Co2 /humidity and hand-held temperature measuring device

Advertisement

Quick Links

CO
/ humidity and hand-held temperature
2
measuring device CO
Technical data
Power supply:
4 x AA batteries (included)
Measuring range:
Temperature:
0 – 50 °C (accuracy ± 0.5 °C)
Humidity:
0 – 95 % RH (version -RH only)
(accuracy ± 3 % at 25 °C
and 10 – 90 % RH, otherwise ± 5 %)
CO
:
0 – 5,000 ppm (version -RH only)
2
(accuracy ± 30 ppm and ± 5 % of reading)
0 – 2,000 ppm (standard version)
(accuracy ± 75 ppm and ± 5 % of reading)
Output:
RS 232 data interface
Perm. Ambient
conditions:
0 – 50 °C, 0 – 95 % RH
(non-condensing)
Housing:
plastic ABS / PC
Dimensions:
70 x 210 x 58 mm (L x W x D)
Weight:
about 180 g (without batteries)
Introduction
Thank you for purchasing this portable CO
measures CO
level, air temp., dew point, wet bulb temp. and
2
humidity (DP, WB, RH are for model version RH) and is an ideal
instrument for indoor air quality (IAQ) diagnosis.
Poor indoor air quality is considered unhealthy because it
causes tiredness, loss of ability to concentrate, and even illness
(ex. Sick Building Syndrome). IAQ monitoring and survey, espe-
cially on CO
level and air ventilation become widely applied in
2
public areas such as offices, classrooms, factories, hospitals and
hotels. It is also suggested in regulations of industrial hygiene in
some countries.
The portable CO
meter uses NDIR (non-dispersive infrared)
2
technology to ensure the reliability and long term stability. It's
useful in verifying HVAC system performance and air ventilation
control.
-TEMP(-RH)-HMG
2
Features:
Triple displays of CO
humidity (version -RH)
NDIR sensor for CO
Statistics of weighted averages
Backlight for working in dark area
Audible CO
Battery and adaptor power supply
Easy manual calibration on CO
Material supplied
This package contains:
Meter, 4pcs AA batteries, Operation manual, Hard carrying case
Power supply
The meter is powered by either 4 AA batteries or a DC adaptor
(9V/1A output).
Install the batteries into the battery compartment on the rear
and make sure they are in correct polarity and good contact.
When an adaptor is used, it will cut off the power supply from
meter. The meter
batteries. The adaptor can't be used as a battery charger.
2
When battery voltage gets low,
the LCD (Fig.1). And beeper sounds. The CO
under low voltage, so it beeps to indicate failed CO
surement (press any key but
readings won't be displayed. Please replace with fresh batteries
or connect with an adaptor.
level, temp. and
2
detection
2
TWA (8 hours weighted average)
STEL (15 minutes weighted average)
warning alarm
2
version -RH
Sensors PG1 | Data sheet No. 14108 | Version 11-2018 | 1 | 4
2
and "Lob" will appear on
sensor can't work
2
mea-
2
to stop the beeps) and the

Advertisement

loading

Summary of Contents for Oppermann Regelgeräte CO2-TEMP-HMG

  • Page 1 / humidity and hand-held temperature measuring device CO -TEMP(-RH)-HMG Features: • Triple displays of CO level, temp. and humidity (version -RH) • NDIR sensor for CO detection • Statistics of weighted averages • TWA (8 hours weighted average) • STEL (15 minutes weighted average) Technical data •...
  • Page 2 / humidity and hand-held temperature measuring device CO -TEMP(-RH)-HMG LCD display Taking measurement The meter starts measurement when power on and update readings every second. In the condition of operating environ- ment change (ex. from high to low temp.), it takes 30 sec to respond for CO sensor and 30 minutes for RH.
  • Page 3 / humidity and hand-held temperature measuring device CO -TEMP(-RH)-HMG Note: Press to increase the value or to decrease. Each press 1. If the meter is turned on for shorter than 15 minutes, the STEL tunes 100 ppm and the alarm range is from 100 to 9900 ppm. value will be the weighted average of readings taken since When the preferred alarm value is set, press to save the...
  • Page 4 / humidity and hand-held temperature measuring device CO -TEMP(-RH)-HMG Troubleshooting Can’t power on Press for more than 0.3 seconds and try again. Check whether batteries are in good contact and correct polarity, or the adaptor is well plugged. Fixed readings Check whether data hold function was activated.

This manual is also suitable for:

Co2-temp-rh-hmg