Operator’s Installation and Instruction Manual Detcon Model 1000 Series H S Analyzer consists of two major assemblies: 1. The Model 1000 Series gas sample-conditioning assembly. 2. The Model DM-624 H S gas sensor (using Alcohol Free - Electrochemical Sensor or Hi-Range Electrochemical Sensor) DETCON, Inc.
Page 2
Model 1000 H2S Echem This page left intentionally blank Model 1000 H2S Echem Instruction Manual...
The analyzer requires a constant, liquid-free, 10±2psig sample pressure, which is provided by the customer or may be supplied by Detcon as a separate gas sample handling system. The on-board gas sample conditioning system includes a stainless steel pressure gauge (0-30psig), and a Genie membrane filter to provide the analyzer with absolute condensate liquid protection.
Model 1000 H2S Echem Detcon MicroSafe™ Model DM-624, toxic sensors are non-intrusive “Smart” sensors designed to detect and monitor for H S gas in the ppm range. A primary feature of the sensor is its method of automatic calibration, which guides the user through each step via instructions displayed on the backlit LCD. The sensor features field adjustable, fully programmable alarms and provides relays for two alarms plus fault as standard.
Model 1000 H2S Echem Figure 3 Functional Block Diagram Figure 4 DM-624 Control Circuit 1.4 Base Connector Board The base connector board is mounted in the enclosure and includes: the mating connector for the control circuit, reverse input and secondary transient suppression, input filter, alarm relays, lugless terminals for all field wiring, and a terminal strip for storing unused programming jumper tabs.
Model 1000 H2S Echem 1.5 Interference Data DM-624 series electrochemical H S sensors are subject to interference from other gases. This interaction is shown in the following table as the relation between the amount of the interfering gas applied to the sensor, and the corresponding reading that will occur (in ppm).
Model 1000 H2S Echem 2. Specifications Sensor Type Electrochemical cell Measurement Range 0-10ppm H S, up to 0-5 % H Accuracy/Repeatability ±10% of reading or ±0.5ppm, whichever is greater Response/Clearing Time T80 < 1 minute Operating Temperature -40°F to +122°F: -40°C to +50°C Outputs Linear 4-20mA DC;...
Model 1000 H2S Echem 3. Installation 3.1 Mounting Securely mount the Model 1000 analyzer panel or NEMA 4 enclosure (optional) in accordance with Figure 7. Figure 7 Unit Dimensions 3.2 Gas Connections and Sampling System Notes 1. Install a length of tubing from the desired sample point to the sample input port (as shown in Figure 8). Sample draw tubing should be 316 stainless steel of ¼"...
Set a flow of 100-200cc/min. and vent to a safe area using ¼” O.D. tubing. 3. Connect to the Over-pressure Relief valve and vent to a safe area. The pressure relief valve is set at Detcon to open at 15-20psig. For convenience sake, the sample bypass and over-pressure relief can be vented together.
Model 1000 H2S Echem Air Inlet Port SAMPLE FLOW FLOW Optional Pressure Guage Acetic Acid Bubbler Sample Inlet Port Bottle Sample Calibrate S Vent Port Sample Bypass and Liquid NOTE: Rejection Exhaust Vent. These can be combined into a Over Pressure Relief Vent single Vent.
Model 1000 H2S Echem 3.4 Relays and RS-485 Setup Program the alarms via the gold plated jumper tab positions located on the CPU board (Figure 11). Alarm 1 and Alarm 2 have three jumper programmable functions: latching/non-latching relays, normally energized/normally de-energized relays, and ascending/descending alarm set points. The fault alarm has two jumper programmable functions: latching/non-latching relay, and normally energized/normally de-energized relay.
Model 1000 H2S Echem 4. Start Up Upon completion of all tubing connections and field wiring the Model 1000 Series Process Analyzer is ready for startup. Note that after power is applied, varying readings may occur during sensor warm-up. Allow at least 1 hour for stabilization (24 hours is best).
Model 1000 H2S Echem 5. Operating Software 5.1 Programming Magnet Operating Instructions Operator interface to MicroSafe™ gas detection products is via magnetic switches located behind the transmitter faceplate. DO NOT remove the glass lens cover to calibrate or change programming parameters. Two switches labeled “PGM 1”...
Model 1000 H2S Echem 5.2.2 Calibration Mode Calibration mode allows for sensor zero and span adjustments. It uses an “AUTO SPAN” sequence with built in diagnostics to ensure adequate sensor response and signal stability “1-ZERO 2-SPAN”. Zero Adjustment Zero is set with no H S target gases present.
Page 19
Model 1000 H2S Echem c) Exit back to normal operations by holding the programming magnet over “PGM 2” for 3 seconds, or, the sensor will automatically return to normal operation in 30 seconds. Alarm 1 Level Adjustment The alarm 1 level is adjustable over the range 10% to 90% of range. For hydrogen sulfide gas sensors, the level is factory set at 20% F.S.
Figure 14 DM-624 Software Flow Chart 6. Calibration and Maintenance The Model 1000 Series H S Process Analyzer is calibrated prior to shipment. Only minimal adjustment should be required at time of commissioning. However, it is highly recommended that a zero and span calibration be performed twice in the first week of operation to assure optimum system performance.
Material Requirements Detcon Part Number 327-000000-000 MicroSafe™ Programming Magnet. Detcon Part Number 942-001123-000 Zero Air cal gas. (Ambient air can be used, if it can be verified that no target gas is present). Span gas cylinder containing H S gas in background N2 (nitrogen) as per the following range table.
Page 22
Model 1000 H2S Echem 1) Verify that the current calibration gas level is equal to the calibration span gas concentration. (Refer to Section 5.2.3 Program Mode, “View Sensor Status”. The item appears as “CAL LEVEL @ ### PPM”.) 2) If the calibration gas level setting is equal to the calibration span gas concentration, proceed to item “3)”. If the calibration gas level setting is not equal to the calibration span gas concentration, adjust the calibration gas level setting so that it is equal to the calibration span gas concentration.
8. Programming Features Detcon MicroSafe™ H S gas sensors incorporate a comprehensive program to accommodate easy operator interface and fail-safe operation. Program features are detailed in this section. Each sensor is factory tested, programmed, and calibrated prior to shipment.
See section 3.5.4-f for details on changing the RS-485 ID number. The following section explains the details of the Modbus™ protocol that the Detcon MicroSafe™ sensor supports. Code 03 - Read Holding Registers is the only code supported by the transmitter. Each transmitter contains 6 holding registers which reflect its current status.
Page 25
Model 1000 H2S Echem Sensor life is an estimated remaining use of the sensor head, between 0% and 100% Example: 85=85% sensor life Register # High Byte Low Byte 40001 Detectable Range i.e. 100 for 0-100 ppm, 50 for 0-50% LEL, etc Register # High Byte Low Byte...
9600 baud, 8 data bits, 1 stop bit, No parity, and half duplex 485. 10. Display Contrast Adjustment Detcon MicroSafe™ sensors feature a 16 character backlit liquid crystal display. Like most LCDs, character contrast can be affected by viewing angle and temperature. Temperature compensation circuitry included in the MicroSafe™...
Model 1000 H2S Echem 12. Sensor Replacement Should the electrochemical H S plug-in sensor (part number 370-323200-000 or 370-3H3H00-000) require replacement, use the following procedure: 1 - (A) If the sensor is mounted in a classified area, system power to the transmitter must first be removed before proceeding further.
Model 1000 H2S Echem Spare Parts DM-624 H S Sensor Part# Description 942-010212-XXX Calibration Span Gas (XXXppm H S, balance nitrogen) (Where XXX denotes the gas level – 010 = 10ppm, 025 = 25ppm, 050 = 50ppm) 942-010213-XXX Calibration Span Gas (XXX% H S, balance nitrogen) (Where XXX denotes the gas level –...
Detcon, Inc., further provides for a five year fixed fee service policy wherein any failed transmitter shall be repaired or replaced as is deemed necessary by Detcon, Inc., for a fixed fee of $65.00. The fixed fee service policy shall affect any factory repair for the period following the two-year warranty and shall end five years after expiration of the warranty.
Update Spare parts to reflect High Range sensor , Added new range (0-5%) to Sample Flow Rate tables 06/02/11 Updated drawings and Spare Parts List to show Detcon Aluminum condulet 09/19/11 Updated notation for VDC input in section 3.3 to match wiring...