  • Page 1 ULTRASONIC SCALER FMS-2530 Flight Dental Systems 21 Kenview Blvd., Unit 9, Ontario, Canada, L6T5G7 www.flightdentalsystems.com Toll Free: 1-866-799-0517...
  • Page 2: Table Of Contents

    FMS-2530 Tabletop Ultrasonic Scaler V.1 Table of Contents Section Section Title Page Number Description of Contents Precautions………………………………………………………………………………………… 3 1.1 Precautions for All Systems Introduction……………………………………………………………………………………….. 4 Glossary of Symbols…………………………………………………………………………….. 4 Intended Use………………………………………………………………………………………. 5 Specification………………………………………….……………………………………………. 5 Contraindications and Warnings……………………………………………………………. 5 6.1 Contraindications 6.2 Warnings...
  • Page 3: Precautions

    FMS-2530 Tabletop Ultrasonic Scaler V.1 Section 1: Safety Precautions Prior to the installation and start-up of the ultrasonic scaler, please read the instructions provided here. 1.1 Precautions for All Systems Do not place the ultrasonic scaler on or next to a radiator or any other heat source. Excessive heat may damage the ultrasonic scaler’s electronics.
  • Page 4: Introduction

    DEVICE TO SALE BY OR ON THE ORDER OF A 2.2 Supplies and Replacement Parts DENTAL PROFESSIONAL. Contact your local Flight Dental Systems dealer to order supplies or replacement parts. There are no user serviceable parts inside this device. Please...
  • Page 5: Intended Use

    FMS-2530 Tabletop Ultrasonic Scaler V.1 Icon to identify electric and electronic devices. The unit must be collected and disposed of according to local laws. Section 4: Intended Use The Ultrasonic Scaler Procedures ‧All general supra and subgingival scaling applications. ‧Periodontal debridement for all types of periodontal diseases.
  • Page 6: Infection Control

    Please carefully follow procedural guide lines in this manual regarding The Flight Dental Systems FMS-2530 Ultrasonic the use of your ultrasonic scaler. Scaler is designed for use in prophylaxis and periodontal treatments, and other areas of ‧ Always flush your Ultrasonic Scaler at least 20 operative dentistry.
  • Page 7 FMS-2530 Tabletop Ultrasonic Scaler V.1 infection control procedures.
  • Page 8: Installation Instructions

    40 psi (276 kPa) maximum. If your dental water system's supply line pressure is ‧ Place the FMS-2530 unit on a level surface. above 40 psi, install a water pressure regulator ‧ Be sure unit is stable and resting on four feet.
  • Page 9: Ultrasonic Scaler Description

    FMS-2530 Tabletop Ultrasonic Scaler V.1 Section 9: Ultrasonic Scaler Description 9.1 Rear Panel Power Switch Solenoid Water Inlet Foot Pedal Connector Power Inlet 9.2 System Control Turbo Indicator Light Sterilizable Sleeve Power Adjustment Lit when Foot Switch is fully of Handpiece...
  • Page 10: Accessories

    FMS-2530 Tabletop Ultrasonic Scaler V.1 9.3 Flight Handpiece Cable Set / Sterilizable Sleeve of Handpiece Assembly ® The ultrasonic scaler is a multi-frequency unit compatible with all Cavitron inserts. The system will automatically detect the insert without a need to switch settings.
  • Page 11: Techniques For Use

    FMS-2530 Tabletop Ultrasonic Scaler V.1 5. Sterilizable sleeve for the handpiece Section 11: Techniques For Use 11.1 Patient Positioning ‧ For optimal access to both the upper and lower ‧ It may be necessary to adjust water flow larger arches, the backrest of the chair should be under "Turbo"...
  • Page 12: System Maintenance And Care

    FMS-2530 Tabletop Ultrasonic Scaler V.1 Section 12: System Maintenance and Care Daily Maintenance 3. Place a sterilized handpiece sleeve on the It is recommended that you perform the following handpiece cable. Set power to minimum. Hold the maintenance procedures to help minimize bio-film...
  • Page 13: Troubleshooting

    ‧ Dispose of the Ultrasonic Scaler in accordance with local and national laws. Section 15: Disclaimer FLIGHT DENTAL SYSTEMS considers itself responsible for the effects on safety, reliability and performance of this product only if: ‧ Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by FLIGHT DENTAL SYSTEMS.
  • Page 14: Warranty

    Section 16: Warranty 16.1 Malfunction Flight Dental Systems hereby warrants that for a period of one year from the delivery date, this device shall be free from defects in material and workmanship. In case the machine is found defective under normal use, Flight Dental Systems will offer parts under warranty for replacement.
  • Page 15 IEC 61000-3-3 Guidance and manufacturer’s declaration-electromagnetic immunity The FMS-2530 is intended for use in the electromagnetic environment specified below. The customer or the user of the FMS-2530 should assure that it is used in such an environment. Immunity test IEC 60601...
  • Page 16 NOTE: UT is the AC mains voltage prior to application of the test level. Guidance and manufacturer’s declaration-electromagnetic immunity The FMS-2530 is intended for use in the electromagnetic environment specified below. The customer or the user of the FMS-2530 should assure that is used in such and environment. Immunity test IEC 60601 test...
  • Page 17 If the measured field strength in the location in which the FMS-2530 should be observed to verify normal operation. If abnormal performance is observed, additional measures my be necessary, such as re-orienting or relocating the FMS-2530.