CO
/ humidity and hand-held temperature
2
measuring device CO
Technical data
Power supply:
4 x AA batteries (included)
Measuring range:
Temperature:
0 – 50 °C (accuracy ± 0.5 °C)
Humidity:
0 – 95 % RH (version -RH only)
(accuracy ± 3 % at 25 °C
and 10 – 90 % RH, otherwise ± 5 %)
CO
:
0 – 5,000 ppm (version -RH only)
2
(accuracy ± 30 ppm and ± 5 % of reading)
0 – 2,000 ppm (standard version)
(accuracy ± 75 ppm and ± 5 % of reading)
Output:
RS 232 data interface
Perm. Ambient
conditions:
0 – 50 °C, 0 – 95 % RH
(non-condensing)
Housing:
plastic ABS / PC
Dimensions:
70 x 210 x 58 mm (L x W x D)
Weight:
about 180 g (without batteries)
Introduction
Thank you for purchasing this portable CO
measures CO
level, air temp., dew point, wet bulb temp. and
2
humidity (DP, WB, RH are for model version RH) and is an ideal
instrument for indoor air quality (IAQ) diagnosis.
Poor indoor air quality is considered unhealthy because it
causes tiredness, loss of ability to concentrate, and even illness
(ex. Sick Building Syndrome). IAQ monitoring and survey, espe-
cially on CO
level and air ventilation become widely applied in
2
public areas such as offices, classrooms, factories, hospitals and
hotels. It is also suggested in regulations of industrial hygiene in
some countries.
The portable CO
meter uses NDIR (non-dispersive infrared)
2
technology to ensure the reliability and long term stability. It's
useful in verifying HVAC system performance and air ventilation
control.
-TEMP(-RH)-HMG
2
Features:
•
Triple displays of CO
humidity (version -RH)
•
NDIR sensor for CO
•
Statistics of weighted averages
•
•
•
Backlight for working in dark area
•
Audible CO
•
Battery and adaptor power supply
•
Easy manual calibration on CO
Material supplied
This package contains:
Meter, 4pcs AA batteries, Operation manual, Hard carrying case
Power supply
The meter is powered by either 4 AA batteries or a DC adaptor
(9V/1A output).
Install the batteries into the battery compartment on the rear
and make sure they are in correct polarity and good contact.
When an adaptor is used, it will cut off the power supply from
meter. The meter
batteries. The adaptor can't be used as a battery charger.
2
When battery voltage gets low,
the LCD (Fig.1). And beeper sounds. The CO
under low voltage, so it beeps to indicate failed CO
surement (press any key but
readings won't be displayed. Please replace with fresh batteries
or connect with an adaptor.
level, temp. and
2
detection
2
TWA (8 hours weighted average)
STEL (15 minutes weighted average)
warning alarm
2
version -RH
Sensors PG1 | Data sheet No. 14108 | Version 11-2018 | 1 | 4
2
and "Lob" will appear on
sensor can't work
2
mea-
2
to stop the beeps) and the