary input, serial output and printing. SUEZ ©2018 71 of 162 DLM 74300-05 EN Rev. A...
  • Page 72: Setting Up Analog Output

    4. Press the Back button. 5. Press one of the Analog Out buttons, either Analog Out 1 , Analog Out 2 , or Analog Out 3 . Set the values for the following: SUEZ ©2018 72 of 162 DLM 74300-05 EN Rev. A...
  • Page 73: Adjusting Analog Output Values

    Setting Up Serial Output To download data from the Analyzer directly to a computer, you must enable the Output option and set the Baud Rate for the serial (RS-232) port. SUEZ ©2018 73 of 162 DLM 74300-05 EN Rev. A...
  • Page 74: Interpreting Serial Output

    HyperTerminal. For information on setting up HyperTerminal to communicate with the Analyzer, see “Using HyperTerminal” on page 157 in the Appendix. Note: For long distance connections (up to 1,000 feet or 305 meters), SUEZ recommends use of either a RS-232 current loop converter or a RS-422 &...
  • Page 75: Issuing Serial Commands To The Analyzer

    5. Press the Interval button to print the average TOC value for the period of time specified. The options are as follows: Every Sample , Every Hour , Every 2 Hrs. , Every 4 Hrs. , or Every 8 Hrs. SUEZ ©2018 75 of 162...
  • Page 76: Activating Binary Input

    6. Press the Power Fail button to select On or Off . The On option activates the alarm trigger when the Analyzer loses power. 7. Press the Menu button when you are done. SUEZ ©2018 76 of 162 DLM 74300-05 EN Rev. A...
  • Page 77: Using The Ethernet Connection And Modbus

    Table 5: Modbus Register Table Register Type Description Function Address Discrete Inputs (read) Status 1000 New Data Error Standby Alarm 1 Alarm 2 Alarm 3 Alarm 4 Activations 1002 Feature 1 SUEZ ©2018 77 of 162 DLM 74300-05 EN Rev. A...
  • Page 78 TC Value High Word 3311 TC Units 3312 IC Value Low Word 3320 IC Value High Word 3321 IC Units 3322 Sample Temp. Value Low Word 3700 Sample Temp. Value High Word 3701 SUEZ ©2018 78 of 162 DLM 74300-05 EN Rev. A...
  • Page 79: Managing Maintenance Information

    The Maintenance tab also provides access to calibration and verification functions and other advanced features; consult Chapter 6: Calibration and System Protocols for details on these functions. SUEZ ©2018 79 of 162 DLM 74300-05 EN Rev. A...
  • Page 80: Displaying Consumables Status

    3. Press each date component button, enter the appropriate value, and press the Enter button. 4. Press each time component button, enter the appropriate value, and press the Enter button. SUEZ ©2018 80 of 162 DLM 74300-05 EN Rev. A...
  • Page 81: Reviewing Warnings And Errors

    The Analyzer automatically clears the corresponding icon W or E from the screen header. To delete the warning or error, press the Remove button. SUEZ ©2018 81 of 162 DLM 74300-05 EN Rev. A...
  • Page 82: Displaying System Information

    For more information about each warning or error, refer to the list of warning and error codes detailed in “Warnings and Error Messages” on page 136. Displaying System Information If you need to contact SUEZ about your Analyzer, you can find most essential information on a single screen. 1. Select the Maintenance tab.
  • Page 83: Exporting Or Printing The Warning And Error History

    To use the restore feature, be sure that the USB flash memory drive containing the previously exported data is attached to the Analyzer’s USB port, repeat Steps 1 through 3 above, and select Restore System . SUEZ ©2018 83 of 162...
  • Page 84: Printing The Constants Values

    2. Press the Advanced button. 3. Press the Advanced Setup button. 4. Press the Location button. 5. Use the keypad to specify a name for the Analyzer and press Enter . SUEZ ©2018 84 of 162 DLM 74300-05 EN Rev. A...
  • Page 85: Changing The Display Mode

    To change the Flow Sensor setting, follow these steps: 1. Select the Maintenance tab. 2. Press the Advanced button. 3. Press the Advanced Setup button. 4. Press the Flow Sensor button. SUEZ ©2018 85 of 162 DLM 74300-05 EN Rev. A...
  • Page 86: Specifying Auto Restart Settings

    5. Press the Frequency button and use the number keypad to enter how often you want to program the TOC Autozero function to work. The Frequency is measured in days; if you enter 7 the TOC SUEZ ©2018 86 of 162...
  • Page 87: Draining The Ios System

    2. Press the Activate Options button. 3. Press the OK button and wait for the Analyzer to detect the USB flash drive. 4. Press the Activate button to activate the new feature. SUEZ ©2018 87 of 162 DLM 74300-05 EN Rev. A...
  • Page 88: Basic Hardware Operation Features

    Slide the iOS System door open, wait 30 seconds for sample to drain from the iOS System, and then insert the vial containing the standard or sample. SUEZ ©2018 88 of 162 DLM 74300-05 EN Rev. A...
  • Page 89: The Sample Inlet Block

    System. The Sample Inlet Block The Sample Inlet Block (see Figure 20) is designed to for on-line sampling only. Sampling from a vial is not possible with this configuration. SUEZ ©2018 89 of 162 DLM 74300-05 EN Rev. A...
  • Page 90: Figure 20 The Sample Inlet Block

    Chapter 4: Basic Analyzer Operation Figure 20: The Sample Inlet Block Warning The Sample Inlet Block may be hot. If the sample water is hot, be careful to avoiding touching the Sample Inlet Block. SUEZ ©2018 90 of 162 DLM 74300-05 EN Rev. A...
  • Page 91: Chapter 5: Password Protection And Dataguard

    The DataGuard feature is available as an upgrade from SUEZ. DataGuard encompasses the same functionality as the Password Protection feature, but adds support for regulation 21 CFR Part 11 by maintaining an audit trail for all user operations and allows for the use of 100 unique User IDs.
  • Page 92: Using Password Protection

    Figure 21: Menu Map with Minimum User Level Using Password Protection Enabling Password protection ensures that only authorized personnel access the Analyzer. Once Password security is enabled, all users must enter the default User ID and Password. SUEZ ©2018 92 of 162 DLM 74300-05 EN Rev. A...
  • Page 93: Enabling Password Protection

    Login Timeout — Specifies the period of inactivity that passes before the use is automatically logged out. The default is 5 minutes. You can enter a value in the range of 1 to 30 minutes. SUEZ ©2018 93 of 162...
  • Page 94: Disabling Password Protection

    DataGuard offers support for regulation 21 CFR Part 11, providing multiple levels of user access to the Analyzer via up to 100 unique User IDs and maintaining an audit trail of all user operations. DataGuard must be purchased as an upgrade from SUEZ; once it is enabled, it cannot be disabled. Enabling DataGuard DataGuard is locked until it is activated by following the steps below.
  • Page 95: Adding User Ids

    1. Select the DataGuard tab. 2. Press the Add User button. 3. Specify a new User ID and press the Enter button. (If you enter an existing User ID, an error message will display.) SUEZ ©2018 95 of 162 DLM 74300-05 EN Rev. A...
  • Page 96: Editing User Information

    • Password — Press the Password button to change the Password for the selected User ID. 5. Press the Menu button to save changes and return to the DataGuard screen. SUEZ ©2018 96 of 162 DLM 74300-05 EN Rev. A...
  • Page 97: Configuring Login Settings

    5. Press the Password button. 6. Specify a new Password and press Enter . 7. Verify the new Password and press Enter . 8. Press the Menu button to return to the DataGuard screen. SUEZ ©2018 97 of 162 DLM 74300-05 EN Rev. A...
  • Page 98: Reactivating Inactivated User Accounts

    User IDs should be maintained. You may want to keep a list of User IDs to avoid duplication; Table 7 on page 101 provides a format for recording current and archived User IDs. SUEZ ©2018 98 of 162 DLM 74300-05 EN Rev. A...
  • Page 99: Viewing, Exporting, And Printing Audit Trails

    (either by printing or exporting). After printing or exporting, you will be prompted to erase the printed or exported audit trail information. Press OK to erase the data or Cancel to keep it. SUEZ ©2018 99 of 162 DLM 74300-05 EN Rev. A...
  • Page 100: Logging Out

    Back or Menu button until any of the main tabs ( Setup , Data , I/O , Maintenance , or DataGuard ) display, and then press the Logout button. SUEZ ©2018 100 of 162 DLM 74300-05 EN Rev. A...
  • Page 101: Table 7 Record Of User Ids

    Chapter 5: Password Protection and DataGuard Table 7: Record of User IDs Date Date User ID User’s Name Created Archived SUEZ ©2018 101 of 162 DLM 74300-05 EN Rev. A...
  • Page 102 Chapter 5: Password Protection and DataGuard SUEZ ©2018 102 of 162 DLM 74300-05 EN Rev. A...
  • Page 103: Chapter 6: Calibration And System Protocols

    The Analyzer can be calibrated using one of two methods: single-point calibration at 1.5 ppm; or multi-point calibration at 500 ppb, 1.0 ppm, and 1.5 ppm. SUEZ recommends single-point calibration for most applications. Multi-point calibration is available as an alternative procedure for customers.
  • Page 104: Required Calibration Supplies

    Note: For all calibration and verification procedures use only Sievers standards purchased from SUEZ. All standards are provided in 40-mL and 30-mL vials. Standards used in each of the system protocols are shown in Table 8 through Table 12. Standards concentrations are shown in ppm and ppb.
  • Page 105: Preparing For Calibration

    2. Press the Advanced button. 3. Press the USB I/O button. Make sure that the USB flash memory drive is attached to the USB port, and then press the Save System button. SUEZ ©2018 105 of 162 DLM 74300-05 EN Rev. A...
  • Page 106: Perform Annual Maintenance Tasks

    When you are ready to calibrate the Analyzer, follow these steps: 1. If the Analyzer is taking measurements, press the Stop Analysis button. 2. Press the Menu button, select the Maintenance tab, and press the Cal/Ver/Validate button. SUEZ ©2018 106 of 162 DLM 74300-05 EN Rev. A...
  • Page 107 8. Remove the vials from the iOS System and slide the iOS door closed. 9. You should now verify the calibration. Proceed to “Accuracy, Precision, and Calibration Verification” on page 110 for instructions. SUEZ ©2018 107 of 162 DLM 74300-05 EN Rev. A...
  • Page 108: Performing A Multi-Point Calibration

    Insert the third TOC Calibration Blank, and press Next . • When prompted, remove the TOC Calibration Blank from the iOS System, insert the TOC Calibration Standard (1.0 ppm KHP), and press Next . SUEZ ©2018 108 of 162 DLM 74300-05 EN Rev. A...
  • Page 109: Figure 23 The First Summary Screen In A Multi-Point Calibration

    9. Remove the vials from the iOS System and slide the iOS door closed. 10. You should now verify the calibration. Proceed to “Accuracy, Precision, and Calibration Verification” on page 110 for instructions. Figure 23: The First Summary Screen in a Multi-Point Calibration SUEZ ©2018 109 of 162 DLM 74300-05 EN Rev. A...
  • Page 110: Accuracy, Precision, And Calibration Verification

    The Sievers 500 RL On-Line TOC Analyzer automatically calculates accuracy, precision, and calibration verification values. Use this procedure to confirm that the Analyzer’s current calibration is accurate. SUEZ recommends performing a verification after replacement of consumables, such as the pump heads (with pump tubing), DI water cartridge, or UV lamp, and after calibration.
  • Page 111: System Suitability Verification

    Sucrose Standard (Rs), and press Next . • When prompted, remove the 500 ppb Sucrose Standard (Rs) from the iOS System, insert the 500 ppb Benzoquinone Standard (Rss), and press Next . SUEZ ©2018 111 of 162 DLM 74300-05 EN Rev. A...
  • Page 112: Figure 25 System Suitability Summary Screen

    6. Press the Exit button to continue. 7. Remove the vials from the iOS System and slide the iOS door closed. Figure 25: System Suitability Summary Screen SUEZ ©2018 112 of 162 DLM 74300-05 EN Rev. A...
  • Page 113: Linearity

    5. Press the Next button to display the second page of the linearity summary, which includes LOD and LOQ data. 6. Press the Exit button to continue. 7. Remove the vials from the iOS System and slide the iOS door closed. SUEZ ©2018 113 of 162 DLM 74300-05 EN Rev. A...
  • Page 114: Reviewing Calibration And Verification History

    Reviewing Calibration and Verification History If you need to review the results summary of past calibrations, follow these steps: 1. Press the Menu button and select the Maintenance tab. 2. Press the Calibrate button. SUEZ ©2018 114 of 162 DLM 74300-05 EN Rev. A...
  • Page 115 The Mode column lists TC Cal and IC Cal for calibration results, TC Ver and IC Ver for verification results, Sys Suit for system suitability results, and Linearity for Linearity results. These results also are included when you export the data history. SUEZ ©2018 115 of 162 DLM 74300-05 EN Rev. A...
  • Page 116: Single-Point Calibration Worksheet

    ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ Diff ________ ________ ________ ________ ________   Calibration Results: Passed Failed   Calibration Action: Applied Canceled SUEZ ©2018 116 of 162 DLM 74300-05 EN Rev. A...
  • Page 117: Multi-Point Calibration Worksheet

    ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ Diff ________ ________ ________ ________ ________   Calibration Results: Passed Failed   Calibration Action: Applied Canceled SUEZ ©2018 117 of 162 DLM 74300-05 EN Rev. A...
  • Page 118: Accuracy, Precision, And Verification Worksheet

    Fail   Accuracy ________ Pass Fail Pass criteria: TOC Precision: RSD of last three measurements of 500 ppb standard  3% • TOC Accuracy: % Difference  ±7% • SUEZ ©2018 118 of 162 DLM 74300-05 EN Rev. A...
  • Page 119: System Suitability Worksheet

     ----------------------------- - Response Efficiency = (Rs-Rw) Response Limit = Rs - Rw Acceptance criteria for USP System Suitability is response efficiency between 85% and 115%   Pass Fail SUEZ ©2018 119 of 162 DLM 74300-05 EN Rev. A...
  • Page 120: Linearity Worksheet

    ________ LOD/LOQ 750 ppb 500 ppb 250 ppb ________ ________ ________ ________ ________ ________ ________ ________ Intercept (Expected Value vs Standard Deviation)________ ________ ________    0.96 Pass Fail SUEZ ©2018 120 of 162 DLM 74300-05 EN Rev. A...
  • Page 121: Chapter 7: Maintenance

    Refer to Table 13, “Replacement Schedule for Consumables,” on page 122 for the recommended maintenance schedule. The in-line filter, UV lamp, DI water cartridge, and sample pump tubing (including pump heads) MUST be purchased from SUEZ. If you need additional assistance, contact Technical Support or your local service provider. Installation and training by a qualified service technician can also be provided.
  • Page 122: Checking And Refilling The Di Water Level

    1. If the Analyzer is taking measurements, press the Stop Analysis button. 2. Turn off power to the Analyzer with the main power switch. 3. Open the Analyzer front panel. 4. Locate the DI water cartridge (see Figure 28 on page 122). SUEZ ©2018 122 of 162 DLM 74300-05 Rev. A...
  • Page 123: Figure 29 Filling The Di Water Cartridge

    DI water cartridge to make sure it is still at the Full line. If necessary, repeat these steps to add more water. Figure 29: Filling the DI Water Cartridge SUEZ ©2018 123 of 162 DLM 74300-05 Rev. A...
  • Page 124: Replacing The Uv Lamp

    4. Locate the UV lamp enclosure at the right of the Analyzer (see Figure 28 on page 122). The UV enclosure has a brass nut and a gray wire with a black connector extending from the nut. SUEZ ©2018 124 of 162...
  • Page 125: Figure 30 Relative Positioning Of Components In The Uv Lamp Assembly

    Make sure the black heat tubing is flush against the brass-colored nut. 11. Tighten the brass-colored nut finger-tight. While tightening the nut, make sure the heat shrink tubing remains flush against the nut (Figure 30). Do NOT use a wrench. SUEZ ©2018 125 of 162 DLM 74300-05 Rev. A...
  • Page 126: Replacing The Sample Pump Tubing And Pump Heads

    The tubing cannot be replaced independently of the sample pump heads, and is provided as a kit with new pump heads. The sample pump tubing and pump heads must be purchased from SUEZ; use of tubing from other sources or failure to replace the tubing on the prescribed replacement schedule will affect Analyzer functionality.
  • Page 127: Replacing The Di Water Cartridge

    The ion exchange resin (resin bed) within the DI water cartridge depletes over time. With typical Analyzer use, the DI water cartridge should be replaced after every 24 months of operation. SUEZ ©2018 127 of 162 DLM 74300-05 Rev. A...
  • Page 128 DI water cartridge, and then gently squeeze the water bottle. Be sure to fill the reservoir until the water reaches the Full line. 18. Connect the fitting labeled “B” to the barb labeled “D” on top of the DI water cartridge. SUEZ ©2018 128 of 162 DLM 74300-05 Rev. A...
  • Page 129 35. Record the installation date for the DI water cartridge in the service log (see Table 15 on page 134). 36. Enter the installation date for the pump heads by following the procedures in “Setting the Installation or Date for New Consumables” on page 132. SUEZ ©2018 129 of 162 DLM 74300-05 Rev. A...
  • Page 130: Replacing The In-Line Filter Element

    System will decrease and can even stop. A simple way to determine if the filter element needs to be changed is to periodically measure the flow rate of water out the waste line from the iOS System and replace the filter element when the flow rate starts to decrease. SUEZ ©2018 130 of 162 DLM 74300-05 Rev. A...
  • Page 131 12. Reconnect the 1/4" Teflon tubing to the outlet of the filter and tighten the nut one-quarter turn past finger-tight. 13. Turn on the water supply at the sampling port. 14. Turn on the Analyzer using the main power switch. SUEZ ©2018 131 of 162 DLM 74300-05 Rev. A...
  • Page 132: Setting The Installation Or Date For New Consumables

    When Password protection or DataGuard is activated, not all users are able to modify settings on this screen. See “Using Password Protection” on page 92 or “Using DataGuard” on page 94 for more information. SUEZ ©2018 132 of 162 DLM 74300-05 Rev. A...
  • Page 133: Cleaning The Analyzer

    30 seconds by selecting the following option: 1. Press the Menu button and select the Maintenance tab. 2. Press the Advanced button, press the Advanced Setup button, and then press the Clean Screen button. SUEZ ©2018 133 of 162 DLM 74300-05 Rev. A...
  • Page 134: Table 15 Analyzer 3-Year Maintenance Worksheet

    Note: The DI water cartridge should be checked and filled as necessary as part of every maintenance. Analyzer Serial number ______________ Date placed into operation __________ Dates of operation encompassed by this worksheet: from __________ to __________ SUEZ ©2018 134 of 162 DLM 74300-05 Rev. A...
  • Page 135: Chapter 8: Troubleshooting

    To display and print the Warnings/Errors list, follow these steps: 1. Select the Maintenance tab. 2. Press the Warnings/Errors button. 3. Use the scroll buttons to browse the warnings and errors. SUEZ ©2018 135 of 162 DLM 74300-05 Rev. A...
  • Page 136: Warnings And Error Messages

    8 - User List Warning Message: “The user list is nearly full.” Explanation: DataGuard allows a maximum of 100 users. This warning displays when 90 user accounts have been created. SUEZ ©2018 136 of 162 DLM 74300-05 Rev. A...
  • Page 137 The Analyzer waits until the next scheduled time and attempts to perform the Autozero again. If this warning persists, there may be an issue with the water system. SUEZ ©2018 137 of 162 DLM 74300-05 Rev. A...
  • Page 138 IC or TC cell conductivity is out of range. Confirm that there are no air bubbles in any tubing inside the Analyzer, and make sure the water level in the DI water cartridge is sufficient. Also SUEZ ©2018 138 of 162...
  • Page 139 Explanation: This error displays when the audit trail is completely filled. To comply with 21 CFR Part 11, you must either Export ( DataGuard tab > Export AuditTrail ) or Print ( DataGuard tab > Print Audit Trail ) the audit trail before proceeding. SUEZ ©2018 139 of 162 DLM 74300-05 Rev. A...
  • Page 140 Stopped message and displays the 16 - Flow Started warning with the message “Sample stream flow restored.” The Analyzer also replaces the E (Error) icon at the top of the screen with a W (Warning) icon. SUEZ ©2018 140 of 162...
  • Page 141: Step 2: Visual Inspection

    Confirm that there are no leaks present at the sample inlet or the waste connections. • Confirm that all cables are securely attached to the Analyzer. • If you are sampling from vials, confirm that the needle in the iOS System is not obstructed. SUEZ ©2018 141 of 162 DLM 74300-05 Rev. A...
  • Page 142: Internal Inspection

    “Checking the Sample Pump” and “Backflushing the Analyzer” Note: Sample will not flow through the Analyzer if the sample pump is not operating or if the 60-µm stainless steel in-line filter is clogged. SUEZ ©2018 142 of 162 DLM 74300-05 Rev. A...
  • Page 143: Checking The Sample Pump

    30 - 300 mL/min. Make sure the Analyzer is powered on at all times, except when maintenance is being performed, to ensure that water flows through the Analyzer’s fluidics module. SUEZ ©2018 143 of 162 DLM 74300-05 Rev. A...
  • Page 144: Backflushing The Analyzer

    1. Disconnect the iOS tubing from the sample inlet. You may also want to place a container or paper towel below the inlet to collect water during the flush procedure. Figure 36: Sample Inlet SUEZ ©2018 144 of 162 DLM 74300-05 Rev. A...
  • Page 145: Figure 37 Measurement Module Tubing - Connected

    Figure 38: Measurement Module Tubing – Disconnected 4. Insert the plastic tip of the water bottle (provided in your accessory kit) into the tube as shown in Figure 39. DO NOT USE A SYRINGE. SUEZ ©2018 145 of 162 DLM 74300-05 Rev. A...
  • Page 146: Negative, High, Or Erratic Measurements

    1. If the Analyzer is taking measurements, press the Stop Analysis button. 2. Press the Menu button, select the Maintenance tab and press the Cal/Ver/Validate button. 3. Press Run TOC Autozero . 4. Wait for the TOC Autozero to complete before proceeding. SUEZ ©2018 146 of 162 DLM 74300-05 Rev. A...
  • Page 147: Toc Offset

    4. Press the fourth button in the left column ( Advanced Setup ). 5. Press the fifth button in the left column ( Select Language ). 6. Select your language by pressing the appropriate button. The program language changes immediately. SUEZ ©2018 147 of 162 DLM 74300-05 Rev. A...
  • Page 148: Problems With The Di Water Pump

    In some instances, after consulting with Technical Support or your local service provider, it will be necessary to return the Analyzer to the factory for repairs. Only return the Analyzer to SUEZ if Technical Support or your local service provider has issued you a Return Authorization (RA) number.
  • Page 149 Warning If the Analyzer is returned in anything other than a SUEZ shipping container, you will be charged for any damage that occurs during shipping. Before repackaging, water must be drained from the Analyzer.
  • Page 150: Figure 40 Removing The John Guest Fittings

    22. Close and latch the Analyzer front panel. You can now repackage the Analyzer and ship it to SUEZ. For international shipments, coordinate with a SUEZ representative to ensure quick passage through customs. Figure 40: Removing the John Guest Fittings SUEZ ©2018...
  • Page 151: Appendix A: Analyzer Dimension Diagrams

    Appendix A: Analyzer Dimension Diagrams Diagrams show Analyzer dimensions and required clearances as follows: inches [millimeters]. Figure 41: Analyzer Dimensions, with Standard iOS System SUEZ ©2018 151 of 162 DLM 74300-05 EN Rev. A...
  • Page 152: Figure 42 Right-Side Analyzer Dimensions, With Standard Ios System

    Appendix A Figure 42: Right-Side Analyzer Dimensions, with Standard iOS System SUEZ ©2018 152 of 162 DLM 74300-05 EN Rev. A...
  • Page 153: Figure 43 Analyzer Dimensions, With Sample Inlet Block

    Appendix A Figure 43: Analyzer Dimensions, with Sample Inlet Block SUEZ ©2018 153 of 162 DLM 74300-05 EN Rev. A...
  • Page 154: Figure 44 Right-Side Analyzer Dimensions, With Sample Inlet Block

    Appendix A Figure 44: Right-Side Analyzer Dimensions, with Sample Inlet Block SUEZ ©2018 154 of 162 DLM 74300-05 EN Rev. A...
  • Page 155: Figure 45 Left-Side Analyzer Dimensions

    Appendix A Figure 45: Left-Side Analyzer Dimensions SUEZ ©2018 155 of 162 DLM 74300-05 EN Rev. A...
  • Page 156 Appendix A SUEZ ©2018 156 of 162 DLM 74300-05 EN Rev. A...
  • Page 157: Appendix B: Connecting To A Pc

    “Determining Your Communications Port Number” on page 158. In the Com Settings window, specify the following settings: SUEZ ©2018 157 of 162 DLM 74300-05 EN Rev. A...
  • Page 158: Determining Your Communications Port Number

    4. Scroll down to Ports (Com and LPT) and expand the menu by clicking the plus sign (+). 5. The communication ports in use should be listed, along with device names. SUEZ ©2018 158 of 162 DLM 74300-05 EN Rev. A...
  • Page 159: Index

    71 Configuration system settings 83 setting the Analyzer mode 65 warnings and errors 81 Constants exporting 83 printing 84 Consumables DI water cartridge 122, 127 displaying status 64, 80 SUEZ ©2018 159 of 162 DLM 74300-05 EN Rev. A...
  • Page 160 System 88 DI water cartridge 122 Sample Pump Heads sample pump tubing and pump heads 126 replacing 126 setting installation dates 132 Screens UV lamp 124 Main 64 worksheet 134 SUEZ ©2018 160 of 162 DLM 74300-05 EN Rev. A...
  • Page 161 Analyzer for shipping 148 visual inspection 141 warnings and errors 136 USB Flash Drive installing 50 User IDs adding 95 archiving 98 changing passwords 97 editing information 96 SUEZ ©2018 161 of 162 DLM 74300-05 EN Rev. A...
  • Page 162 Notes SUEZ ©2018 162 of 162 DLM 74300-05 EN Rev. A...