BIOCON-1100 QUICK START GUIDE
When not in use, the BioCon-1100 probe should be securely stored in its trolley or carry case.
SCANNER OVERVIEW
Power on/off
Touchscreen display
POWER & CHARGING
Press the POWER button once to switch on. Mains or battery operated: recharge when the battery level indicator is on one bar.
HOME SCREEN & SELECT PATIENT TYPE
Saved scan results/
Help
storage capacity
Adult
Child
Scan depth 23cm
Scan depth 8cm
PROBE PREPARATION - USE ONLY A RECOGNISED ULTRASOUND TRANSMISSION GEL
Clean the probe before and after with detergent wipes in
accordance with manufacturer's guidelines. Apply
ultrasound gel to the probe.
Ensure the patient is in the supine position with abdominal
muscles relaxed.
Position the probe midline on the patient's abdomen
approximately 3cm above the pubic bone.
Ensure the patient icon on the top of the probe is facing
towards the patient's head/feet.
Angle the probe slowly under the patient's pubic bone –
towards the patient's feet (without moving the head of the
probe).
Press scan (once) either on the probe or on the console
touchscreen – you are now in the pre-scan window.
It is not necessary to keep moving the probe head just a slight adjustment of angle or change of pressure.
Please refer to the operator's manual for full instructions.
Insert SD card
Connect barcode reader
Time and battery
percentage
Scan
View saved
scans
Change settings
Child setting: if the bladder (black image) encroaches
beyond the yellow child scan depth marker, select exit and
scan again using the appropriate adult setting (for a larger
scan depth of 23cm).
Probe connector
Connect mains lead
Male setting: for any adult male.
Female setting: allows the scanner to compensate for the
presence of fluid in the uterus (approx. ±10ml).
Female H setting: for females who have had a full hysterectomy.
Child setting: a shorter scan depth for any patient less than
25kg/55lbs (includes neonates).
Yellow child scan depth marker indicates scan depth (8cm).
Scan button
Probe

Advertisement

loading

Summary of Contents for de smit medical CUBEscan BioCon-1100

  • Page 1 BIOCON-1100 QUICK START GUIDE Please refer to the operator’s manual for full instructions. When not in use, the BioCon-1100 probe should be securely stored in its trolley or carry case. SCANNER OVERVIEW Power on/off Insert SD card Probe connector Scan button Touchscreen display Connect barcode reader Connect mains lead...
  • Page 2 MANUFACTURER’S GUIDANCE AND TRUST RECOMMENDATIONS. OTHER SUPPORTING MATERIALS ARE AVAILABLE. de Smit Medical Systems, Unit 15 Falcons Gate, Dean Road, Yate, Bristol, BS37 5NH Tel: 0845 345 4226 || Fax: 0845 345 4227 || Email: sales@desmitmedical.com || Web: www.desmitmedical.com V4 – 22/07/2020 - Z:\PRODUCTS\MCUBE\BIOCON-1100\QRG– Page 2...