INSTRUCTION MANUAL
HI98318
Waterproof EC/TDS Tester
Thank You
Thank you for choosing a Hanna Instruments product. Please read
this instruction manual carefully before using the instrument.
For more information about Hanna and our products,
visit www.hannainst.com or e-mail us at sales@hannainst.com.
For technical support, contact your local Hanna office or e-mail us
at tech@hannainst.com
Find your local Hanna office on www.hannainst.com
Preliminary Examination
Remove the meter from the packing material and examine it carefully to make sure
that no damage has occurred during shipment. If noticeable damage is evident, contact
your local Hanna office.
Each meter is supplied with:
• CR2032 battery (1 pc.)
• Storage / Protection sleeve
• Instruction manual
• Quality Certificate
• Quick Cal pH and EC one point calibration solution sachet (4 pcs.)
Note: Save all packing material until you are sure that the instrument functions
correctly. All defective items must be returned in the original packaging with the
supplied accessories.
Preparation
The probe is shipped dry. Before using the electrode, remove the protective cap. Then
follow the calibration procedure.
• Turn the tester on by pressing ON/OFF button.
• Remove the protective cap and immerse the tip of the probe in the sample to
be tested.
• Stir gently and wait for a stability tag to disappear.
• The electrode automatically compensates for temperature variations.
• The reading on display is directly expressed in: mS/cm or ppm.
• For best results, recalibrate periodically.
• After use rinse the probe with water.
• Always replace the protective cap after each use.
Operation
Liquid Crystal
Display
CAL button
ON/OFF button
EC Probe
Temperature
Sensor (inside)
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.
Specifications
0.00 to 6.00 mS/cm
0 to 3000 ppm (0.5)
Range
EC/TDS
0 to 4000 ppm (0.7)
0.0 to 50.0 °C (32.0 to 122.0 °F)
0.01 mS/cm
10 ppm (0.5)
Resolution
EC/TDS
10 ppm (0.7)
0.1 °C (0.1 °F)
±2% F.S.
Accuracy
EC/TDS
(@25 °C/77 °F)
±0.5 °C (±1.0 °F)
TDS Conversion factor
0.5 (500 ppm) or
(CF)*
0.7 (700 ppm)
Calibration
HI50036
HI70031
Solutions
(Quick Cal)
(1.41 mS/cm)
Calibration
automatic,single point (1.41 mS/cm)
one point calibration using Quick Cal pH and EC
Quick calibration
one point calibration solution
HI50036
Temperature
automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF)
Compensation
Battery type
CR2032 Li-Ion 3V (1 pc.)
Battery life
approximately 250 hours of continuous use
0 to 50 ºC (32 to 122 ºF);
Environment
RH 100% max
Weight
68 g (2.4 oz.)
(without battery)
*1000 µS/cm =500 ppm with 0.5 CF
Hanna Instruments reserves the right to modify the design,
construction, or appearance of its products without advance notice.
Need help?
Do you have a question about the Gro Line HI98318 and is the answer not in the manual?
Questions and answers