Download Print this page

ACE INSTRUMENTS ACE one Operating Manual page 14

Hide thumbs Also See for ACE one:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 11
Battery notices
The device associated with this operating manual contains 2 AAA Alkaline batteries (1.5 V) in
delivery condition. These batteries are not made for recharging. The use of a rechargeable
battery is not possible with this device and can damage it. If the included batteries are deple-
ted, they must not be thrown into the household waste. Used batteries may contain harmful
substances that can cause environmental or health damage. Please return all depleted
batteries to electrical appliance stores, who collect them for free.
Battery exchange
Apply a small amount of pressure on the battery cover located on the back of the unit and
push down and away from the unit to release the cover. See the following diagram.
Attention: Mind the battery polarity in the process (+/-).
In the EU the return of used batteries is mandatory by law. All depleted batteries and also
rechargeable batteries will be recycled. This way valuable resources like iron, zinc and nickel
can be retrieved and are reusable then. Battery recycling is the easiest lesson in environmen-
tal protection – so please support it!
17
The shown garbage bins mean that batteries or rechargeable batteries must not be thrown
into the household waste. The signs below the garbage bins are acronyms for the chemical
substances inside the battery:
Error / Warning Notifications
Display
Cause & Action
notification
0.50
If the result is higher
than 0.5‰ (‰ = default
setting) a "WARN" signal
is displayed along with an
"ALARM" sound.
READY
BATT
WARN
LOW
Hot
If the result is above
4.00‰ a "Hot" signal is
displayed along with an
alarm sound.
READY
BATT
WARN
LOW
CAL
When you have done 500
tests, then the unit shows
"CAL". Even if the unit still
works, a re-calibration is
required. Consult ACE
READY
BATT
WARN
Instruments and send your
LOW
device to the calibration
service.
Hg: Battery contains Mercury
Cd: Battery contains Cadmium
Pb: Battery contains Lead
Display
Cause & Action
notification
bAt
The battery voltage is
too low to perform a test.
The device will turn off
automatically after showing
this signal. Replace the
READY
BATT
WARN
2 pieces of 1.5V Alkaline
LOW
Batteries.
FLo
The user had a short
exhale time during the
test, which likely means
the user did not exhale
steadily and continuously
READY
BATT
WARN
for at least 3-5 seconds.
LOW
Push the power button to
perform another test.
OUt
When the ambient tem-
perature is out of working
temperature range (5°C ~
40°C) the unit shows "OUT"
and then turns off. Keep
READY
BATT
WARN
and use the unit within the
LOW
recommended tempera-
ture range.
18