Hanwei JL269LED Operation Manual

Portable gas leak detector

Advertisement

JL269LED
Portable gas leak detector
Operation manual
Please read this manual carefully before operation
Ver: HWWM161229CG

Advertisement

Table of Contents
loading

Summary of Contents for Hanwei JL269LED

  • Page 1 JL269LED Portable gas leak detector Operation manual Please read this manual carefully before operation Ver: HWWM161229CG...
  • Page 3 Safety Information Before using the device, please read this manual carefully and follow the operation requirements as below:  Please don't use the defective detector. Before using, please check if there is crack or spare parts’ missing. If there is, please contact the seller.
  • Page 4 be discarded by qualified withdrawers.  Please change the rechargeable batteries if they have been used for 2 years or charged for 300 times.  It's forbidden to disassemble, adjust or repair the device without permission.  Please avoid the device falling down from high place or serious shocking.
  • Page 5: Table Of Contents

    Table of contents 1. BRIEF INTRODUCTION ..............- 1 - 2. MAIN TECHNICAL SPECIFICATIONS ..........- 1 - 3. STRUCTURE AND FUNCTIONS ............- 2 - 4. OPERATION ................... - 2 - 4.1 Power on ..................- 2 - 4.2 Impact Test ................. - 3 - 4.3 Detecting ..................
  • Page 6: Brief Introduction

    1. Brief introduction JL269 gas detector has a fast and stable performance and wide detecting range. It can be used to detect methane, propane, natural gas, LPG, and hydrogen gas, It will help you to find out the gas leakage sources easily. Main function &...
  • Page 7: Structure And Functions

    Working Time: >8h(working continuously in normal working status) Sensor Life: 2 years Dimensions: 180mm× 72mm× 36mm Weight: About 300g 3. Structure and Functions 4. Operation 4.1 Power on Press and hold “ ” for more than 2 seconds, the detector will power on and enters into self-testing status.
  • Page 8: Impact Test

    or zero translation doesn’t finish within 5 seconds, please keep waiting. Check “ OL” light. After it turns off, zero translation finishes and the detector will enter into detecting status. 4.2 Impact Test It is suggested to do Impact Test before using every day, so as to make sure the detector work normally.
  • Page 9: Open / Close Audible Alarm

    Color Voltage Green Enough Yellow Not enough Low voltage, need charge 4.5 Open / close audible alarm In detecting status, the initial setting of audible alarm is on. The more gas leaks, the higher the audible alarm frequency will be. If the detecting spot is too noisy, operator can use the earphone purchased by himself.
  • Page 10: Charging

    calibration is done. After zero calibration, one yellow LED will be lighted on which means the detector enters into the calibration status of the second point. Put the detector into the standard gas (level equals to the second point) and keep it steady for about 30 seconds. Then press to save the second point information.
  • Page 11: Sensor Using And Replacement

     After the detector turns off automatically, please charge the battery within 12 hours, so as to avoid that the battery cannot work due to too voltage too much low.  If the detector won’t be used for a long time, please take out the batteries and put it in dry environment.