OmniSWD
Shortwave Diathermy System
User Manual
Issued 08.01.2017
Part No. 2903030M Rev 3
This manual and all content therein is owned exclusively by Accelerated Care Plus ("ACP") and is protected by copyright. This manual or any
portion thereof may not be photocopied, reproduced or translated to another language without the express prior written consent of ACP. This
manual may only be used by entities that have purchased the equipment or have implemented the ACP program and are covered by an executed
lease agreement. This manual may not be used for any other purpose.
Any additional copies of the Manual shall be ordered from ACP. No changes or modifications shall be made to the Manual without prior review
and written authorization from ACP. No authorization is given to market, sell, disclose, or exploit this Manual except as for purposes of using the
Equipment as contemplated by the Lease Agreement.
ACCELERATED CARE PLUS MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MANUAL, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Accelerated
Care Plus shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance
or use of this Manual.
The information contained in this document is subject to change without notice.
COPYRIGHT © 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED

Advertisement

Table of Contents
loading

Summary of Contents for Hanger ACP OmniSWD

  • Page 1 ® OmniSWD Shortwave Diathermy System User Manual Issued 08.01.2017 Part No. 2903030M Rev 3 This manual and all content therein is owned exclusively by Accelerated Care Plus ("ACP") and is protected by copyright. This manual or any portion thereof may not be photocopied, reproduced or translated to another language without the express prior written consent of ACP. This manual may only be used by entities that have purchased the equipment or have implemented the ACP program and are covered by an executed lease agreement.
  • Page 2 ® OMNISWD USER MANUAL ® OmniSWD ACP manufactures a premier line of rehabilitation technologies to assist health care professionals with improved outcomes and quality-of-life for Patients. The ACP product line includes Pain Control Systems, Muscle Stimulators, Interferential Therapy, Therapeutic Ultrasound, Pulsed Shortwave Diathermy devices, and advanced Therapeutic ®...
  • Page 3: Symbols On The Product

    ® OMNISWD USER MANUAL SYMBOLS ON THE PRODUCT Symbol Used for Symbol Used for Type BF medical device per: Serial number IEC 60601 Date of manufacture Manufacturer Caution, consult Caution, electrical precautions accompanying documents CE mark of confidence Classification against water IP2X compliant to MDD intrusion...
  • Page 4: Table Of Contents

    ® OMNISWD USER MANUAL TABLE OF CONTENTS SYMBOLS ON THE PRODUCT ........................1 DIATHERMY INDICATIONS & CONTRAINDICATIONS ................3 Indications ............................3 Contraindications ..........................4 DIATHERMY WARNINGS & PRECAUTIONS ..................... 5 Warnings ............................. 5 Precautions ............................7 ® THE OmniSWD ............................
  • Page 5: Diathermy Indications & Contraindications

    ® OMNISWD USER MANUAL DIATHERMY INDICATIONS & CONTRAINDICATIONS CAUTION: Federal law restricts this device for sale or use by, or on the order of, a Practitioner licensed by the laws of the state in which he/she practices to use or order the use of the device. Please notice that Accelerated Care Plus cannot provide medical advice.
  • Page 6: Contraindications

    ® OMNISWD USER MANUAL Contraindications  Do not apply shortwave diathermy if the patient does not understand the potential risks.  Do not apply shortwave diathermy if the patient is not able to cooperate with the operator in maintaining the proper position and in reporting the presence of a heating sensation which is the only indication of an adequate or excessive dose.
  • Page 7: Diathermy Warnings & Precautions

    ® OMNISWD USER MANUAL DIATHERMY WARNINGS & PRECAUTIONS CAUTION: Federal law restricts this device for sale or use by, or on the order of, a Practitioner licensed by the laws of the state in which he/she practices to use or order the use of the device. Please notice that Accelerated Care Plus cannot provide medical advice.
  • Page 8 ® OMNISWD USER MANUAL heated significantly by shortwave diathermy units. Some fabric blends that contain synthetics can also be heated. Additionally, synthetics can trap moisture near the body, which can rapidly heat and burn a patient.  Always keep cables spaced apart. Route cables as designed for the unit, using all spacing insulators and cable supports provided.
  • Page 9: Precautions

    ® OMNISWD USER MANUAL Precautions  Application site and settings should be based on the guidance of the prescribing practitioner.  Do not connect this device to any wall outlet that has not been properly grounded, or to any electrically non- isolated medical device.
  • Page 10: The Omniswd

    ® OMNISWD USER MANUAL ® THE OmniSWD ® Delivery of the OmniSWD Upon receipt of the unit check for any damage which may have occurred in transit. If any signs of damage are found then retain all packing material and inform the carrier and the Company or its agent from whom the unit was purchased.
  • Page 11: Controls And Functions

    ® OMNISWD USER MANUAL Controls and Functions Switch on the base of the stand ® FIGURE 1 - OmniSWD - Front Panel The letter associated with each control or indicator in this section corresponds to the labeling in FIGURE 1. A.
  • Page 12: Treatment Guidelines

    ® OMNISWD USER MANUAL TREATMENT GUIDELINES Inductive Shortwave A coiled conductor is placed in the drum inductive applicator which is usually in the shape of a drum. The current flowing through the coils produces a magnetic field at right angles to the alignment of the coils. The magnetic field induces more current into the muscle and less in the subcutaneous fat than the capacitive technique.
  • Page 13: Depth Of Penetration

    ® OMNISWD USER MANUAL NEWER SHIELDED DRUM INDUCTIVE APPLICATOR EACH FLAP RF SHIELD CONTAINS COIL OLDER UNSHIELDED INDUCTIVE APPLICATOR Depth of Penetration Pulsed shortwave with magnetic field monode drum inductive applicators should be used to treat deeper tissue. Penetration depth for fat at 27 MHz is approximately 62.6 inches (159cm) whereas depth for muscle is 5.5 inches (14cm).
  • Page 14: Infection Control Equipment And Principles Of Use

    ® OMNISWD USER MANUAL INFECTION CONTROL EQUIPMENT AND PRINCIPLES OF USE Definitions  Barrier Film – One-time use, disposable plastic film for use over touch/operator surfaces of equipment to reduce risks of cross-contamination and need for high level disinfection of equipment between patients. ...
  • Page 15: Cleaning And Low Level Disinfection

    ® OMNISWD USER MANUAL Cleaning and Low Level Disinfection This is a recommended daily housekeeping practice to keep the equipment clean and free of contaminants which could contribute to transmission of infection. The following practices are recommended for use when treating intact skin without the presence of physiologic fluids such as blood and urine.
  • Page 16: Use Of Barriers

    ® OMNISWD USER MANUAL Use of Barriers The use of an all-purpose barrier film that provides surface protection from cross-contamination from a variety of applications should be used whenever dealing with non-intact skin or the chance of coming in contact with bodily fluids.
  • Page 17: Modes Of Operation

    ® OMNISWD USER MANUAL MODES OF OPERATION Thermal Effect Variables Tissue Energy Dissipation The extent of the thermal effect in tissue is based on the tissue energy dissipation. Specific Absorption Rate (SAR) is a measure of the rate at which energy is absorbed by the human body when exposed to a radio frequency (RF) electromagnetic field.
  • Page 18 ® OMNISWD USER MANUAL appropriate (assuming the patient has normal sensation). Preprogrammed doses may be selected on the front panel for ease of operator selection. Safety Due to the design of the drum inductive applicator, a strong magnetic field is produced with only a small incidental amount of electrical field.
  • Page 19: Operating Instructions

    ® OMNISWD USER MANUAL Operating Instructions Numbers in circles in the following operating instructions refer to the front panel drawing FIGURE below. ® Connect the OmniSWD to a suitable AC outlet, and energize the unit at the Mains AC Power switch. This is a two position rocker switch, up for ON, down for OFF.
  • Page 20: User Interface

    ® OMNISWD USER MANUAL User Interface ® The OmniSWD utilizes a touch screen to activate selections with the touch of a finger. All screens and functions can be accessed from this ‘Home’ screen. The “Home” icon in the top left corner of all other screens will always return to this screen.
  • Page 21: Setup Procedure For Drum Inductive Applicator

    ® OMNISWD USER MANUAL Setup Procedure for Drum Inductive Applicator The drum inductive applicator has an indicator lamp which illuminates when the treatment is on. If the lamp is not illuminated, check your output mode to determine that output has been initiated. Note the lamp glows from dim to bright as the output power is increased.
  • Page 22: Testing The Output

    ® OMNISWD USER MANUAL System Fault Warning This pop-up window appears if the machine detects a communications error between its two processors. This should be a very rare event if it happens at all, and may not recur if the mains power is switched off and on again. If it persists, then contact ACP Customer Support at (1- 800-350-1100).
  • Page 23: Output Tuning Is Recommended (>10% Variance)

    ® OMNISWD USER MANUAL Tuning POP-UPs Output Tuning is Recommended (>10% variance) If the Output tuning is greater than >10% variance, but less than < 20%, an “Output tuning is recommended” caution will pop –up on the screen. User has the option to select IGNORE to continue without tuning, or TUNE to be taken to the Device Output Tuning screen.
  • Page 24: Drum Inductive Applicator Removal

    ® OMNISWD USER MANUAL Drum Inductive Applicator Removal It may be necessary to remove the Drum Applicator from the articulating arm to apply to a patient lying in a bed. To remove the Drum Applicator, first loosen the locking collar turning counter-clockwise until you can see the white nylon washer as shown.
  • Page 25: Troubleshooting

    ® OMNISWD USER MANUAL TROUBLESHOOTING ® The following table lists OmniSWD problem symptoms and possible areas to check for the problem causes. If these suggested measures do not correct the machine malfunction, call your ACP Customer Support representative. PROBLEM CAUSE REMEDY ...
  • Page 26: Technical Specifications

    ® OMNISWD USER MANUAL TECHNICAL SPECIFICATIONS ® Technical Data - OmniSWD GENERAL: Power Supply: 120VAC 50/60Hz Power consumption: 250W max Externally Replaceable Fuses: 2 x T 2A 20mm Size: Approx. H: 870 mm (34.25 in) / D: 410 mm (16 in) / W: 410 mm (16 in) Weight: 39 kg (86 lbs.) FDA Classification:...
  • Page 27: Diagrammatic Representation Of Pulse Trains

    ® OMNISWD USER MANUAL Diagrammatic Representation of Pulse Trains 20 MICROSECOND (Sec) PULSES - 800 PULSES PER SECOND (PPS) 65 Sec PULSES - 400 PPS 200 Sec PULSES - 200 PPS PULSE PULSE “ON” “OFF” PERIOD PERIOD 400 Sec PULSES - 100 PPS CONTINUOUS TRAIN OF PULSES: PULSES CONTAIN 27.12 MHz SINUSOIDAL R.F.
  • Page 28: Omniswd Average Output Power

    ® OMNISWD USER MANUAL ® OmniSWD Average Output Power PULSE DURATION AVERAGE POWER IN WATTS (W) EFFECT 400 Sec 4.0W 8.0W 16.0W 32.0W 48.0W 64.0W 200 Sec 2.0W 4.0W 8.0W 16.0W 24.0W 32.0W 100 Sec 1.0W 2.0W 4.0W 8.0W 12.0W 16.0W 65 Sec .65W...
  • Page 29: Omniswd Standard And Optional Accessories

    ® OMNISWD USER MANUAL ® OmniSWD STANDARD AND OPTIONAL ACCESSORIES ITEM ITEM NO. DESCRIPTION ® OmniSWD Shortware Diathermy System 200 Watt peak power Shortwave Diathermy 1903030 with Delta T control Shipping Weight: 101 lbs. ITEM ITEM NO. DESCRIPTION ® OmniSWD Locking Power Cord, Hospital 89456 Grade, 10ft.
  • Page 30: Standard Limited Product Warranty

    ® OMNISWD USER MANUAL STANDARD LIMITED PRODUCT WARRANTY The warranty information provided in this section is applicable only to products purchased from ACP, directly or through an authorized dealer. This section does not apply to leased products. The terms of maintenance and repair of any leased products are detailed in the separately executed agreement between the parties.
  • Page 31 ® OMNISWD USER MANUAL Return of Defective Equipment Any Equipment returned to the ACP Service Center under warranty coverage must have the Warranty coverage validated and must receive authorization from ACP Customer Support prior to being received at the Service Center. Shipping charges, insurance, and any other costs incurred in sending product to ACP Service Center is the responsibility of the customer and will not be refunded.