Page 2
Fluke Biomedical. Copyright Release Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Page 3
United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.
Page 4
Email: service.international@fluke.com Certification This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards us- ing accepted test procedures.
Tester Familiarization ................4 Front Panel..................4 Back Panel ..................4 Accessories ................... 5 Operation Overview................5 Operating the DPM1B ..............6 Setup ....................6 Zeroing the Monitor and Tester ............7 Pressure Measurement ..............8 Testing Transducer Linearity ..............10 Transducer Sensitivity Test..............
Page 6
Multiplication Factors............. 12 List of Figures Figure Title Page DPM1B Front Panel ............5 DPM1B to Monitor Setup..........7 DPM1B Vented to Atmosphere........9 DPM1B Configured to Apply Pressure to Transducer ... 10 Sample Plot of Transducer Linearity Using the DPM1B 11...
DPM1B Pneumatic Transducer Tester Introduction This document is the Operators manual for the DPM1B Pneumatic Transducer Tester (hereafter “the Tester”). It contains general information about the Tester, a description of its components and instructions or documentation required to service the unit. If a problem develops, the user should contact Fluke Biomedical.
DPM1B Operators Manual Summary of Features The Tester is a hand-held pressure measuring device, designed as an aid in the performance testing and troubleshooting of pneumatic or hydraulic systems. A liquid crystal display (LCD) reading indicates pressures manually generated from an internal chamber.
Pneumatic Transducer Tester Specifications Tester using a Y-connector. Pressure is applied to the system as normal, usually with a squeeze bulb. The results can be compared between the Tester and the manometer under test. Testing Capabilities Verifying the proper operation of a blood pressure transducer can be a tedious and time-consuming task.
DPM1B Operators Manual Display..............0.5 “ (1.3 cm) LCD with LO BAT indicator Dimensions ............5.8 “ x 3.6 “ x 1.5 “ (14.6 cm x 15.9 cm x 3.8 cm) Weight..............10 oz. (260 g) Tester Familiarization The Tester incorporates a highly accurate digital pressure meter and pneumatic pressure-generating cylinder into a small, hand-held package for determining the accuracy of most blood pressure measurement systems.
Connection Site Thumbwheel ehf1.eps Figure 1. DPM1B Front Panel Operation Overview When operating according to the instructions given in this section, the Tester produces and displays a pressure for comparison to the displayed pressure on the monitor the blood pressure transducer under test is connected to. A conversion table is found in Multiplication Factor section that quickly provides the multiplication factors to compare various units of measure.
DPM1B Operators Manual Operating the DPM1B Occasionally a blood pressure transducer under test may not have the sensitivity specified by the manufacturer of the monitor the transducer under test is connected to. By matching gain adjustments on the monitor, deviations in sensitivity can usually be compensated for.
Filter BP Input ehf2.eps Figure 2. DPM1B to Monitor Setup 3. When the system is to be filled with liquid, position Stopcock 1 as shown in Figure 2. 4. Attach a liquid-filled syringe to the top port of Stopcock 2.
DPM1B Operators Manual Note Allow approximately 3 minutes warm-up time before taking readings; this allows the electronics to stabilize in the Tester. 4. Zero the Tester by turning the ZERO knob on the front panel of the Tester (See Figure 1 for location of ZERO knob).
Page 15
Pneumatic Transducer Tester Operation Overview DPM-1B Filter Test Transducer Vent Port ehf3.eps Figure 3. DPM1B Vented to Atmosphere To interpret the results, proceed to the following three sections: • Testing Transducer Linearity • Transducer Sensitivity Test • Testing Other Devices...
Test Transducer Vent Port ehf4.eps Figure 4. DPM1B Configured to Apply Pressure to Transducer Testing Transducer Linearity Several measurements are required to determine if a transducer is linear over a range of pressure. See Figure 5 for a sample plot of transducer linearity determined by using the Tester Pneumatic Transducer Tester.
Step 3. If the readings from Step 2 are within the range calculated in Step 3, the transducer is within the manufacturer’s specifications. Linear Plot DPM-1B Readings (mmHg) Nonlinear Plot Monitor Readings(mmHg) ehf5.eps Figure 5. Sample Plot of Transducer Linearity Using the DPM1B...
DPM1B Operators Manual Testing Other Devices A variety of pressure-measuring and pressure-generating devices can be tested using the Tester. The Tester can be used as a pressure standard to easily test water manometers, mercury manometers, and sphygmomanometers. Set up the system as described in the section titled Pressure Measurement.
Maintenance Maintenance Cleaning Outside Surfaces The DPM1B outside surfaces can be cleaned with a cloth dampened with water and mild detergent or alcohol. Battery Replacement The battery compartment of the Tester is found in the rear panel of the tester.
Need help?
Do you have a question about the DPM1B and is the answer not in the manual?
Questions and answers