8. Whitening procedure – operation of the lamp
Plug power cable in electrical outlet 100-240 V.
Switch on power. See instruction of WHITEsmile
gel how to apply the fläsh whitening gel.
The following special instruction for working in
combination with fläsh Whitening Lamp must be
followed: All teeth that should not be bleached and
are therefore not covered with gel must be covered
with moist gauze in order to prevent heating. (to be
covered have to be restorations like crowns, bridges,
implant suprastructures and veneers or in case
of single tooth whitening the other teeth). Gingiva
protector can be used for better fixation.
Restorations must
be covered with a
moist gauze.
fläsh Light Whitening Gel becomes clear after a
few minutes, showing the strong activation of
the fläsh Whitening Lamp. The gel is still active
and remains on the teeth until the end of the
cycle and the lamp goes off.
16
At single tooth whitening
all other teeth must be
covered with moist gauze.
Working times of different fläsh gels:
fläsh Light Whitening 6 % HP:
4 x 15 minutes in one session
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
fläsh Light Whitening 32 % HP:
3 x 15 minutes (maximum 4 x 15 minutes)
in one session
Place lamp in front of the patient slightly
touching the lip retractor.
Make sure that patient and operator are wearing orange
safety goggles according DIN EN 166:2002-04.