Figure 1–1
The front panel of the
NIBP Controller
14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Checking the NIBP Controller

The unit passes quality control inspection before leaving the factory. However,
there is a small chance that damage may occur in transit.
1. Check there are no obvious signs of damage to the outside casing.
2. Check there are no obvious signs of internal damage (like rattling).
If you find a problem, please contact your ADInstruments distributor
immediately.

The Front Panel

Cuff connection
NIBP Controller
Cuff
Status
Power
Power indicator

Status indicator

The Power Indicator

The Power indicator will light when the unit is on. If not, check that the unit
is properly connected to the PowerLab, and that the PowerLab is properly
connected to a power socket and is switched on.
Status Indicator
The Status indicator light can be in one of three states:
Off: Idle, no measurement in progress
Green: Inflating cuff
Orange: Deflating cuff

Cuff Pressure Connection

This port has a luer lock fitting and is where the tail cuff is attached. The
NIBP system can be purchased with either a cuff suitable for rats or a cuff
Filter setting switch
Pulse Input
Pulse Range (BPM)
40 - 150
90 - 420
240 - 600
Pulse Input connector
NIBP Controller Owner's Guide
Start
Stop
Start/Stop button