INSTALLATION GUIDE
The FLOMEC QS200 provides a robust ultrasonic
flowmeter for irrigation applications. The QS200 uses
Percentage of Reading (RD) Accuracy which is superior to
an impeller flow sensor/flowmeter that uses Percentage
of Full Scale (FS) Accuracy
ACCURACY WHITE PAPER for more information on RD and
FS information @
http://bit.ly/flomec-accuracy-wp). Also,
the QS200 does not have any moving wetted parts, so
there is no need to replace damaged parts.
IMPORTANT:
Read the General Safety Instructions for this flowmeter (see last page) before beginning any installation procedures.
GENERAL INSTALLATION PROCEDURE
The following lengths of straight pipe should be connected to the QS200 FLOMEC Tee (or NON-FLOMEC Tee):
10 times the pipe's inner diameter upstream of the meter
-
5 times the pipe's inner diameter downstream of the meter
-
The pipe needs to be full of irrigation water with no air inside. Built-up sediment on the QS200's transducers or large
debris can greatly hinder the meter's performance. Keep the following in mind when installing:
Install QS200 at an angle as shown in Figure 1a.
-
1 inch pipe installations will need a larger angle
-
of 45 degrees.
Flow direction arrows on the QS200 insert must
-
match flow direction of water through the pipe
(see Figure 1a).
FLOMEC Tee is bidirectional for flow (the QS200
-
insert IS NOT bidirectional - the flow direction
matters).
Flow direction of the QS200 should be horizontal
-
or upward. DO NOT install flowmeter in a
downward flow direction, to avoid partially filled
pipes (see Figure 1b).
FLOMEC
(download the FLOWMETER
QS200
®
This meter uses a two-wire system that is connected to an
irrigation controller's flow sensor input (or decoder's flow
sensor port) and meets the cable distance recommended
by Irrigation controller/decoder manufacturers. The
two-wire system communicates via a square wave pulse
that is converted to flow rate at the controller (identical
to an impeller flow sensor communicating to irrigation
controller/decoder). The QS200's power is supplied
through these same two wires by the controller's/
decoder's flow sensor input/port
(voltage range of 7.5 V
When installing below ground, install meter in a
-
valve box with a thick layer of gravel below the
meter.
DO NOT direct bury QS200 flowmeter.
-
If using a Non-FLOMEC Tee, verify the QS200
-
insert physically fits into the Non-FLOMEC Tee.
DO NOT pull on the QS200 wires under any
-
circumstances.
Confirm that the Controller/Decoder flow sensor
-
input (or port) will supply the QS200 two-wire
meter with the appropriate power
(7.5 V
to 36 V
DC
-1-
to 36 V
).
DC
DC
).
DC
08/20 IND-1126 Rev A QS200 Installation Guide

Advertisement

loading

Summary of Contents for Flomec QS200

  • Page 1 5 times the pipe’s inner diameter downstream of the meter The pipe needs to be full of irrigation water with no air inside. Built-up sediment on the QS200’s transducers or large debris can greatly hinder the meter’s performance. Keep the following in mind when installing: Install QS200 at an angle as shown in Figure 1a.
  • Page 2: Flow Direction

    NOTE: It is recommended to install Tee at an angle. The water lines are not supposed to have air in them, however, sometimes very minute pockets of air can get trapped in the QS200. Installing at an angle helps reduce the consequences PAGE 8 of this issue.
  • Page 3 (see NOTE: When replacing another manufacturer’s insert Figure 5). with the QS200 Retrofit Insert, the old insert could have a sediment or mineral buildup and may need to be leveraged out of the tee.
  • Page 4 5. A levering tool by slipping the U-formed end QS200 meter or other brands of tee type insertion meters. around the insert between the insert lip and the PAGE 16 & 17 PAGE 9 &...
  • Page 5 INSTALLING QS200 METER INTO AN EXISTING TEE INSTALLATION (continued) Remove Old Insert From Tee with FLOMEC Multi-Tool PRY UP - To break a stuck insert loose DRIVING WEDGE - To lift insert more LEVERAGE - To Lift insert up and out...
  • Page 6 QS200 WIRING INSTALLATION (Tee and Meter should be installed before splice connections) Make sure wiring is only done on low DC voltage (7.5 V to 36 V ). Make sure to follow all of the Controller/Decoder manufacturer’s directions for electrical safety. Also, refer to the General Safety Instruction section provided in this document.
  • Page 7: Selection Chart

    SPLICE KIT FLOW INSERT Figure 13 Calibration Set K-factor and Offset per QS200 FLOW INSERT SELECTION CHART below (reference QS200 Data Sheet - FLOW INSERT SELECTION CHART - Download DATA SHEET HERE. These calibrations will be configured in the irrigation controller.
  • Page 8: Information Resources

    NOTE: Be sure O-rings and seals are kept in good repair. Do not allow this meter to be used with steam. Great Plains Industries, Inc. Australia +61 2 9540 4433 FLOMECmeters.com © 2020 Great Plains Industries, Inc. All Rights Reserved. 08/20 IND-1126 Rev A QS200 Installation Guide...