The aging process due to the length of time required for com- pletion of the test, is seldom monitored, so a heater failure may be serious. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Everyone’s consumable requirements will be different, and replace- ment quantities needed will depend upon the number of test performed on a daily and/or weekly basis. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Periodically check the inside of Aging Cells for stress or pit corrosion. Oper- ating cells in high or low pH concentrations or with concentrated saline solu- tions (Chlorides) will promote corrosion. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
8. Press the “h” and “min” buttons to set the correct time for the “OFF” cycle to end. The “DOWN” arrow will blink. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 9
Then press and hold the “+1h / CLEAR” button for 3 seconds. Temperature Controller 1. Upper Temperature is the current temperature. 2. Lower Temperature is the temperature set point. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Teflon (#175-46) and Buna N (#175-54) o-rings are available. For tests above 300°F (148.9°C), use Viton o-rings. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 12
“Heat” switch is on. For faster cooling, open the door. 7. After the test is complete, turn the “Heat” and “Motor” switches off. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
5. Press the “h” and “min” buttons to set the correct local time. 6. Press “WRITE” when finished. 7. Set the “P1, P2, Run” button to P1 to show the time. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 14
10. Set the switch (“P1, P2, Run”) to the right of the display to “RUN”. The bar (-) will display the day(s) of operation. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 15
5. Only one of the two channels are utilized. 6. Season switching, cyclic operation, and non-operating day (Holiday) settings are non-functional as the timer may be programmed daily for such functions. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
(switch them both to the right for 115V and to the left for 230V). d. Re-install the top panel. Voltage Speed Control Conversion Knob Switch Voltage Conversion Switch OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 17
Over time it will become rigid and hard and will not insulate as well. If this happens, replace the insulation (#172-10). OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 18
The cover on the back of the oven will need to be removed. Always re- place it after the necessary repairs have been completed. Roller Oven – Rear With Back Panel in Place Chain Drive and Sprockets OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
11. Press the up arrow key once. The display will read “EIit / yes”. This will reboot the temperature controller. After the reboot, the display will show the new units. Page Scroll OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties...
Need help?
Do you have a question about the 173-00-C and is the answer not in the manual?