CALIBRATION
When the scale is operated for the first time, and when it is moved to a new location, a calibration is recommended
to ensure accurate weighing results.
CALIBRATION WEIGHTS
Before performing the calibration, be sure to have the appropriate calibration weights.
CALIBRATION VALUES
ULINE MODEL NO.
Span Calibration Mass (g)
Linearity Calibration Mass (g)
SECURITY SWITCH
Ensure the security switch is set to the unlocked position.
(See Figure 7)
Locked
Unlocked
SPAN CALIBRATION (SINGLE POINT)
Span calibration uses two calibration points: zero and a
specified calibration weight.
1.
Remove all weight from the platform.
2. Press and hold MENU until
the button is released, the display will show
If no APW exists, the balance will display
proceed to step 5.
3. Press YES to accept.
4. Press YES to begin the span calibration.
displayed while zero reading is stored.
5. Next, the display shows the calibration weight value
(press NO to toggle value). Place the specified
calibration mass on the pan.
while the reading is stored.
6. The display will show
successful. The balance returns to the previous
application mode and is ready for use.
PAGE 4 OF 33
OPERATION CONTINUED
H-7293
H-5848
H-5849
0, 200
0, 200
0, 400
0, 100,
0, 100,
0, 200,
200
200
400
MENU
is displayed. When
Put.10
SPAN
will then be shown.
--C--
--C--
will be displayed
donE
if the calibration was
H-5850
H-7294
0, 600
0, 2,000
0, 300,
0, 1,000,
600
2,000
LINEARITY CALIBRATION (MULTI-POINT)
Linearity calibration uses three calibration points: zero,
mid-scale and full scale.
1.
Remove all weight from platform.
Figure 7
2. Press and hold MENU until
the button is released, the display will show
3. Press YES to accept.
4. Press NO until
5. Press YES to begin the linearity calibration.
be displayed while zero reading is stored.
6. Next, the display shows the calibration weight value
(press NO until the mid-scale value is shown). Place
the specified calibration mass on the pan.
be displayed while the reading is stored.
7.
Next, the display shows the calibration weight value
(press NO until the full-scale value is shown). Place
the specified calibration mass on the pan.
be displayed while the reading is stored.
8. The display will show
successful. The balance returns to the previous
application mode and is ready for use.
ZERO
C.A.L.
,
If there is a minor weight displayed without anything on
the platform, press the ZERO button to clear the display.
Zero indicator will be shown on the display.
TARE
will be
When weighing an item that is in a container, taring
stores the container weight to memory so only the item
weight is displayed.
To store the tare weight into memory and set the display
to zero:
1.
Place an empty container on the platform.
H-5851
H-5852
H-5853
0, 400
0, 600
0, 2,000
0, 200,
0, 300,
0, 1,000,
400
600
2,000
MENU
SPAN
will then be shown.
LIn
is shown.
donE
if the calibration was
H-5854
H-5855
0, 6,000
0, 8,000
0, 3,000,
0, 4,000,
6,000
8,000
is displayed. When
C.A.L.
--C--
will
--C--
will
--C--
will
0218 IH-5848