7 Remote Operation

Using the Relay Control
7.1
NOTE:
1.
Flip the Main Power ON/OFF (I/0) switch to the OFF (0) position and
remove the power cord from the unit.
2.
Using the Control Cable (P/N 38000-M03-022), apply a 5-24VDC
control voltage (from a PLC, computer I/O device, or equivalent) to the
Input connector on the rear panel. Refer to the image below for
connection information.
RJ10 -
CONTROL
Plug – pin end view
NOTE:
NOTE:
38000-L01-001_Rev_J_SugarCUBE_Manual
Relay Control and the On/Stby button can both be
used to turn the LED on or off. The unit must be in
Standby Mode to ensure proper operation while using
the Relay Control. The necessary accessory cable can
be purchased at
www.ushio.com
1 - Black
2 - Red
3 - Green
4 - Yellow
USER SUPPLIED
TERMINAL STRIP
The Relay Control terminals are NOT polarity specific.
For units having the Analog Control option:
Pin 1 – Black is Analog GND
Pin 4 – Yellow is Analog Control Voltage
5 – 24 VDC
N/C
STBY
ON
N/C
15