Flight DEFENDME Aerosol Evacuator R.1.2 2.0 Symbols If the instructions are not followed properly, operation of this device WARNING may lead to hazards for the user/patient and failure of the equipment NOTE Additional information: Explanation of operation and performance. Serial number...
3.2 Safety Instructions 1. Please read the manual before use. 2. The DEFENDME system must be placed upright while being used. Laying the unit on its side or upside down is prohibited, doing so will cause damage to the machine and shorten its service life.
Flight DEFENDME Aerosol Evacuator R.1.2 4.0 Installation 4.1 Removing the Protective Foam Before using the machine for the first time, please remove the foam at the bottom of the main filter assembly otherwise the machine will not work properly. Please refer to the following steps to remove the foam.
Flight DEFENDME Aerosol Evacuator R.1.2 4.2 Installing DefendMe Aerosol Evacuator Step Diagram Description Installation of the Suction Nozzle: Fasten the suction nozzle to the directional duct by screwing the nut onto the duct Installation of the Directional Duct: Plug the direction duct to the top of the cover.
Flight DEFENDME Aerosol Evacuator R.1.2 Connect the Speed Controller: Plug the connector of the speed controller into the main housing as shown in the picture. 5.0 User Interface Turn power On Press Power Switch to “ “ to turn on.
Flight DEFENDME Aerosol Evacuator R.1.2 Turn on the Controller: Press speed Controller power switch to “ON” Speed Setting: Rotate Speed Setting knob clockwise to increase air volume from 1 to 10 and counter clockwise to reduce the speed. Turn of the Controller: 5.
Flight DEFENDME Aerosol Evacuator R.1.2 display “OFF”. To disconnect the power supply please press the power switch on the main enclosure box. Condensation may appear on the suction nozzle if the temperature of the suction nozzle is much lower than the atmospheric temperature. If this occurs,...
Flight DEFENDME Aerosol Evacuator R.1.2 components. Please contact your authorized dealer for repair or replacement of any components. 7.1.2 General Recommendations • The user is responsible for disinfecting the product before use. • For you own safety, please wear the appropriate personal protective equipment (PPE) (gloves, safety glasses, gowns etc) when using and cleaning this device.
Flight DEFENDME Aerosol Evacuator R.1.2 Secondary Filter Assembly 3-6 months Displays 050 or a multiple of Eg.050/100/150 Main Filter Assembly 12 Months Display 100 or a multiple of it Eg. 100/200/300 Replace relevant filter according to the duration (months) or if the flow value has been reached.
Technical Data Manufacturer HR Dental Products Inc. o/a Flight Dental Systems Model DefendMe Aerosol Evacuator Model # AE-1000 Dimensions 16 x 9 x 18”...
10.0 EMC Tables Guidance and manufacturer’s declaration – electromagnetic emissions The DEFENDME is intended for use in the electromagnetic environment specified below. The customer or the user of the DEFENDME should assure that it is used in such an environment.
Page 15
Flight DEFENDME Aerosol Evacuator R.1.2 Surge Line to line: Line to line: Mains power quality should be that of IEC 61000-4-5 ±0.5kV, ±1kV ±0.5kV, ±1kV a typical commercial or hospital environment. Line to earth: Line to earth: ±0.5kV, ±1kV, ±0.5kV, ±1kV, ±2kV...
Page 16
Note: UT: rated voltage(s); E.g. 25/30 cycles means 25 cycles at 50Hz or 30 cycles at 60Hz Guidance and manufacturer’s declaration – electromagnetic immunity The DEFENDME is intended for use in the electromagnetic environment specified below. The customer or the user of the DEFENDME should assure that it is used in such an environment.
Page 17
When they are used in close proximity to medical equipment and/or systems, the medical equipment and/or systems’ basic safety and essential performance may be affected. The DEFENDME has been tested with the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2014.