OfiTE 174-00 Instruction Manual

OfiTE 174-00 Instruction Manual

Portable roller oven

Advertisement

Portable Roller Oven
174-00 (115 Volt)
174-00-1 (230 Volt)
Instruction Manual
Updated 12/7/2020
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com
©
Ver. 4
Copyright OFITE 2015

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 174-00 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for OfiTE 174-00

  • Page 1 Portable Roller Oven 174-00 (115 Volt) 174-00-1 (230 Volt) Instruction Manual Updated 12/7/2020 Ver. 4 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com Copyright OFITE 2015 ©...
  • Page 2: Table Of Contents

    Table of Introduction..................2 Description ..................2 Contents Components ...................3 Specifications .................4 Safety ....................4 Operation..................5 Timer Operation ................7 Maintenance ..................9 Warranty and Return Policy ............11 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 3: Introduction

    2. Rolling Mode Only: Ball Mill Roller To make homogenous mixtures of liquids To make homogenous mixtures of powders To agitate chemicals into solutions To de-aerate liquids OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 4: Components

    Bearing for Roller Shaft (Glass Impregnated Teflon); Qty: 8 #172-13 Fuse Light Holder; Qty: 2 #172-20 150-Watt Heater; Qty: 2 #172-24 Solid State Relay; 240 Volt - 25 Amp; Qty: 2 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 5: Specifications

    Do not overfill the cell. Refer to the instruction manual for Aging Cells #175-25, #175-30, #175-50 to determine the appropriate fluid volume for the test conditions. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 6: Operation

    N (#175-54) o-rings are available. For tests above 300°F (148.9°C), use Teflon only. If for some reason the temperature controller fails, the safety thermostat will cut power to the heater at 500°F (260°C). OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 7 The temperature controller will indicate the oven temperature as long as the heat switch is in the “On” position. For faster cooling, open the oven door. Shelf OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 8: Timer Operation

    (in minutes) that you want the timer program to run. 6. Press the “Page” button three times to return to the temperature display screen. The timer program will now begin to run. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 9 The display will read “AUTO”. Simultaneously press and hold the “Page” and “Scroll” buttons for four seconds. This will return the temperature controller back to the normal user mode. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 10: Maintenance

    For 115V, use part no. 165-47. For 230V, use part no. 164-30. Plug Speed Control Recepticle Knob Circuit Board Voltage Speed Control Conversion Knob Switch Voltage Conversion Switch OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 11 Refer to the photo on page 5. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 12: Warranty And Return Policy

    In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied,...

This manual is also suitable for: